2026-W44 · 2026-10-26 to 2026-11-01
FDA enforcement, week of 2026-10-26
In 2026-W44 (2026-10-26 to 2026-11-01), FDA posted 10 warning letters from its drug, biologics and device offices, and listed 224 Form 483s in its Electronic Reading Room for the first time. FDA's Enforcement Report added 120 drug and device recalls, 16 of them Class I. FDAgent recorded 1 new Official Action Indicated (OAI) classification in FDA's inspection data. FDA published 2 firm entries on its import alert lists, all of them on a Red List.
- Warning letters posted
- 10
- Posted 2026-10-27
- Form 483s listed
- 224
- In the reading room · plus 3 re-posted
- Recalls
- 120
- 16 Class I · 103 II · 1 III
- New OAI classifications
- 1
- Newly listed by FDA
Ask FDAgent about this week
FDAgent reads the warning letters FDA posted this week and answers in plain English, with links to the letters.
Warning letters
10 letters FDA posted on its warning letters page this week, from its drug, biologics and device offices. Each letter's issue date can be weeks before its posting date. Warning letters last synced Nov 22, 2026, 04:45 UTC.
Drug Product/Misbranded, False & Misleading Claims
The Office of Prescription Drug Promotion (OPDP) · issued 2026-08-12 · posted 2026-10-27
Unapproved New Drugs/Unlicensed Biological Product Violations
CBER · issued 2026-09-25 · posted 2026-10-27
Unapproved New Drugs/Unlicensed Biological Product Violations
CBER · issued 2026-10-13 · posted 2026-10-27
CGMP/QSR/Medical Devices/Adulterated
CDRH · issued 2026-10-02 · posted 2026-10-27
Northland Nuclear Medicine, LLC
CGMP/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-10-22 · posted 2026-10-27
Medical Device/Adulterated/Misbranded/Lacks PMA and/or 510(k)
CDRH · issued 2026-08-26 · posted 2026-10-27
Investigational Device Exemptions (IDE)/Premarket Approval Application (PMA)
CDRH · issued 2026-10-05 · posted 2026-10-27
Sanrace Biotechnology Co., Ltd.
CGMP/Finished Pharmaceuticals/Adulterated
CDER · issued 2026-10-22 · posted 2026-10-27
Request for Records/CGMP/Finished Pharmaceuticals/Adulterated/OTC
CDER · issued 2026-10-22 · posted 2026-10-27
An FEI links to its facility page only when the letter names exactly one FEI and FDAgent holds FDA records for it.
Import alert entries
2 firm entries FDA published on its import alert pages this week, dated by each entry's date published: 2 on a Red List. FDA lists firms by name and address; FDAgent links an entry to a facility only when FDA printed the firm's FEI. Import alerts last synced Nov 20, 2026, 03:55 UTC.
66-40Red List1 entryAlert on fda.govShow allDetention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPsAlert on fda.gov
- Published 2026-10-27
COHESION BIOTEC PRIVATE LIMITED
INDIA
95-05Red List1 entryAlert on fda.govShow allDetention Without Physical Examination of Electronic Products that Fail to Comply with Performance Standards or to Have Required CertificationAlert on fda.gov
- Published 2026-10-29
JINAN CONSURE ELECTRONIC TECHNOLOGY CO LTD
CHINA
Form 483s listed in FDA's Electronic Reading Room
224 Form 483s FDA listed for the first time in its Electronic Reading Room this week, dated by the publish date FDA gives each. FDA sometimes lists a 483 days after that date, so a week can still gain 483s after it ends. FDA also re-posted 3 Form 483s it had listed before. Re-posts don't count as new 483s. Each is marked below. FDA posts only some 483s, often months after the inspection. Reading room last synced Nov 22, 2026, 15:42 UTC.
- Listed 2026-10-30
FDA record date 2019-08-02
- Listed 2026-10-30
FDA record date 2019-09-18
- Listed 2026-10-30
Medlmmune UK Limited (AstraZeneca/Supply Biologics)
FDA record date 2022-06-14
- Listed 2026-10-30
FDA record date 2023-11-10
- Listed 2026-10-30
Boehringer lngelheim RCV GmbH & Co KG
FDA record date 2023-12-05
- Listed 2026-10-30
Autolus Limited (referred to as The Nucleus)
FDA record date 2024-04-30
- Listed 2026-10-30
FDA record date 2024-09-18
- Listed 2026-10-30
FDA record date 2024-11-22
- Listed 2026-10-30
Baxalta Belgium Manufacturing SA
FDA record date 2024-11-22
- Listed 2026-10-30
InVia Fertility Specialists, PLLC
FDA record date 2025-02-05
- Listed 2026-10-30
FDA record date 2025-02-20
- Listed 2026-10-30
FDA record date 2025-02-24
- Listed 2026-10-30
Lin, James P MD Inc. d.b.a. Reproductive Feritlity Center
FDA record date 2025-03-12
- Listed 2026-10-30
FDA record date 2025-04-09
- Listed 2026-10-30
Lifesouth Community Blood Centers - Montgomery Region
FDA record date 2025-04-14
- Listed 2026-10-30
Baptist Health Deaconess Madisonville Inc
FDA record date 2025-04-24
- Listed 2026-10-30
FDA record date 2025-05-07
- Listed 2026-10-30
FDA record date 2025-05-09
- Listed 2026-10-30
FDA record date 2025-05-20
- Listed 2026-10-30
FDA record date 2025-05-30
Show 207 more
- Listed 2026-10-30
Lux Therapeutics LLC dba Ponya Therapeutics LLC
FDA record date 2025-06-06
- Listed 2026-10-30
FDA record date 2025-07-15
- Listed 2026-10-30
FDA record date 2025-07-24
- Listed 2026-10-30
Lake Superior Centre For Regenerative Medicine
FDA record date 2025-09-12
- Listed 2026-10-30
FDA record date 2026-03-06
- Listed 2026-10-30
FDA record date 2026-04-30
- Listed 2026-10-30
CellTrans Islet Isolation facility
FDA record date 2021-06-11
- Listed 2026-10-30
FDA record date 2024-03-12
- Listed 2026-10-30
FDA record date 2024-03-15
- Listed 2026-10-30
FDA record date 2024-04-30
- Listed 2026-10-30
FDA record date 2024-12-06
- Listed 2026-10-30
Central California Blood Center/ Clovis Donor Center
FDA record date 2024-12-11
- Listed 2026-10-30
FDA record date 2024-12-12
- Listed 2026-10-30
Jackson Memorial Hospital Transfusion Medicine Services
FDA record date 2024-12-17
- Listed 2026-10-30
FDA record date 2024-12-17
- Listed 2026-10-30
FDA record date 2025-02-07
- Listed 2026-10-30
Lions Medical Eye Bank & Research Center of Eastern VA
FDA record date 2025-02-11
- Listed 2026-10-30
FDA record date 2025-04-02
- Listed 2026-10-30
Clinical Laboratories of Hawaii Inc
FDA record date 2025-05-06
- Listed 2026-10-30
Ascension Saint John Medical Center
FDA record date 2025-05-09
- Listed 2026-10-30
FDA record date 2025-06-17 · Re-posted: FDA listed this 483 before, so it is not counted again
- Listed 2026-10-30
Aurora St. Luke's Medical Center
FDA record date 2025-06-24
- Listed 2026-10-30
FDA record date 2025-11-25
- Listed 2026-10-27
FDA record date 2012-06-21
- Listed 2026-10-27
Baxalta Us Inc. (Baxalta, now part of Shire Plc)
FDA record date 2016-06-24
- Listed 2026-10-27
Baxalta Belgium Manufacturing S.A.
FDA record date 2019-05-31
- Listed 2026-10-27
FDA record date 2023-11-17
- Listed 2026-10-27
Baxter Pharmaceutical Solutions LLC
FDA record date 2024-07-26
- Listed 2026-10-27
FDA record date 2024-09-19
- Listed 2026-10-27
Fertility Care of Orange County, A Medical Corporation
FDA record date 2024-10-07
- Listed 2026-10-27
Janssen Biopharmaceuticals, Inc.
FDA record date 2024-11-01
- Listed 2026-10-27
FDA record date 2024-11-27
- Listed 2026-10-27
Philips Medical Systems Nederland B.V.
FDA record date 2025-01-17
- Listed 2026-10-27
FDA record date 2025-01-28
- Listed 2026-10-27
FDA record date 2025-02-14
- Listed 2026-10-27
FDA record date 2025-04-28
- Listed 2026-10-27
FDA record date 2025-04-30
- Listed 2026-10-27
FDA record date 2025-05-09
- Listed 2026-10-27
FDA record date 2025-05-09
- Listed 2026-10-27
Central Coast Multispecialty Medical Group, Inc.
FDA record date 2025-05-16 · Re-posted: FDA listed this 483 before, so it is not counted again
- Listed 2026-10-27
FDA record date 2025-06-26
- Listed 2026-10-27
Center for Translational Medicine
FDA record date 2025-07-11
- Listed 2026-10-27
FDA record date 2025-07-30
- Listed 2026-10-27
FDA record date 2025-07-31
- Listed 2026-10-27
LifeSouth Community Blood Centers, Inc.
FDA record date 2025-08-18
- Listed 2026-10-27
Lightbody Medical Technologies Inc dba Artemis Surgical
FDA record date 2025-08-27
- Listed 2026-10-27
FDA record date 2025-10-24
- Listed 2026-10-27
Collaborative Care Diagnostics, LLC.
FDA record date 2026-02-09 · Re-posted: FDA listed this 483 before, so it is not counted again
- Listed 2026-10-27
FDA record date 2026-04-17
- Listed 2026-10-27
FDA record date 2026-04-20
- Listed 2026-10-27
FDA record date 2026-04-28
- Listed 2026-10-27
FDA record date 2026-04-30
- Listed 2026-10-27
FDA record date 2025-02-19
- Listed 2026-10-27
FDA record date 2025-12-16
- Listed 2026-10-26
FDA record date 2015-05-29
- Listed 2026-10-26
FDA record date 2017-06-22
- Listed 2026-10-26
Plasma Industries Belgium SCRL
FDA record date 2018-03-16
- Listed 2026-10-26
Pliva Croatia Ltd. c/o Teva Pharmaceuticals, USA
FDA record date 2018-05-31
- Listed 2026-10-26
GlaxoSmithKline Biologicals SA
FDA record date 2019-06-24
- Listed 2026-10-26
FDA record date 2019-08-23
- Listed 2026-10-26
FDA record date 2019-12-06
- Listed 2026-10-26
FDA record date 2020-03-05
- Listed 2026-10-26
Serum Institute of India Pvt. Ltd.
FDA record date 2022-04-19
- Listed 2026-10-26
ID Biomedical Corporation of Quebec
FDA record date 2022-05-20
- Listed 2026-10-26
FDA record date 2022-06-23
- Listed 2026-10-26
FDA record date 2020-01-28
- Listed 2026-10-26
Octapharma Pharmazeutika Produktionsgesellschaft m.b.H. OPG
FDA record date 2022-12-16
- Listed 2026-10-26
FDA record date 2023-02-03
- Listed 2026-10-26
Octapharma Produktionsgesellschaft Deutschland mbH.
FDA record date 2023-03-21
- Listed 2026-10-26
FDA record date 2023-03-27
- Listed 2026-10-26
FDA record date 2023-05-19
- Listed 2026-10-26
FDA record date 2023-05-26
- Listed 2026-10-26
FDA record date 2023-07-27
- Listed 2026-10-26
Charles River Laboratories Inc
FDA record date 2023-08-25
- Listed 2026-10-26
FDA record date 2023-08-29
- Listed 2026-10-26
FDA record date 2023-09-08
- Listed 2026-10-26
FDA record date 2023-09-19
- Listed 2026-10-26
FDA record date 2023-09-29
- Listed 2026-10-26
FDA record date 2023-09-29
- Listed 2026-10-26
FDA record date 2023-12-15
- Listed 2026-10-26
FDA record date 2024-03-01
- Listed 2026-10-26
FDA record date 2024-03-06
- Listed 2026-10-26
FDA record date 2024-06-07
- Listed 2026-10-26
Bio-Rad Medical Diagnostics GmbH
FDA record date 2024-06-14
- Listed 2026-10-26
Grifols Diagnostic Solutions Inc.
FDA record date 2024-06-28
- Listed 2026-10-26
Neurotech Pharmaceuticals, Inc.
FDA record date 2024-07-26
- Listed 2026-10-26
Argon Medical Productions & Vertriebs Gesellschaft MBH & Co
FDA record date 2024-07-31
- Listed 2026-10-26
FDA record date 2024-08-13
- Listed 2026-10-26
FDA record date 2024-08-23
- Listed 2026-10-26
FDA record date 2024-09-12
- Listed 2026-10-26
FDA record date 2024-10-16
- Listed 2026-10-26
Rejuvenating Fertility Center PLLC/AllCare Fertility Center
FDA record date 2024-10-25
- Listed 2026-10-26
FDA record date 2024-10-29
- Listed 2026-10-26
FDA record date 2024-10-30
- Listed 2026-10-26
FDA record date 2024-11-01
- Listed 2026-10-26
FDA record date 2024-11-08
- Listed 2026-10-26
FDA record date 2024-11-12
- Listed 2026-10-26
Reproductive Technologies, Inc.
FDA record date 2024-11-18
- Listed 2026-10-26
FDA record date 2024-11-21
- Listed 2026-10-26
FDA record date 2024-11-25
- Listed 2026-10-26
FDA record date 2024-11-25
- Listed 2026-10-26
Advocate Lutheran General Hospital Blood Bank
FDA record date 2024-12-05
- Listed 2026-10-26
Dominion Fertility & Endocrinology
FDA record date 2024-12-06
- Listed 2026-10-26
FDA record date 2024-12-11
- Listed 2026-10-26
Grifols Diagnostic Solutions Inc.
FDA record date 2024-12-13
- Listed 2026-10-26
FDA record date 2024-12-13
- Listed 2026-10-26
FDA record date 2024-12-13
- Listed 2026-10-26
FDA record date 2024-12-19
- Listed 2026-10-26
FDA record date 2024-12-20
- Listed 2026-10-26
FDA record date 2024-12-20
- Listed 2026-10-26
Pacific Reproductive Center, Inc.
FDA record date 2025-01-13
- Listed 2026-10-26
FDA record date 2025-01-21
- Listed 2026-10-26
FDA record date 2025-01-30
- Listed 2026-10-26
FDA record date 2025-01-31
- Listed 2026-10-26
Reproductive Science Center of the San Francisco Bay Area, a Medical Corporation
FDA record date 2025-02-05
- Listed 2026-10-26
FDA record date 2025-02-06
- Listed 2026-10-26
FDA record date 2025-02-06
- Listed 2026-10-26
FDA record date 2025-02-13
- Listed 2026-10-26
FDA record date 2025-02-19
- Listed 2026-10-26
FDA record date 2025-02-26
- Listed 2026-10-26
FDA record date 2025-02-27
- Listed 2026-10-26
Richard Joseph Schmidt, MD, Inc .
FDA record date 2025-03-05
- Listed 2026-10-26
FDA record date 2025-03-07
- Listed 2026-10-26
Red Star Contract Manufacturing, Inc.
FDA record date 2025-03-13
- Listed 2026-10-26
Fertility Center of Southern California
FDA record date 2025-03-14
- Listed 2026-10-26
FDA record date 2025-03-14
- Listed 2026-10-26
FDA record date 2025-03-18
- Listed 2026-10-26
FDA record date 2025-03-20
- Listed 2026-10-26
MP Biomedicals Asia Pacific Pte Ltd
FDA record date 2025-03-21
- Listed 2026-10-26
Lifesouth Community Blood Centers Inc. East Alabama Region
FDA record date 2025-03-26
- Listed 2026-10-26
Department of Defense - Air Force, 99th Medical Group
FDA record date 2025-03-27
- Listed 2026-10-26
FDA record date 2025-04-01
- Listed 2026-10-26
FDA record date 2025-04-08
- Listed 2026-10-26
FDA record date 2025-04-09
- Listed 2026-10-26
FDA record date 2025-04-11
- Listed 2026-10-26
Department of Defense - Air Force
FDA record date 2025-04-14
- Listed 2026-10-26
Department of Defense - Navy (U.S. Naval Hospital Naples)
FDA record date 2025-04-15
- Listed 2026-10-26
FDA record date 2025-04-18
- Listed 2026-10-26
Primary Children's Hospital - Miller Campus
FDA record date 2025-04-24
- Listed 2026-10-26
FDA record date 2025-04-24
- Listed 2026-10-26
Louisville Reproductive Center
FDA record date 2025-04-28
- Listed 2026-10-26
FDA record date 2025-04-29
- Listed 2026-10-26
Advanced Fertility Associates Medical Group, Inc.
FDA record date 2025-05-01
- Listed 2026-10-26
FDA record date 2025-05-02
- Listed 2026-10-26
FDA record date 2025-05-05
- Listed 2026-10-26
Labcorp Early Development Laboratories, Inc.
FDA record date 2025-05-07
- Listed 2026-10-26
FDA record date 2025-05-08
- Listed 2026-10-26
FDA record date 2025-05-14
- Listed 2026-10-26
FDA record date 2025-05-14
- Listed 2026-10-26
FDA record date 2025-05-15
- Listed 2026-10-26
Mission Hospitals Transfusion Service
FDA record date 2025-05-15
- Listed 2026-10-26
FDA record date 2025-05-16
- Listed 2026-10-26
FDA record date 2025-05-22
- Listed 2026-10-26
FDA record date 2025-05-22
- Listed 2026-10-26
FDA record date 2025-05-27
- Listed 2026-10-26
FDA record date 2025-05-27
- Listed 2026-10-26
FDA record date 2025-06-05
- Listed 2026-10-26
FDA record date 2025-06-05
- Listed 2026-10-26
FDA record date 2025-06-05
- Listed 2026-10-26
FDA record date 2025-06-06
- Listed 2026-10-26
Innovis Health, LLC d.b.a. Essentia Health
FDA record date 2025-06-11
- Listed 2026-10-26
FDA record date 2025-06-12
- Listed 2026-10-26
Performance Orthopedics Associates Inc
FDA record date 2025-06-13
- Listed 2026-10-26
Reading Scientific Services Ltd
FDA record date 2025-06-16
- Listed 2026-10-26
FDA record date 2025-06-17
- Listed 2026-10-26
FDA record date 2025-06-19
- Listed 2026-10-26
FDA record date 2025-06-20
- Listed 2026-10-26
Hendrick Regional Laboratory Transfusion Services
FDA record date 2025-07-02
- Listed 2026-10-26
FDA record date 2025-07-30
- Listed 2026-10-26
FDA record date 2025-08-18
- Listed 2026-10-26
FDA record date 2025-08-19
- Listed 2026-10-26
Qilu Pharmaceutical (Hainan) Co., Ltd.
FDA record date 2025-08-22
- Listed 2026-10-26
FDA record date 2025-09-04
- Listed 2026-10-26
FDA record date 2025-09-12
- Listed 2026-10-26
FDA record date 2025-11-05
- Listed 2026-10-26
FDA record date 2025-11-06
- Listed 2026-10-26
FDA record date 2025-11-06
- Listed 2026-10-26
FDA record date 2025-11-19
- Listed 2026-10-26
FDA record date 2025-12-11
- Listed 2026-10-26
FDA record date 2025-12-16
- Listed 2026-10-26
FDA record date 2025-12-18
- Listed 2026-10-26
Serum Institute of India Private Ltd.
FDA record date 2026-01-30
- Listed 2026-10-26
FDA record date 2026-02-09
- Listed 2026-10-26
FDA record date 2026-02-20
- Listed 2026-10-26
FDA record date 2026-02-20
- Listed 2026-10-26
FDA record date 2026-02-26
- Listed 2026-10-26
Tan-Alize Kozmetik ve Temizlik Urunleri Sanayi ve Ticaret A.S.
FDA record date 2026-02-27
- Listed 2026-10-26
FDA record date 2026-02-27
- Listed 2026-10-26
FDA record date 2026-03-02
- Listed 2026-10-26
FDA record date 2026-03-04
- Listed 2026-10-26
FDA record date 2026-03-05
- Listed 2026-10-26
North Shore University Hospital Northwell Health
FDA record date 2026-03-11
- Listed 2026-10-26
FDA record date 2026-03-17
- Listed 2026-10-26
FDA record date 2026-03-17
- Listed 2026-10-26
FDA record date 2026-03-19
- Listed 2026-10-26
FDA record date 2026-03-24
- Listed 2026-10-26
FDA record date 2026-03-26
- Listed 2026-10-26
FDA record date 2026-03-26
- Listed 2026-10-26
Fagron Analytical Services, LLC
FDA record date 2026-03-26
- Listed 2026-10-26
FDA record date 2026-03-31
- Listed 2026-10-26
FDA record date 2026-04-01
- Listed 2026-10-26
FDA record date 2026-04-03
- Listed 2026-10-26
Reproductive Sciences Medical Center, Inc.
FDA record date 2026-04-08
- Listed 2026-10-26
Scott & White Memorial Hospital
FDA record date 2026-04-13
- Listed 2026-10-26
FDA record date 2026-04-15
- Listed 2026-10-26
Banner - University Medical Center Tuscon
FDA record date 2026-04-22
- Listed 2026-10-26
Eugia Steriles Private Limited
FDA record date 2026-06-19
Recalls by class
Drug and device recalls in FDA's Enforcement Report dated this week, from openFDA, which FDA last updated Nov 11, 2026.
Class I16 recallsShow all
- D-0084-2027openFDA record
Greenwich Rx
Glutathione Injection (Glutathione-L, Ascorbic Acid, Sodium Hydroxide, Bacteriostatic Water), 200 mg/mL, 10 mL multi-dose vial, Rx only, Subcutaneous Only, Greenwich Rx.
Reason: Microbial Contamination of Sterile Products
Drug · initiated 2026-10-01 · Enforcement Report 2026-10-28 · Ongoing
- D-0082-2027openFDA record
Par Health USA, LLC
Argatroban Injection, USP, 250 mg/2.5mL (100 mg/mL), 2.5 mL x 1 Single Dose Vial, Rx only, Manufactured for: Endo USA, Malvern,PA 19355/Argatroban Injection, 250 mg/2.5mL (100 mg/mL), 2.5 mL x 1 Single Dose Vial, Rx only, Distributed by: Par Pharmaceutical, Chestnut Ridge, NY 10…
Reason: Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.
Drug · initiated 2026-10-13 · Enforcement Report 2026-10-28 · Ongoing
- D-0083-2027openFDA record
Par Health USA, LLC
Argatroban Injection, USP, 250 mg/2.5mL (100 mg/mL), 2.5 mL x 1 Single Dose Vial, Rx only, Manufactured for: Endo USA, Malvern,PA 19355/ Argatroban Injection, 250 mg/2.5mL (100 mg/mL), 2.5 mL x 1 Single Dose Vial, Rx only, Distributed by: Par Pharmaceutical, Cranbury, NJ 08512,…
Reason: Particulate matter: particulate matter identified as cellulose, polystyrene, polyethylene teraphalate, stainless steel, or stopper material in the product solution.
Drug · initiated 2026-10-13 · Enforcement Report 2026-10-28 · Ongoing
- Z-0152-2027openFDA record
Philips Ultrasound, LLC
Philips Lumify Diagnostic Ultrasound System
Reason: Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/p…
Device · initiated 2026-09-24 · Enforcement Report 2026-10-28 · Ongoing
- Z-0122-2027openFDA record
Ventec Life Systems, Inc.
VOCSN+Pro, Model Number:PRT-00490-001, Package PN:PRT-00853-000. used for ventilatory support.
Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing
- Z-0123-2027openFDA record
Ventec Life Systems, Inc.
VOCSN+Pro (Japan), Model Number: PRT-00490-100, Package PN:PRT-00853-100. used for ventilatory support.
Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing
- Z-0124-2027openFDA record
Ventec Life Systems, Inc.
VC+Pro Model Number: PRT-00900-001, Package PN:PRT-00973-000. used for ventilatory support.
Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing
- Z-0125-2027openFDA record
Ventec Life Systems, Inc.
VC+Pro (Japan) Model Number: PRT-00900-100, Package PN:PRT-00973-100. used for ventilatory support.
Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing
- Z-0126-2027openFDA record
Ventec Life Systems, Inc.
VOCSN Model Number: PRT-01100-000, Package PN:PRT-01126-000. used for ventilatory support.
Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing
- Z-0127-2027openFDA record
Ventec Life Systems, Inc.
VOCSN (Japan) Model Number: PRT-01100-100, Package PN:PRT-01126-100. used for ventilatory support.
Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing
- Z-0128-2027openFDA record
Ventec Life Systems, Inc.
VCSN+Pro Model Number: PRT-01184-000, Package PN:PRT-01172-000. used for ventilatory support.
Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing
- Z-0129-2027openFDA record
Ventec Life Systems, Inc.
V+Pro Model Number: PRT-01185-000, Package PN:PRT-01198-000; Model Number: PRT-01185-001, Package PN: N/A; Model Number: PRT-01185-002, Package PN:N/A. used for ventilatory support.
Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing
- Z-0130-2027openFDA record
Ventec Life Systems, Inc.
VC Model Number: PRT-01098-000, Package PN:PRT-01125-000. used for ventilatory support.
Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing
- Z-0131-2027openFDA record
Ventec Life Systems, Inc.
VC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support.
Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing
- Z-0132-2027openFDA record
Ventec Life Systems, Inc.
V+ Pro Emergency, Model Number: PRT-01201-000, Package PN:PRT-01173-000,PRT-01222-000. used for ventilatory support.
Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing
- Z-0133-2027openFDA record
Ventec Life Systems, Inc.
V*Home, Model Number: PRT-01444-000, Package PN:PRT-01445-000. used for ventilatory support.
Reason: Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
Device · initiated 2026-09-17 · Enforcement Report 2026-10-28 · Ongoing
Class II103 recallsShow all
- Z-0191-2027openFDA record
Abbott
Amplatzer Amulet Left Atrial Appendage Occluder Model Numbers 1. 9-ACP2-007-016 2. 9-ACP2-007-018 3. 9-ACP2-007-020 4. 9-ACP2-007-022 5. 9-ACP2-010-025 6. 9-ACP2-010-028 7. 9-ACP2-010-031 8. 9-ACP2-010-034
Reason: Changes to Instructions for Use (IFU) for the device regarding risk of Pulmonary Artery Injury (PAI).
Device · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0045-2027openFDA record
Ascend Laboratories, LLC
Amlodipine and Olmesartan Medoxomil Tablets, 10mg/40mg, 30 Tablets, Rx only, Manufactured by: Alkem Laboratories Ltd., INDIA, Distributed by: Ascend Laboratories, LLC, Parsippany, NJ 07054, NDC 67877-502-30.
Reason: Out of specification due to dissolution.
Drug · initiated 2026-10-06 · Enforcement Report 2026-10-28 · Ongoing
- Z-0282-2027openFDA record
AVID Medical, Inc.
Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH31.
Reason: A non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications.
Device · initiated 2026-09-08 · Enforcement Report 2026-10-28 · Ongoing
- Z-0283-2027openFDA record
AVID Medical, Inc.
Halyard ANTERIOR HIP PACK. Model/Catalog Number: SOAH30.
Reason: A non-sterile component was piggybacked and incorrectly placed outside of the sterile barrier and not subjected to EO sterilization per kit specifications.
Device · initiated 2026-09-08 · Enforcement Report 2026-10-28 · Ongoing
- Z-0198-2027openFDA record
Axonics, Inc.
PNE Lead Implant Kit REF: 1701
Reason: Retroactively Reported: 10mL luer lock syringes included in kit may leak , break or experience other issues that may pose a risk to users.
Device · initiated 2024-07-15 · Enforcement Report 2026-10-28 · Ongoing
- Z-0193-2027openFDA record
B Braun Medical Inc
B Braun SAFSITE Injection Site, RV1000NC SAFSITE VALVE W/O EXTRA CAP. Catalog Number: 415068.
Reason: Potential for leakage and backflow at the Safsite valve.
Device · initiated 2026-09-14 · Enforcement Report 2026-10-28 · Ongoing
- Z-0201-2027openFDA record
Beckman Coulter, Inc.
ROTOR ASSY-SW40, Reference Number 331302, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers…
Reason: Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparatio…
Device · initiated 2026-09-21 · Enforcement Report 2026-10-28 · Ongoing
- Z-0202-2027openFDA record
Beckman Coulter, Inc.
ROTOR ASSY-SW41, Reference Number 331362, when the rotors are installed in the following centrifuges: Optima XPN-100 (Model Numbers A99846 and A94469), Optima XPN-90 (Model Numbers A99842 and A94468), Optima XPN-80 (Model Numbers A99839 and A95765), Optima XE-100 (Model Numbers…
Reason: Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centri…
Device · initiated 2026-09-21 · Enforcement Report 2026-10-28 · Ongoing
- Z-0223-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/105; Material Number: M001195200
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0224-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/130; Material Number: M001195210
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0225-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/155; Material Number: M001195220
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0226-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/1RO/105; Material Number: M001195230
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0227-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/1RO/130; Material Number: M001195240
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0228-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/1RO/155; Material Number: M001195250
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0229-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/105; Material Number: M001195260
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0230-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/130; Material Number: M001195270
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0231-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/J/1RO/155; Material Number: M001195280
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0232-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/105; Material Number: M001195290
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0233-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/130; Material Number: M001195300
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0234-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/SWAN/1RO/155; Material Number: M001195310
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0235-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/2RO/130; Material Number: M001195320
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0236-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/2RO/155; Material Number: M001195330
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0237-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/2RO/130; Material Number: M001195340
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0238-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/021/BERN/2RO/155; Material Number: M001195350
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0239-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/105; Material Number: M001195400
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0240-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/130; Material Number: M001195410
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0241-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/STRAIGHT/1RO/155; Material Number: M001195420
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0242-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/105; Material Number: M001195430
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0243-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/130; Material Number: M001195440
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0244-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/BERN/1RO/155; Material Number: M001195450
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0245-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/J/1RO/130; Material Number: M001195470
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0246-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/J/1RO/155; Material Number: M001195480
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0247-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/SWAN/1RO/130; Material Number: M001195500
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0248-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/SINGLE/027/SWAN/1RO/155; Material Number: M001195510
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0249-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/130; Material Number: M001195610
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0250-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/1RO/155; Material Number: M001195620
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0251-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/1RO/130; Material Number: M001195640
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0252-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/1RO/155; Material Number: M001195650
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0253-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/STRAIGHT/2RO/130; Material Number: M001195660
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0254-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/FATHOM/021/BERN/2RO/130; Material Number: M001195670
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0255-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/FATHOM/027/STRAIGHT/1RO/130; Material Number: M001195710
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0256-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/FATHOM/027/STRAIGHT/1RO/155; Material Number: M001195720
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0257-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/FATHOM/027/BERN/1RO/130; Material Number: M001195740
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0258-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/FATHOM/027/BERN/1RO/155; Material Number: M001195750
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0259-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/TRANSEND/021/STRAIGHT/1RO/130; Material Number: M001195810
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0260-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/TRANSEND/021/BERN/1RO/130; Material Number: M001195840
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0261-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/TRANSEND/021/BERN/1RO/155; Material Number: M001195850
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0262-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/TRANSEND/027/STRAIGHT/1RO/130; Material Number: M001195910
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0263-2027openFDA record
Boston Scientific Cork, Ltd.
Direxion Torqueable Microcatheter, NGMC/TRANSEND/027/BERN/1RO/130; Material Number: M001195940
Reason: The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- Z-0204-2027openFDA record
CooperVision, Inc.
Generic Name: Softmed Breathables Advanced Toric (fanfilcon A) Soft (Hydrophilic) Contact Lens; Model/Catalog Number: Softmed Breathables Advanced Toric Product Description: 55% water content, toric, silicone hydrogel contact lens for daily wear, frequent replacement. Blister pa…
Reason: The affected products bear an incorrect statement on the carton packaging indicating that the product is "Indicated for Extended Wear". However, this product is intended for daily wear, and the extended-wear statement should not have appeared on the affected carton labeling. Thi…
Device · initiated 2026-09-22 · Enforcement Report 2026-10-28 · Ongoing
- Z-0200-2027openFDA record
Greiner Bio-One North America, Inc.
Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)
Reason: Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
Device · initiated 2026-07-29 · Enforcement Report 2026-10-28 · Ongoing
- D-0076-2027openFDA record
Hikma Pharmaceuticals USA INC.
Abirtega (abiraterone acetate) USP, 250 mg, 120-count bottles, Rx only, Rx only, Mfd for: Hikma Pharmaceuticals USA Inc., Berkley Heights, NJ 07922 USA by: Arab Pharmaceuticals Manufacturing PSC PO Box 41, Sahab-Jordon, NDC 82249-011-12.
Reason: Failed Tablet/Capsule Specifications: Failed tablet specifications for weight uniformity.
Drug · initiated 2026-10-06 · Enforcement Report 2026-10-28 · Ongoing
- Z-0211-2027openFDA record
Life Technologies Holdings Pte Ltd
Ion Torrent Genexus Dx with software, Catalog: A40003539, which supports the Ion Torrent Genexus Dx Integrated Sequencer
Reason: DNA or RNA sequencing system software has vulnerabilities.
Device · initiated 2026-08-14 · Enforcement Report 2026-10-28 · Ongoing
- Z-0269-2027openFDA record
Medline Industries, LP
MEDLINE DRAPE C-SECTION W/POUCH ST 7/CS, Medline Product Number/SKU DYNJP6101
Reason: Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Device · initiated 2026-09-09 · Enforcement Report 2026-10-28 · Ongoing
- Z-0270-2027openFDA record
Medline Industries, LP
MEDLINE DRAPE C-SEC FEN CLEAR SCRN WIRE, Medline Product Number/SKU DYNJP6106
Reason: Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Device · initiated 2026-09-09 · Enforcement Report 2026-10-28 · Ongoing
- Z-0271-2027openFDA record
Medline Industries, LP
MEDLINE DRAPE C-SECTION FEN CLR SCREEN W, Medline Product Number/SKU DYNJP6108
Reason: Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Device · initiated 2026-09-09 · Enforcement Report 2026-10-28 · Ongoing
- Z-0272-2027openFDA record
Medline Industries, LP
MEDLINE medical convenience kits labeled as: FETOSCOPY CDS-LF, Medline Product Number/SKU CDS840215P
Reason: Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Device · initiated 2026-09-09 · Enforcement Report 2026-10-28 · Ongoing
- Z-0273-2027openFDA record
Medline Industries, LP
MEDLINE medical convenience kits labeled as: C-SECTION PACK, Medline Product Number/SKU DYNJ48455A
Reason: Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Device · initiated 2026-09-09 · Enforcement Report 2026-10-28 · Ongoing
- Z-0274-2027openFDA record
Medline Industries, LP
MEDLINE medical convenience kits labeled as: 1) FETAL SURGERY CDS-LF, Medline Kit Number/SKUCDS840214R; 2) MAJOR, Medline Kit Number/SKUDYNJ904851I; 3) ABDOMINAL TRANSPLANT, Medline Kit Number/SKUDYNJ908124B.
Reason: Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Device · initiated 2026-09-09 · Enforcement Report 2026-10-28 · Ongoing
- Z-0275-2027openFDA record
Medline Industries, LP
MEDLINE medical convenience kits labeled as (list truncated)
Reason: Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Device · initiated 2026-09-09 · Enforcement Report 2026-10-28 · Ongoing
- Z-0276-2027openFDA record
Medline Industries, LP
MEDLINE medical convenience kits labeled as: 1) BIRTHDAY KIT-DOUBLE BASIN-LF, Medline Kit Number/SKU DYNJ16127N; 2) NYGH-DELIVERY PACK-LF, Medline Kit Number/SKU DYNJ27206D; 3) OB/GYN, Medline Kit Number/SKU DYNJ911652.
Reason: Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Device · initiated 2026-09-09 · Enforcement Report 2026-10-28 · Ongoing
- Z-0277-2027openFDA record
Medline Industries, LP
MEDLINE medical convenience kits labeled as: 1) LAPAROTOMY GYN CDS, Medline Kit Number/SKU CDS983529O; 2) GYN ROBOTIC PK-LF, Medline Kit Number/SKU DYNJ52432M; 3) TAH PACK, Medline Kit Number/SKU DYNJ60646A; 4) TUBAL LIGATION RFD, Medline Kit Number/SKU DYNJ906291F.
Reason: Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Device · initiated 2026-09-09 · Enforcement Report 2026-10-28 · Ongoing
- Z-0278-2027openFDA record
Medline Industries, LP
MEDLINE medical convenience kits labeled as: (list truncated)
Reason: Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Device · initiated 2026-09-09 · Enforcement Report 2026-10-28 · Ongoing
- Z-0279-2027openFDA record
Medline Industries, LP
MEDLINE medical convenience kits labeled as: TUBAL LIGATION RFD, Medline Product Number/SKU DYNJ906291G
Reason: Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Device · initiated 2026-09-09 · Enforcement Report 2026-10-28 · Ongoing
- Z-0280-2027openFDA record
Medline Industries, LP
MEDLINE medical convenience kits labeled as: C-SECTION TRAY-LF, Medline Kit Number/SKU OBN471K
Reason: Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Device · initiated 2026-09-09 · Enforcement Report 2026-10-28 · Ongoing
- Z-0281-2027openFDA record
Medline Industries, LP
MEDLINE medical convenience kits labeled as: 1) HIP SCOPE PACK, Medline Kit Number/SKU DYNJ64103C; 2) HIP SCOPE PACK, Medline Kit Number/SKU DYNJ64155B.
Reason: Reports of problems with product being difficult to remove from patient and leaving adhesive residue on patient's skin.
Device · initiated 2026-09-09 · Enforcement Report 2026-10-28 · Ongoing
- Z-0203-2027openFDA record
Medtronic MiniMed, Inc.
CareLink Clinic, REF: MMT-7350
Reason: CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ra…
Device · initiated 2026-09-11 · Enforcement Report 2026-10-28 · Ongoing
- Z-0265-2027openFDA record
Medtronic MiniMed, Inc.
Generic Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 (iOS), MMT-8601 (Android) Software Version: Software version 1.2.1 Product Description: The MiniMed Go app for iOS or MiniMed Go app for Android is compatible with the following components:…
Reason: A software defect in which existing users who had previously configured and enabled the long-acting dose reminder feature in version 1.2.0 or earlier experienced failure of the long-acting dose reminder after updating their app to version 1.2.1. The long-acting dose reminder is…
Device · initiated 2026-09-02 · Enforcement Report 2026-10-28 · Ongoing
- Z-0194-2027openFDA record
Medtronic Navigation, Inc.
Medtronic NDI Passive Spheres REF: 8801071, Sterile EO, 1/Tray 12 pk., Rxonly
Reason: Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.
Device · initiated 2026-08-31 · Enforcement Report 2026-10-28 · Ongoing
- Z-0195-2027openFDA record
Medtronic Navigation, Inc.
Medtronic NDI Passive Spheres REF: 8801074, Sterile EO, 4/Tray 12 Pk., Rxonly
Reason: Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.
Device · initiated 2026-08-31 · Enforcement Report 2026-10-28 · Ongoing
- Z-0196-2027openFDA record
Medtronic Navigation, Inc.
Medtronic NDI Passive Spheres REF: 8801075, Sterile EO, 5/Tray 12 Pk., Rxonly
Reason: Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.
Device · initiated 2026-08-31 · Enforcement Report 2026-10-28 · Ongoing
- Z-0197-2027openFDA record
Medtronic Navigation, Inc.
Medtronic Biopsy Needle Kit, REF: 9733068, Passive, Sterile EO, Rxonly
Reason: Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.
Device · initiated 2026-08-31 · Enforcement Report 2026-10-28 · Ongoing
- Z-0266-2027openFDA record
Medtronic Neuromodulation
InterStim SureScan MRI Leads 978A1 & 978B1: LEAD 978A128 ISTM MICRO 28CM SSMRI TINED 978A128 LEAD 978A133 ISTM MICRO 33CM SSMRI TINED 978A133 LEAD 978A141 ISTM MICRO 41CM SSMRI TINED 978A141 LEAD 978B128 ISTM II 28CM SSMRI TINED 978B128 LEAD 978B133 ISTM II 33CM SSMRI TINED 978B…
Reason: Wrong lead or percutaneous extension selected. Use of an incompatible lead or percutaneous extension will prevent the system from working as intended and may result in delay of surgery or need for surgical replacement of the product.
Device · initiated 2026-09-08 · Enforcement Report 2026-10-28 · Ongoing
- Z-0267-2027openFDA record
Medtronic Neuromodulation
InterStim Percutaneous Extensions EXT 3560022 ISTM II PERC 3560022 EXT 3560030 ISTM MICRO PERC 3560030
Reason: Wrong lead or percutaneous extension selected. Use of an incompatible lead or percutaneous extension will prevent the system from working as intended and may result in delay of surgery or need for surgical replacement of the product.
Device · initiated 2026-09-08 · Enforcement Report 2026-10-28 · Ongoing
- D-0051-2027openFDA record
Pfizer
Levoxyl (levothyroxine sodium tablets, USP), tablets, 175 mcg, 1000-count bottle, Rx only, Distributed by Pfizer Inc., New York, NY 10017, Made in Austria. NDC 60793-859-10
Reason: Subpotent Drug: confirmed out-of-specification (OOS) assay during stability testing.
Drug · initiated 2026-09-29 · Enforcement Report 2026-10-28 · Ongoing
- D-0052-2027openFDA record
Pfizer
Levoxyl (levothyroxine sodium tablets, USP), tablets, 175 mcg, 100 tablet bottle, Rx only, Distributed by Pfizer Inc., New York, NY 10017, Made in Austria. NDC 60793-859-01.
Reason: Subpotent Drug: confirmed out-of-specification (OOS) assay during stability testing.
Drug · initiated 2026-09-29 · Enforcement Report 2026-10-28 · Ongoing
- Z-0199-2027openFDA record
Philips North America
Philips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;
Reason: Unreliable wireless roaming behavior of Smart-hopping 1.0 wireless clients during transport when connecting to Smart-hopping 2.0 infrastructure, resulting in loss of connectivity to the central monitoring system (PIC iX).
Device · initiated 2026-09-16 · Enforcement Report 2026-10-28 · Ongoing
- D-0053-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
Brush Buddies, Barbie, Anticavity Fluoride Toothpaste, (sodium fluoride 0.243%) Bubble Gum Flavor, Net Wt. 4.0 oz (113g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-001-40, Item# 88022, UPC 8 10147 88022 5
Reason: CGMP Deviations
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0054-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
Brush Buddies, Hot Wheels, Anticavity Fluoride Toothpaste, (sodium fluoride 0.243%) Strawberry Flavor, Net Wt. 4.0 oz (113g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-002-40, Item # 88047, UPC 8 10147 88047 8
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0055-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
Brush Buddies, Crayola, Silly Scents, Anticavity Fluoride Toothpaste, (sodium fluoride 0.243%) Bubble Gum Flavor, Net Wt. 4.0 oz (113g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-003-40, Item # 32731, UPC 8 50048 32731 3
Reason: CGMP Deviations
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0056-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
Brush Buddies, Crayola, Silly Scents, Anticavity Fluoride Toothpaste, (sodium fluoride0.243%) Watermelon Flavor, Net Wt. 4.0 oz (113g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-004-40, Item # Product code: 32730, UPC 8 50048 32730 6.
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0057-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
Brush Buddies, Crayola Silly Scents, Anticavity Fluoride Toothpaste, (sodium fluoride 0.243%) Strawberry Flavor, Net Wt. 4.0 oz (113g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-005-40, Item # 32732, UPC 8 50048 32732 0
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0058-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
Brush Buddies, My Little Pony Scents Anticavity Fluoride Toothpaste, (sodium fluoride 0.243%) Bubble Gum Flavor, Net Wt. 4.0 oz (113g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-008-40, Item# 88270, UPC 8 10147 88270 0.
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0059-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
evo, Advanced Whitening, Anticavity Fluoride Toothpaste, (sodium fluoride 0.24%) Peppermint, Net Wt. 3.5 oz (99g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-886-19, Item# 88619, UPC 8 10147 88619 7
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0060-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
evo, Sensitive, Anticavity Fluoride Toothpaste, (potassium nitrate 5%, sodium fluoride 0.24%) Sweet Mint, Net Wt. 3.5 oz (99g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-620-35, Item # 88620 UPC8 10147 88620 3
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0061-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
Brush Buddies, Vanilla Ice Cream Flavored Toothpaste, Anticavity Fluoride Toothpaste, (sodium fluoride 0.243%) Net Wt. 3.5 oz (99g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-517-35, Item # 88517, UPC 8 10147 88517 6
Reason: CGMP Deviations
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0062-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
Brush Buddies, Mint Chocolate Chip Ice Cream Flavored Toothpaste, Anticavity Fluoride Toothpaste, (sodium fluoride 0.243%) Net Wt. 3.5 oz (99g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-520-35, Item # 88520, UPC 8 10147 88520 6
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0063-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
Brush Buddies, Cookies 'n Cream Ice Cream Flavored Toothpaste, Anticavity Fluoride Toothpaste, (sodium fluoride 0.243%) Net Wt. 3.5 oz (99g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-519-35, Item # 88519, UPC 8 10147 88519 0
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0064-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
Brush Buddies, Chocolate Ice Cream Flavored Toothpaste, Anticavity Fluoride Toothpaste, (sodium fluoride 0.243%) Net Wt. 3.5 oz (99g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-518-35, Item # 88518, UPC 8 10147 88518 3
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0065-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
Brush Buddies, Whitening Anti-cavity Fluoride Toothpaste, with Baking Soda, (sodium monofluorophosphate 0.76%) Cool Mint, Net Wt. 3.5 oz (99g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-007-07, ITEM # 999131, 8 50010 84249 3
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0066-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
BB (Brush Buddies) Cavity Protection, Anti-Cavity Fluoride Toothpaste, (sodium monofluorophosphate 0.76%) Cool Mint, 2 Value Pack, Two 3.5 oz (99g) tubes, Total Net Wt 7 oz (198g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-035-10, Item # 88543, UPC…
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0067-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
BB (Brush Buddies) Advanced Whitening, Anti-Cavity Fluoride Toothpaste, (sodium fluoride 0,24%) Peppermint, Net Wt 3.5 oz (99g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-885-42, Item # 88542, UPC 8 10147 88542 8
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0068-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
ARU, Toothpaste, Gum Health, 9 Ingredients, Soothing Mint, (stannous fluoride 0.45%) Net Wt 3.5 oz (99g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-786-35, Item # 88786 UPC 8 10147 88786 6
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- D-0069-2027openFDA record
PRITAM INTERNATIONAL PRIVATE LIMITED
ARU, Toothpaste, Cavity Protection, 9 Ingredients, Cool Mint, (sodium fluoride 0.243%) Net Wt 3.5 oz (99g), Distributed by Ashtel Studios Inc., Ontario, California 91761, NDC 84140-785-35, Item # 88785, UPC 8 10147 88785
Reason: CGMP Deviations.
Drug · initiated 2026-09-10 · Enforcement Report 2026-10-28 · Ongoing
- Z-0264-2027openFDA record
restor3d Inc.
Generic Name: restor3d MTP Implant Generic Name: MTP Implant Model/Catalog Number: 1103-161620 Software Version: None Product Description: The restor3d Metatarsophalangeal (MTP) Implant is a sterile, single-use implant grade metal device, available in varied articulating surface…
Reason: Implants were found in the field with orange sterilzation indicators which means implants were not exposed to sterilization.
Device · initiated 2026-09-04 · Enforcement Report 2026-10-28 · Ongoing
- D-0070-2027openFDA record
Right Value Drug Stores LLC
Testosterone Cypionate, 200 mg/mL, packaged in a) 5 mL vials (NDC 73271-1225-1) and b) 30 m vials ( NDC 73271-1226-2) This Is A Compounded Dug, Intramuscular Use, Office Use Only, 8400 Esters Blvd, STE 190, Irving, TX 75063.
Reason: Lack of sterility assurance.
Drug · initiated 2026-10-09 · Enforcement Report 2026-10-28 · Ongoing
- D-0050-2027openFDA record
Shenzhen Techdow Pharmaceutical Co., Ltd.
Heparin Sodium Injection, USP, 5,000 USP Units/mL, 1mL Multi dose vial x 25 vials per carton, Distributed by: Hepalink USA Inc., Langhome, PA 19047-1885. Carton NDC: NDC 81952-111-06, Vial NDC: 81942-111-01
Reason: Lack of Assurance of Sterility
Drug · initiated 2026-10-01 · Enforcement Report 2026-10-28 · Ongoing
- Z-0268-2027openFDA record
Solstice Corporation
Xact mole marker for mammography Model/Catalog Number: M15RM
Reason: The radiopaque material may detach from the marker. In some cases, artifact was noted and attributed to the detached material.
Device · initiated 2026-09-22 · Enforcement Report 2026-10-28 · Ongoing
- D-0079-2027openFDA record
STRIDES PHARMA INC
Cinacalcet Tablets, 30 mg, 30-count bottle, Rx only, Manufactured by: Strides Pharma Science Ltd. Puducherry, 605 014, India; Distributed by: Strides Pharma Inc. Bridgewater, NJ 08807, NDC 64380-883-04
Reason: Failed dissolution specifications
Drug · initiated 2026-10-07 · Enforcement Report 2026-10-28 · Ongoing
- Z-0205-2027openFDA record
TECHNO-PATH MANUFACTURING LTD.
Techno-Path Multichem AE Unassayed Level 1 REF:AE621X
Reason: A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supp…
Device · initiated 2026-08-14 · Enforcement Report 2026-10-28 · Ongoing
- Z-0206-2027openFDA record
TECHNO-PATH MANUFACTURING LTD.
Techno-Path Multichem AE Unassayed Level 2 REF: AE622X
Reason: A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supp…
Device · initiated 2026-08-14 · Enforcement Report 2026-10-28 · Ongoing
- Z-0207-2027openFDA record
TECHNO-PATH MANUFACTURING LTD.
Techno-Path Multichem AE Unassayed Level 3, REF:AE623X
Reason: A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supp…
Device · initiated 2026-08-14 · Enforcement Report 2026-10-28 · Ongoing
- Z-0192-2027openFDA record
Think Surgical, Inc.
Think Surgical Square Driver REF: 109126
Reason: Robotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin wobbling during drilling.
Device · initiated 2026-08-20 · Enforcement Report 2026-10-28 · Ongoing
Class III1 recallShow all
- D-0049-2027openFDA record
Viatris, Inc.
Telmisartan and Amlodipine, Tablets, USP, 40mg/5 mg, 30-count Bottle, Rx only, Manufactured for: Mylan Pharmaceuticals Inc., Morgantown, WV 26505 USA, Made in India, NDC 0378-1075-93.
Reason: Superpotent Drug: potential for the Telmisartan content in these lots to exceed specifications
Drug · initiated 2026-10-01 · Enforcement Report 2026-10-28 · Ongoing
New OAI classifications
1 inspection classified Official Action Indicated that FDAgent first recorded from FDA's inspection classification data this week, after its reads of that data in the 8 days before had not found them. Older inspections FDAgent adds to its archive are left out. Dates are inspection end dates. Inspection classifications last synced Nov 22, 2026, 04:55 UTC.
- Inspection ended 2026-08-19FEI 3029514624
Pharmalogic New York City, LLCOAI
Bronx, New York, United States · Drugs · Drug Quality Assurance · recorded 2026-10-26
Get this page by email every Monday
A free weekly email with each week’s warning letters, new Form 483s, OAI classifications and Class I recalls, counted the same way as this page. Just your email address.
About this data
- Weeks are ISO weeks, Monday to Sunday, in UTC. Each section is dated by its own FDA date: posting date for warning letters, reading-room publish date for Form 483s (each inspection's 483 counts once, when FDA first listed it; FDA's re-posts and amended 483s with the FEI and record date of one it listed before are shown apart), Enforcement Report date for recalls, and the date published FDA prints beside each firm on an import alert. FDA publishes no date for an inspection classification, so an OAI classification counts as new on the day FDAgent first recorded the inspection, and only when FDAgent's complete reads of FDA's inspection data in the 8 days before had covered the inspection's end date without finding it. Older inspections FDAgent adds to its archive, by a backfill or a wider read, never count as new, however late FDA posted them. Inspections FDAgent recorded before it began keeping a record of its reads, in November 2026, count only when they ended in the 13 weeks before, and the Veterinary inspections of FDAgent's first read of them, that same month, never count.
- Import alert entries come from FDA's import alert pages, where FDA lists firms by name and address, not by FEI. FDAgent links an entry to a facility only when FDA printed the firm's FEI, never by name. Earlier weeks list only entries still on FDA's pages, because FDA deletes an entry when it removes a firm, and FDAgent keeps removed entries only from the day it began recording them. FDAgent lists an entry here only when its name shows it is a firm, through a legal form such as Ltd, Inc or GmbH, a business word such as Laboratories or Trading, or a web address, or when FDA's inspection or drug registration records hold a firm under exactly that name; entries that may name a person are shown only to signed-in users.
- Import refusals are not listed by week: FDA reports a refusal weeks after it happens, so a week's count would keep changing after the week closed. Each facility page lists its own refusals.
- Warning letters are those issued by FDA's drug, biologics and device centers and their field and quality offices. Letters from FDA's food, tobacco, veterinary, cosmetics and dietary supplement offices are left out.
- FDA posts only some Form 483s to its Electronic Reading Room, so a missing 483 does not mean none was issued.
- FDA publishes an inspection's classification once it is final and updates its inspection data weekly. Source: FDA inspections FAQ.
- Recalls come from openFDA, which republishes FDA's weekly Enforcement Report; a report's recalls appear here once openFDA publishes it.
- FDAgent is independent of FDA. Every row links to its FDA source; check it before relying on a record.