AGC Biologics, Inc. Form 483, April 2025
FDA issued this Form 483 to AGC Biologics, Inc. in Bothell, Washington (FEI 3000209996) at the end of a biologics inspection that ended April 18, 2025. It lists 2 observations. FDA classified the inspection Voluntary Action Indicated (VAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Oct 26, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Apr 18, 2025
- Biologics · Blood and Blood Products
- FDA's outcome
- VAI
- Voluntary Action Indicated
- Posted by FDA
- Oct 26, 2026
- 556 days after the inspection
The 2 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Inadequate review of laboratory data
Review of laboratory data for drug substance in-process and release testing is not adequate: [illegible] does not match a) We observed raw data from the analytical packet for lot #[illegible] used to document the [illegible] final results. This data is generated as part of Final Bulk Drug Substance release testing for residual [illegible] The analysis packet was reviewed by a QC reviewer on 19 NOV 2024 and did not catch the error. b) On 15 APR 2025, your quality control associate (analyst) who performs the [illegible] testing method for final bulk drug substance, demonstrated (with typical system access rights) the ability to delete original electronic data from the [illegible] used for test method, among others). There is no routine system-level review to ensure the veracity of electronic data files submitted. c) On 8 Jun 2016, instrument for in-process monitoring of [illegible] and in Building [illegible] were qualified by testing reference [illegible] at various concentrations. Along with deviations noted in the protocol, the final report was reviewed and approved in April 2024. [illegible] used in the qualification is the [illegible] which was evaluated by [illegible] to show average results of [illegible] The qualification of the instrument [illegible] are still in use today for monitoring bio-reactor cell viability including for [illegible] production. [illegible] has no acceptance criteria for accuracy.
- 2
Inaccurate working cell bank inventory
Documentation of critical material inventory is not accurate. On 14 APR 2025, we observed the inventory system [illegible] indicating [illegible] vials of working cell bank (WCB lot #[illegible]) located within liquid nitrogen storage vessel #[illegible] at Building [illegible] records of the same lot show conflicting inventory of [illegible] vials. Upon observation of the storage location, only [illegible] vials were found in cryo-storage. This working cell bank is the starting material for upstream culture that will produce the drug substance under BLA [illegible] As stated by your Senior Director, Quality Systems and Compliance, no established procedures exist for conducting inventory/cycle counts of WCB stored in the release and quarantine liquid storage vessels (#[illegible] and #[illegible] respectively) in Building [illegible]
Other Form 483s FDA posted for this facility
Every Form 483 FDA posts under FEI 3000209996, newest first.
- Form 483, February 2023Inspection ended Feb 10, 2023 · Posted Jan 31, 2024VAI
- Form 483, April 2016FDA record date Apr 18, 2016 · Posted Oct 20, 2026
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.