FDA issued this Form 483 to Implantica CE Reflux Limited in San Gwann, Malta (FEI 3015217460) at the end of a device inspection that ended October 24, 2025. It lists 4 observations. FDA classified the inspection No Action Indicated (NAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 27, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Oct 24, 2025
Devices · Bioresearch Monitoring
FDA's outcome
NAI
No Action Indicated
Posted by FDA
Oct 27, 2026
368 days after the inspection
Facility
FEI 3015217460
San Gwann, Malta

The 4 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Inadequate CAPA procedures and documentation

    Procedures for corrective and preventive action have not been adequately established. Specifically, the following deficiencies were found in these CAPA records: CAPA 00001: Inadequate Corrective Actions (ineffective training) and Effectiveness Check Plan (no interim checks) CAPA 00003: Inadequate Effectiveness verification plan for systemic QMS issue (no interim checks) CAPA 00002: 1. Investigation - Missing Process-Parameter Documentation The investigation does not contain any documented evidence that the (b) (4) tumbler was the equipment actually used throughout production. There is no record showing that the speed, rpm, and other parameters employed on the RS-300 devices match the values specified in the original validation reports (IQ/OQ/PQ). 2. Corrective Action - Incomplete Batch-Record Amendment The corrective-action portion of the CAPA fails to provide documented proof that all affected batch records, including previously released batch records (job-router forms, batch-release sheets, and any downstream documentation) have been updated to include the tumbler model-number (e.g., (b) (4) as the equipment used for the tumbling operation. 3. Scope & Preventive Action - Insufficient Coverage The CAPA scope is limited to the single subcontractor (b) (4) and the one batch in question. The scope should be expanded to verify that the same omission (absence of equipment identification) does not exist in any other batch-record packages, whether they involve the same tumbler or other subcontractors, and appropriate preventive actions should be instituted to prevent recurrence.

  2. 2

    Inadequate sampling procedures for validation activities

    Procedures to ensure that sampling methods are adequate for their intended use have not been adequately established. Specifically, the firm's procedures for establishing sample sizes in validation activities lack clear guidance on when an external standard (such as an ANSI/ISO sampling requirement) can override or replace the organization's own internal criteria. The procedures do not specify the necessary elements such as acceptance criteria, justification, rationale, approval process, methodology, or how to resolve any conflict between the two sources. Consequently, the choice of sample size in validation studies may be unsupported by statistical justification and applied inconsistently. For example: A sample size of six was initially selected in Protocol (DT_DVP0007 01) Manual Cleaning Validation Plan on the suggestion as per ANSI/AAMI ST98: 2022 [9]. ANSI/AAMI ST98: 2022. This sample size was changed to 9, a more conservative sample size, as reflected in the validation report DT_DVR0007 03 Manual Cleaning Validation Report. Collectively, these validation reports did not adequately document the following: • the factors considered when selecting the initial six-unit size (expert judgment, industry practice, risk assessment, etc.). • the rationale or explanation why a sample size of 9 was sufficient, even if recommended by the standard that could include a clear, risk-based rationale for the sample size used, is sufficient for the specific device, process and risk level, and explain the deviation from your own internal SOP.

  3. 3

    Inadequate organizational structure for quality system

    21 CFR 820

    The organizational structure has not been adequately established and maintained to ensure that devices are manufactured in accordance with 21 CFR Part 820. Specifically, between approximately December 2024 and September 2025, the firm's approach to managing the quality system was, in part, triage-oriented rather than maintaining normal quality system operations. The departure of several key staff members resulted in delays and partial interruptions in the processing of CAPAs, change requests, nonconformances, and quality plans.

  4. 4

    Process validation documentation inaccurate and incongruent

    Process validation activities have not been adequately documented. Specifically, the stated purpose of the validation protocol (test instructions) RS DVP0004 "RefluxStop Radiopacity and Barium Sulphate," was not congruent with the conclusions in the corresponding output report RS DVR0004. The protocol documentation was inaccurate.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.