Implantica CE Reflux Ltd
Sir Temi Zammit Bldgs, 0 Malta Life Sciences Park, Level, San Gwann, Malta
FEI 3015217460
FDA has one inspection on file for Implantica CE Reflux Ltd in San Gwann, Malta (FEI 3015217460). It ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA posted the Form 483 from that inspection. FDAgent read its 4 observations from the PDF; they are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 1, in 2025
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 1 request, latest Apr 2, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringPosted 483 · 4 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Inadequate CAPA procedures and documentation
Procedures for corrective and preventive action have not been adequately established. Specifically, the following deficiencies were found in these CAPA records: CAPA 00001: Inadequate Corrective Actions (ineffective training) and Effectiveness Check Plan (no interim checks) CAPA 00003: Inadequate Effectiveness verification plan for systemic QMS issue (no interim checks) CAPA 00002: 1. Investigation - Missing Process-Parameter Documentation The investigation does not contain any documented evidence that the (b) (4) tumbler was the equipment actually used throughout production. There is no record showing that the speed, rpm, and other parameters employed on the RS-300 devices match the values specified in the original validation reports (IQ/OQ/PQ). 2. Corrective Action - Incomplete Batch-Record Amendment The corrective-action portion of the CAPA fails to provide documented proof that all affected batch records, including previously released batch records (job-router forms, batch-release sheets, and any downstream documentation) have been updated to include the tumbler model-number (e.g., (b) (4) as the equipment used for the tumbling operation. 3. Scope & Preventive Action - Insufficient Coverage The CAPA scope is limited to the single subcontractor (b) (4) and the one batch in question. The scope should be expanded to verify that the same omission (absence of equipment identification) does not exist in any other batch-record packages, whether they involve the same tumbler or other subcontractors, and appropriate preventive actions should be instituted to prevent recurrence.
- Observation 2
Inadequate sampling procedures for validation activities
Procedures to ensure that sampling methods are adequate for their intended use have not been adequately established. Specifically, the firm's procedures for establishing sample sizes in validation activities lack clear guidance on when an external standard (such as an ANSI/ISO sampling requirement) can override or replace the organization's own internal criteria. The procedures do not specify the necessary elements such as acceptance criteria, justification, rationale, approval process, methodology, or how to resolve any conflict between the two sources. Consequently, the choice of sample size in validation studies may be unsupported by statistical justification and applied inconsistently. For example: A sample size of six was initially selected in Protocol (DT_DVP0007 01) Manual Cleaning Validation Plan on the suggestion as per ANSI/AAMI ST98: 2022 [9]. ANSI/AAMI ST98: 2022. This sample size was changed to 9, a more conservative sample size, as reflected in the validation report DT_DVR0007 03 Manual Cleaning Validation Report. Collectively, these validation reports did not adequately document the following: • the factors considered when selecting the initial six-unit size (expert judgment, industry practice, risk assessment, etc.). • the rationale or explanation why a sample size of 9 was sufficient, even if recommended by the standard that could include a clear, risk-based rationale for the sample size used, is sufficient for the specific device, process and risk level, and explain the deviation from your own internal SOP.
Show observations 3 to 4
- Observation 321 CFR 820
Inadequate organizational structure for quality system
The organizational structure has not been adequately established and maintained to ensure that devices are manufactured in accordance with 21 CFR Part 820. Specifically, between approximately December 2024 and September 2025, the firm's approach to managing the quality system was, in part, triage-oriented rather than maintaining normal quality system operations. The departure of several key staff members resulted in delays and partial interruptions in the processing of CAPAs, change requests, nonconformances, and quality plans.
- Observation 4
Process validation documentation inaccurate and incongruent
Process validation activities have not been adequately documented. Specifically, the stated purpose of the validation protocol (test instructions) RS DVP0004 "RefluxStop Radiopacity and Barium Sulphate," was not congruent with the conclusions in the corresponding output report RS DVR0004. The protocol documentation was inaccurate.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Apr 2, 2026. It asked for a Form 483.
Inspection date not given
1 request. FDA released records in answer to a request for its Form 483 on Jun 22, 2026. FDAgent holds no copy that FDA posted.
- Apr 2, 2026FDA-FOIA-2026-3324Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3015217460 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3015217460. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.