Skip to content

Miltenyi Biotec Gmbh Form 483, May 2015

FDA issued this Form 483 to Miltenyi Biotec Gmbh in Bergisch Gladbach, Germany (FEI 3004112148) at the end of a biologics inspection that ended May 29, 2015. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 26, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
May 29, 2015
Biologics · Blood and Blood Products
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Oct 26, 2026
4,168 days after the inspection
Facility
FEI 3004112148
Bergisch Gladbach, Germany

FDA's 4 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Lack of System for Event Evaluations

    The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.

  2. Records of complaint investigation

    Records of complaint investigations do not include required information.

  3. Maintained

    Records of complaint investigations are not maintained by the formally designated unit.

  4. Lack of or inadequate complaint procedures

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.