Miltenyi Biotec Gmbh
Friedrich-Ebert-Str. 68, Bergisch Gladbach, Germany
FEI 3004112148
FDA has 2 inspections on file for Miltenyi Biotec Gmbh in Bergisch Gladbach, Germany (FEI 3004112148), from 2011 to 2015. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA posted the Form 483 from that inspection. FDA published 4 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
3 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2011–2015
- 0 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 1 of 2
- Form 483s posted by FDA
- ↗
- Warning letters
- None on file
- Import refusals
- 3 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsPosted 483 · 4 citations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.
Records of complaint investigation
Records of complaint investigations do not include required information.
Maintained
Records of complaint investigations are not maintained by the formally designated unit.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood ProductsCitations not published
Regulations FDA cited here
The 2 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 820.198Complaint files1 of 1
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3004112148, one row per refused product line.
3 refusals from 2017-06-29 to 2024-10-31 · charged UNAPPROVED on 2, FRNMFGREG on 1 · no FDA sample analysis
FDA firm profile- 2024-10-31
BLOOD SPECIMEN COLLECTION KIT (EXCLUDES HIV TESTING)73OFZ
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Northeast ImportsSCS-6727181-1/11/1
- 2022-11-22
IN-VITRO DIAG PRODUCTS, N.E.C.57VH99
UNAPPROVED 505(a), 801(a)(3); UNAPPROVED NEW DRUG
Division of Northeast ImportsDSV-8579854-1/121/1
- 2017-06-29
REVLIMID61CIY03
UNAPPROVED 505(a), 801(a)(3); UNAPPROVED NEW DRUG
Division of Northern Border ImportsUPS-3704524-3/12/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004112148 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004112148. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.