21 CFR 820.198: Complaint files

FDA's inspection data lists 5,617 Form 483 citations of 21 CFR 820.198 at 3,866 facilities, from October 2008 to July 14, 2026. 82 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
5,617
At 5,020 inspections
Facilities cited
3,866
Distinct FEIs
Most recent citation
Jul 14, 2026
Bionix, Llc
Warning letters
82
State a violation of it

The regulation

Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).

§ 820.198 Complaint files.

(a) Each manufacturer shall maintain complaint files. Each manufacturer shall establish and maintain procedures for receiving, reviewing, and evaluating complaints by a formally designated unit. Such procedures shall ensure that:

(1) All complaints are processed in a uniform and timely manner;

(2) Oral complaints are documented upon receipt; and

(3) Complaints are evaluated to determine whether the complaint represents an event which is required to be reported to FDA under part 803 of this chapter, Medical Device Reporting.

(b) Each manufacturer shall review and evaluate all complaints to determine whether an investigation is necessary. When no investigation is made, the manufacturer shall maintain a record that includes the reason no investigation was made and the name of the individual responsible for the decision not to investigate.

(c) Any complaint involving the possible failure of a device, labeling, or packaging to meet any of its specifications shall be reviewed, evaluated, and investigated, unless such investigation has already been performed for a similar complaint and another investigation is not necessary.

(d) Any complaint that represents an event which must be reported to FDA under part 803 of this chapter shall be promptly reviewed, evaluated, and investigated by a designated individual(s) and shall be maintained in a separate portion of the complaint files or otherwise clearly identified. In addition to the information required by § 820.198(e), records of investigation under this paragraph shall include a determination of:

(1) Whether the device failed to meet specifications;

(2) Whether the device was being used for treatment or diagnosis; and

(3) The relationship, if any, of the device to the reported incident or adverse event.

(e) When an investigation is made under this section, a record of the investigation shall be maintained by the formally designated unit identified in paragraph (a) of this section. The record of investigation shall include:

(1) The name of the device;

(2) The date the complaint was received;

(3) Any unique device identifier (UDI) or universal product code (UPC), and any other device identification(s) and control number(s) used;

(4) The name, address, and phone number of the complainant;

(5) The nature and details of the complaint;

(6) The dates and results of the investigation;

(7) Any corrective action taken; and

(8) Any reply to the complainant.

(f) When the manufacturer's formally designated complaint unit is located at a site separate from the manufacturing establishment, the investigated complaint(s) and the record(s) of investigation shall be reasonably accessible to the manufacturing establishment.

(g) If a manufacturer's formally designated complaint unit is located outside of the United States, records required by this section shall be reasonably accessible in the United States at either:

(1) A location in the United States where the manufacturer's records are regularly kept; or

(2) The location of the initial distributor.

21 CFR Part 820: Quality Management System Regulation, Subpart MRead on eCFR

Citations by year

Form 483 citations of 21 CFR 820.198 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 506 in 2012

Form 483 citations of 21 CFR 820.198 by inspection year
YearForm 483 citations
200845
2009341
2010416
2011472
2012506
2013409
2014432
2015397
2016343
2017390
2018341
2019334
202092
2021125
2022216
2023247
2024224
2025273
202614

How FDA words this citation

FDA's short descriptions and citation text on these 5,617 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Lack of or inadequate complaint procedures3,094 citations
  • Investigation of device failures733 citations
  • Complaints561 citations
  • Records of complaint investigation298 citations
  • General211 citations
Show 3 more
  • Review and evaluation for investigation180 citations
  • Evaluation, timeliness, identification106 citations
  • Rationale documented for no investigation83 citations

Citation text

  • “Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.”2,363 citations
  • “Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.”730 citations
  • “Complaint files are not adequately maintained.”428 citations
  • “Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.”326 citations
Show 2 more
  • “Records of complaint investigations do not include required information.”298 citations
  • “Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated and investigated where necessary.”119 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 820.198, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

  • Bionix, Llc2026-07-14 · FEI 1526854Records of MDR Investigation
  • INSPIREMD Inc2026-06-18 · FEI 3032814119Evaluation, timeliness, identification
  • Marson Medical, Inc.2026-01-16 · FEI 1000550978Lack of or inadequate complaint procedures
  • Apex Medical Technologies, Inc.2026-01-15 · FEI 2025773Lack of or inadequate complaint procedures
  • MIDWAY MEDICAL2026-01-15 · FEI 3015536408Lack of or inadequate complaint procedures
  • Expanse Medical Inc.2026-01-15 · FEI 3017823876Lack of or inadequate complaint procedures
Show 4 more
  • Conformis Inc.2026-01-13 · FEI 3004153240Lack of or inadequate complaint procedures
  • Nuclein LLC2026-01-13 · FEI 3014515702Investigation of device failures
  • Innova Vascular Inc2026-01-12 · FEI 3020256033Lack of or inadequate complaint procedures
  • ZIIP INC2026-01-12 · FEI 3029740697Lack of or inadequate complaint procedures

What the observations say

Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 820.198 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 4 observations tie this way. The highlighted sentence is the matched wording.

  • Foremount Enterprise Co., Ltd.2025-06-19 · Observation 2

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established. Specifically, your firm is a Manufacturer of Class II Non-Rebreathing Valves, Product Code CBP and Emergency Manual (Resuscitator) Ventilators, Product Code BTM, medical devices exported and intended for the US market, and your firm's Customer Complaint Handling Procedure, QP824, V.05, dated May 27, 2025, has no requirement to …

    This observation opens with FDA's standard wording for 21 CFR 820.198(a).

  • Philips Medical Systems Nederland B.V.2025-01-17 · Observation 3

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established. Specifically, A. Review of data downloaded from your (b)(4) reporting IntelliSpace Cardiovascular malfunctions (with no deaths or serious injuries reported) were identified as not reportable as Medical Device Reports ("No" entered under "Potentially Reportable Event"); however, there is no valid justification …

    This observation opens with FDA's standard wording for 21 CFR 820.198(a).

  • Argon Medical Productions & Vertrieb Gesellschaft MBH & CO2024-07-31 · Observation 4

    Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary. Specifically, complaints of failure of dental implants to osseointegrate are not investigated, for

    This observation opens with FDA's standard wording for 21 CFR 820.198(c).

Show 1 more
  • Legacy Pharmaceutical Packaging LLC2020-01-16 · Observation 5

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established. Specifically, Legacy does not have complaint procedures that ensure complaints are evaluated for MDR reportability.

    This observation opens with FDA's standard wording for 21 CFR 820.198(a).

Warning letters stating a violation of 21 CFR 820.198

82 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.

“Failure to review, evaluate and investigate any complaint involving the possible failure of a device, labeling or packaging to meet any of its specifications, as required by 21 CFR 820.198(c).”
“Failure to establish and maintain procedures for receiving, reviewing, and evaluating complaints by a formally designated unit, as required by 21 CFR 820.198(a).”
“Failure to maintain complaint files and establish and maintain procedures for receiving, reviewing, and evaluating complaints by a formally designated unit, as required by 21 CFR 820.198(a).”
“Failure to adequately establish and maintain procedures for receiving, reviewing, and evaluating complaints by a formally designated unit, as required by 21 CFR 820.198(a).”

Each letter below states the violation in these words.

“Failure to establish and maintain procedures for receiving, reviewing, and evaluating complaints by a formally designated unit as required by 21 CFR 820.198(a).”

Often cited at the same inspection

Of the 5,020 inspections where FDA cited 21 CFR 820.198, the share where it also cited each of these sections.

  • 21 CFR 820.100Corrective and preventive action2,869 (57%)
  • 21 CFR 820.30Design controls1,801 (36%)
  • 21 CFR 820.50Purchasing controls1,515 (30%)
  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1,355 (27%)
  • 21 CFR 820.90Nonconforming product1,136 (23%)
  • 21 CFR 820.80Receiving, in-process, and finished device acceptance1,125 (22%)

Ask FDAgent about 21 CFR 820.198

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

Topics that count this section:Device CAPA and complaints

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.198; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.