21 CFR 820.90: Nonconforming product
FDA's inspection data lists 2,478 Form 483 citations of 21 CFR 820.90 at 1,892 facilities, from October 2008 to January 16, 2026. 39 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 2,478
- At 2,279 inspections
- Facilities cited
- 1,892
- Distinct FEIs
- Most recent citation
- Jan 16, 2026
- Marson Medical, Inc.
- Warning letters
- 39
- State a violation of it
The regulation
Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).
§ 820.90 Nonconforming product.
(a) Control of nonconforming product. Each manufacturer shall establish and maintain procedures to control product that does not conform to specified requirements. The procedures shall address the identification, documentation, evaluation, segregation, and disposition of nonconforming product. The evaluation of nonconformance shall include a determination of the need for an investigation and notification of the persons or organizations responsible for the nonconformance. The evaluation and any investigation shall be documented.
(b) Nonconformity review and disposition. (1) Each manufacturer shall establish and maintain procedures that define the responsibility for review and the authority for the disposition of nonconforming product. The procedures shall set forth the review and disposition process. Disposition of nonconforming product shall be documented. Documentation shall include the justification for use of nonconforming product and the signature of the individual(s) authorizing the use.
(2) Each manufacturer shall establish and maintain procedures for rework, to include retesting and reevaluation of the nonconforming product after rework, to ensure that the product meets its current approved specifications. Rework and reevaluation activities, including a determination of any adverse effect from the rework upon the product, shall be documented in the DHR.
21 CFR Part 820: Quality Management System Regulation, Subpart IRead on eCFR
Citations by year
Form 483 citations of 21 CFR 820.90 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 201 in 2015
| Year | Form 483 citations |
|---|---|
| 2008 | 20 |
| 2009 | 124 |
| 2010 | 164 |
| 2011 | 173 |
| 2012 | 191 |
| 2013 | 167 |
| 2014 | 190 |
| 2015 | 201 |
| 2016 | 155 |
| 2017 | 190 |
| 2018 | 174 |
| 2019 | 172 |
| 2020 | 39 |
| 2021 | 56 |
| 2022 | 99 |
| 2023 | 124 |
| 2024 | 115 |
| 2025 | 119 |
| 2026 | 5 |
How FDA words this citation
FDA's short descriptions and citation text on these 2,478 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Nonconforming product, Lack of or inadequate procedures1,381 citations
- Nonconforming product control348 citations
- Product rework procedures, Lack of or inadequate procedures263 citations
- Procedures for product review,disposition lack of/inadequate236 citations
- Product rework documentation, DHR {see also 820.184}136 citations
Show 3 more
- Documentation of disposition, justification, signature29 citations
- Specific non-conforming product procedures28 citations
- Product rework adverse effects {see also 820.184}26 citations
Citation text
- “Procedures have not been adequately established to control product that does not conform to specified requirements.”996 citations
- “Products that do not conform to specifications are not adequately controlled.”348 citations
- “Procedures have not been established to control product that does not conform to specified requirements.”330 citations
- “Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.”188 citations
Show 2 more
- “Procedures for rework of nonconforming product have not been adequately established.”157 citations
- “Rework and reevaluation activities have not been fully documented in the device history record.”86 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 820.90, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Marson Medical, Inc.2026-01-16 · FEI 1000550978Nonconforming product, Lack of or inadequate procedures
- DASISIMULATIONS, LLC2026-01-16 · FEI 3030707552Nonconforming product, Lack of or inadequate procedures
- Vortex Surgical Inc.2026-01-08 · FEI 3013002167Nonconforming product, Lack of or inadequate procedures
- Solomon Park Research Laboratories, Inc.2026-01-08 · FEI 3023437Procedures for product review,disposition lack of/inadequate
- Peter Schiff Enterprises2025-12-31 · FEI 1000112898Nonconforming product, Lack of or inadequate procedures
Show 4 more
- Hatch Product Development LLC2025-12-11 · FEI 3030438338Nonconforming product, Lack of or inadequate procedures
- Superior Felt & Filtration, L.L.C.2025-12-05 · FEI 2000044808Nonconforming product, Lack of or inadequate procedures
- Safesource Direct LLC2025-12-05 · FEI 3021486363Nonconforming product, Lack of or inadequate procedures
- Creative Foam Corporation - Bremen2025-11-24 · FEI 3001814065Nonconforming product, Lack of or inadequate procedures
What the observations say
Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 820.90 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 2 observations tie this way. The highlighted sentence is the matched wording.
Legacy Pharmaceutical Packaging LLC2020-01-16 · Observation 2
Rework and reevaluation activities have not been documented in the device history record. Specifically, the packaging project manager said mislabeled packages were rejected in I (o) (4) performance qualification production lots. He said the contents I (6) (4) were removed from the rejected cartons and reused. The mislabeled packages were destroyed. There was no documentation in the device history records for the I (ID-:=-@1) performance …
This observation opens with FDA's standard wording for 21 CFR 820.90(b)(2).
Warning letters stating a violation of 21 CFR 820.90
39 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.
“Failure to establish and maintain procedures to control product that does not conform to specified requirements, as required by 21 CFR 820.90(a).”
Each letter below states the violation in these words.
- Olympus Medical Systems CorporationIssued 2026-10-07
- Peter Schiff EnterprisesIssued 2026-08-26
“Failure to establish and maintain procedures to control product that does not conform to specified requirements, including the identification, documentation, evaluation, segregation, and disposition of nonconforming product. The evaluation of nonconformance shall include a determination of the need for an investigation and notification of the persons or organizations responsible for the nonconformance. The evaluation and any investigation shall be documented, as required by 21 CFR 820.90(a).”
- Longhorn Vaccines and Diagnostics LLCIssued 2026-02-26
“Failure to adequately establish procedures to control product that does not conform to specified requirements, as required by 21 CFR § 820.90(a).”
- Diasol, Inc.Issued 2026-01-29
“Failure to adequately establish and maintain procedures to control product that does not conform to specified requirements, as required by 21CFR 820.90(a).”
- Beta Bionics, Inc.Issued 2026-01-28
“Failure to adequately establish and maintain procedures to control product that does not conform to specified requirements, as required by 21 CFR 820.90(a).”
- Physitemp Instruments, LLCIssued 2025-11-12
Often cited at the same inspection
Of the 2,279 inspections where FDA cited 21 CFR 820.90, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action1,418 (62%)
- 21 CFR 820.198Complaint files1,136 (50%)
- 21 CFR 820.30Design controls795 (35%)
- 21 CFR 820.50Purchasing controls763 (33%)
- 21 CFR 820.80Receiving, in-process, and finished device acceptance705 (31%)
- 21 CFR 820.70Production and process controls613 (27%)
Ask FDAgent about 21 CFR 820.90
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
Topics that count this section:Device CAPA and complaints
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.90; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.