Skip to content

21 CFR 820.50: Purchasing controls

FDA's inspection data lists 3,174 Form 483 citations of 21 CFR 820.50 at 2,283 facilities, from October 2008 to April 16, 2026. 44 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
3,174
At 2,942 inspections
Facilities cited
2,283
Distinct FEIs
Most recent citation
Apr 16, 2026
Emerald Supply Inc
Warning letters
44
State a violation of it

The regulation

Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).

§ 820.50 Purchasing controls.

Each manufacturer shall establish and maintain procedures to ensure that all purchased or otherwise received product and services conform to specified requirements.

(a) Evaluation of suppliers, contractors, and consultants. Each manufacturer shall establish and maintain the requirements, including quality requirements, that must be met by suppliers, contractors, and consultants. Each manufacturer shall:

(1) Evaluate and select potential suppliers, contractors, and consultants on the basis of their ability to meet specified requirements, including quality requirements. The evaluation shall be documented.

(2) Define the type and extent of control to be exercised over the product, services, suppliers, contractors, and consultants, based on the evaluation results.

(3) Establish and maintain records of acceptable suppliers, contractors, and consultants.

(b) Purchasing data. Each manufacturer shall establish and maintain data that clearly describe or reference the specified requirements, including quality requirements, for purchased or otherwise received product and services. Purchasing documents shall include, where possible, an agreement that the suppliers, contractors, and consultants agree to notify the manufacturer of changes in the product or service so that manufacturers may determine whether the changes may affect the quality of a finished device. Purchasing data shall be approved in accordance with § 820.40.

21 CFR Part 820: Quality Management System Regulation, Subpart ERead on eCFR

Citations by year

Form 483 citations of 21 CFR 820.50 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 266 in 2012

Form 483 citations of 21 CFR 820.50 by inspection year
YearForm 483 citations
200829
2009168
2010183
2011232
2012266
2013231
2014231
2015216
2016187
2017220
2018239
2019219
202044
202187
2022137
2023169
2024136
2025175
20265

How FDA words this citation

FDA's short descriptions and citation text on these 3,174 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Purchasing controls, Lack of or inadequate procedures1,632 citations
  • Evaluation of suppliers, contractors, etc., requirements396 citations
  • Acceptable supplier records, inadequate records242 citations
  • Documented evaluation229 citations
  • Supplier oversight215 citations
Show 3 more
  • Supplier notification of changes185 citations
  • Evaluation and Selection, Suppliers, Contractors, etc.167 citations
  • Approval, inadequate purchasing data88 citations

Citation text

  • “Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.”1,086 citations
  • “Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.”462 citations
  • “Requirements that must be met by suppliers have not been adequately established.”124 citations
  • “The evaluation of potential suppliers was not documented.”118 citations
Show 2 more
  • “Records of acceptable suppliers have not been adequately established.”102 citations
  • “The type and extent of control to be exercised over suppliers was not clearly defined.”82 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 820.50, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

  • Emerald Supply Inc2026-04-16 · FEI 3011714211Approval, inadequate purchasing data
  • MIDWAY MEDICAL2026-01-15 · FEI 3015536408Purchasing controls, Lack of or inadequate procedures
  • Pleural Dynamics, Inc.2026-01-13 · FEI 3029300753Evaluation of suppliers, contractors, etc., requirements
  • Osseofuse International Inc.2026-01-07 · FEI 3008494722Documented evaluation
  • Peter Schiff Enterprises2025-12-31 · FEI 1000112898Purchasing controls, Lack of or inadequate procedures
  • ZIOSOFT, INC.2025-12-18 · FEI 3006621332Purchasing controls, Lack of or inadequate procedures
Show 4 more
  • My Medical Store LLC2025-12-18 · FEI 3024988980Evaluation of suppliers, contractors, etc., requirements
  • NeuroCareAI, Inc. dba SaveLife AI, Inc.2025-12-12 · FEI 3034656245Evaluation of suppliers, contractors, etc., requirements
  • Specialty Health Products CO, LLC2025-12-11 · FEI 2027347Purchasing controls, Lack of or inadequate procedures
  • SMC2025-12-11 · FEI 3011221234Supplier notification of changes

What the observations say

Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 820.50 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 1 observation ties this way. The highlighted sentence is the matched wording.

  • FRUEH VERPACKUNGSTECHNIK AG2023-03-24 · Observation 2

    Requirements that must be met by suppliers have not been adequately established. Specifically, components of (b)(4) drug product are not adequately controlled for bioburden levels. Specifications for (b)(4) do not include a bioburden requirement and incoming certificates of conformity do not certify that components meet a bioburden requirement.

    This observation opens with FDA's standard wording for 21 CFR 820.50(a).

Warning letters stating a violation of 21 CFR 820.50

44 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.

“Failure to establish and maintain procedures to ensure that all purchased or otherwise received product, and services conform to specified requirements, as required by 21 CFR 820.50.”
“Failure to establish and maintain procedures to ensure that all purchased or otherwise received product and services conform to specified requirements, as required by 21 CFR 820.50.”

Each letter below states the violation in these words.

“Failure to adequately establish and maintain the requirements, including quality requirements, that must be met by suppliers, contractors, and consultants, as required by 21 CFR 820.50(a).”
“Failure to adequately establish requirements that must be met by contractors and consultants, as required by 21 CFR § 820.50(a).”
“Failure to adequately establish procedures to ensure that all purchased or otherwise received products and services conform to specified requirements, as required by 21 CFR 820.50.”

Often cited at the same inspection

Of the 2,942 inspections where FDA cited 21 CFR 820.50, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 820.50

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

Topics that count this section:Supplier controls

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.50; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.