21 CFR 820.22: Quality audit
FDA's inspection data lists 2,151 Form 483 citations of 21 CFR 820.22 at 1,742 facilities, from October 2008 to January 16, 2026. 22 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 2,151
- At 2,076 inspections
- Facilities cited
- 1,742
- Distinct FEIs
- Most recent citation
- Jan 16, 2026
- Marson Medical, Inc.
- Warning letters
- 22
- State a violation of it
The regulation
Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).
§ 820.22 Quality audit.
Each manufacturer shall establish procedures for quality audits and conduct such audits to assure that the quality system is in compliance with the established quality system requirements and to determine the effectiveness of the quality system. Quality audits shall be conducted by individuals who do not have direct responsibility for the matters being audited. Corrective action(s), including a reaudit of deficient matters, shall be taken when necessary. A report of the results of each quality audit, and reaudit(s) where taken, shall be made and such reports shall be reviewed by management having responsibility for the matters audited. The dates and results of quality audits and reaudits shall be documented.
21 CFR Part 820: Quality Management System Regulation, Subpart BRead on eCFR
Citations by year
Form 483 citations of 21 CFR 820.22 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 205 in 2012
| Year | Form 483 citations |
|---|---|
| 2008 | 27 |
| 2009 | 132 |
| 2010 | 177 |
| 2011 | 204 |
| 2012 | 205 |
| 2013 | 156 |
| 2014 | 166 |
| 2015 | 139 |
| 2016 | 122 |
| 2017 | 125 |
| 2018 | 128 |
| 2019 | 131 |
| 2020 | 31 |
| 2021 | 41 |
| 2022 | 91 |
| 2023 | 90 |
| 2024 | 75 |
| 2025 | 107 |
| 2026 | 4 |
How FDA words this citation
FDA's short descriptions and citation text on these 2,151 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Quality audits - Lack of or inadequate procedures1,011 citations
- Quality Audit/Reaudit - conducted528 citations
- Quality Audits - defined intervals318 citations
- Quality audits - auditor independence243 citations
- Quality Audit/Reaudit - documentation33 citations
Show 2 more
- Quality audit - audit report review11 citations
- Quality audit corrective action, reaudits {see also 820.100}7 citations
Citation text
- “Procedures for quality audits have not been adequately established.”586 citations
- “Quality audits have not been performed.”393 citations
- “Procedures for quality audits have not been established.”335 citations
- “Individuals who conduct quality audits have direct responsibility for the matters being audited.”243 citations
Show 2 more
- “Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.”151 citations
- “Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.”67 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 820.22, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Peter Schiff Enterprises2025-12-31 · FEI 1000112898Quality audits - Lack of or inadequate procedures
Show 4 more
- Specialty Health Products CO, LLC2025-12-11 · FEI 2027347Quality audits - Lack of or inadequate procedures
Warning letters stating a violation of 21 CFR 820.22
22 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.
“Failure to establish procedures for quality audits and conduct such audits to assure that the quality system is in compliance with the established quality system requirements, as required by 21 CFR 820.22.”
- BlephEx, LLCIssued 2026-06-03
“Failure to establish procedures for quality audits and conduct such audits to assure that the quality system is in compliance with the established quality system requirements and to determine the effectiveness of the quality system, as required by 21 CFR 820.22.”
- Longhorn Vaccines and Diagnostics LLCIssued 2026-02-26
“Your firm has failed to establish and maintain adequate procedures for quality audits, as required by 21 CFR 820.22.”
- Technological Medical Advancements LLCIssued 2025-09-26
“Failure to establish procedures for quality audits, as required by 21 CFR § 820.22.”
- SeniorLife Technologies, Inc.Issued 2025-08-21
“Failure to establish procedures for quality audits and conduct such audits to assure that the quality system is in compliance with the established quality system requirements and to determine the effectiveness of the quality system, as required by 21 CFR § 820.22.”
- Spectra Therapy, LLCIssued 2025-07-22
“Failure to establish procedures for quality audits and conduct such audits to assure that the quality system is in compliance, as required by 21 CFR 820.22.”
- O3UV, LLCIssued 2025-07-07
Often cited at the same inspection
Of the 2,076 inspections where FDA cited 21 CFR 820.22, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action1,232 (59%)
- 21 CFR 820.198Complaint files1,114 (54%)
- 21 CFR 820.20Management responsibility936 (45%)
- 21 CFR 820.50Purchasing controls912 (44%)
- 21 CFR 820.30Design controls899 (43%)
- 21 CFR 820.80Receiving, in-process, and finished device acceptance596 (29%)
Ask FDAgent about 21 CFR 820.22
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.22; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.