FDA has 5 inspections on file for Peter Schiff Enterprises in Cookeville, Tennessee (FEI 1000112898), from 2010 to 2025. The latest ended on and was classified OAI, official action indicated. FDA has issued 2 warning letters under this FEI, the latest on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 12 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No court action or import refusal is filed under this FEI.

Seven regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
5, 2010–2025
2 OAI · 3 VAI · 0 NAI
Inspections with 483 observations shown
4 of 5
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Feb 25, 2026
Warning letters
· CDRH ↗
2 on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR - not or inadequately maintained

      A device history record has not been maintained.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Management review - Lack of or inadequate procedures

      Procedures for management review have not been adequately established.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been.

    5. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    Show 7 more citations
    1. Devices subject to device identification GUDID data submission requirements.

      The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Design history file

      The design history file was not established.

    4. Label to bear a unique device identifier

      The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    6. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    7. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    2. Management review - Lack of or inadequate procedures

      Procedures for management review have not been established.

    3. Not approved or obsolete document retrieval

      Documents that were obsolete were observed at a location where they are being used.

    4. DHR documentation of label release {see also 820.184}

      The DHR does not include records of examination and release of device labeling, including date and signature of the examiner.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    Show 3 more citations
    1. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    3. Quality policy and objectives

      The quality policy and quality objectives was/were not established by management with executive responsibility.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    2. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    3. Management review - Lack of or inadequate procedures

      Procedures for management review have not been established.

    4. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

    5. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Acceptance status

      The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria.

    2. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.

    3. Lack of procedures, or not maintained

      Document control procedures have not been established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    5. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    Show 4 more citations
    1. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    2. Quality System Procedures

      Quality system procedures and instructions have not been established.

    3. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a method, process, and procedure have not been established.

    4. Management review - Lack of or inadequate procedures

      Procedures for management review have not been established.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Feb 25, 2026. It asked for a Form 483.

  • Inspection ending Dec 31, 2025OAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 25, 2026. FDAgent holds no copy that FDA posted.

    1. Feb 25, 2026FDA-FOIA-2026-1958Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 17 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.

Show all 17

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000112898 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000112898. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.