21 CFR 820.86: Acceptance status

FDA's inspection data lists 102 Form 483 citations of 21 CFR 820.86 at 98 facilities, from April 2009 to December 9, 2025. 5 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
102
At 102 inspections
Facilities cited
98
Distinct FEIs
Most recent citation
Dec 9, 2025
Moldworks Worldwide LLC
Warning letters
5
State a violation of it

The regulation

Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).

§ 820.86 Acceptance status.

Each manufacturer shall identify by suitable means the acceptance status of product, to indicate the conformance or nonconformance of product with acceptance criteria. The identification of acceptance status shall be maintained throughout manufacturing, packaging, labeling, installation, and servicing of the product to ensure that only product which has passed the required acceptance activities is distributed, used, or installed.

21 CFR Part 820: Quality Management System Regulation, Subpart HRead on eCFR

Citations by year

Form 483 citations of 21 CFR 820.86 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 11 in 2012

Form 483 citations of 21 CFR 820.86 by inspection year
YearForm 483 citations
20094
20105
20118
201211
20136
20145
20158
201611
201711
20186
20196
20201
20212
20223
20237
20245
20253

How FDA words this citation

FDA's short descriptions and citation text on these 102 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Acceptance status99 citations
  • Maintained3 citations

Citation text

  • “The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria.”57 citations
  • “The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria and maintained.”24 citations
  • “The acceptance status of product was not maintained.”16 citations
  • “The acceptance status of product was not clearly identified.”2 citations
Show 2 more
  • “The acceptance status of product was not clearly identified throughout labeling of the product.”1 citations
  • “The acceptance status of product was not clearly identified throughout manufacturing of the product.”1 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 820.86, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

Show 4 more
  • Ecoguard Inc.2024-04-23 · FEI 3016667418Acceptance status
  • WHPM Inc.2024-01-30 · FEI 3003102569Acceptance status
  • Johnson Medtech LLC2023-11-27 · FEI 3013522941Acceptance status
  • Control-X Medical Zrt.2023-08-10 · FEI 3012093201Acceptance status

What the observations say

Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 820.86 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 2 observations tie this way. The highlighted sentence is the matched wording.

  • Foremount Enterprise Co., Ltd.2025-06-19 · Observation 7

    The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria. Specifically, on June 18, 2025, during the inspection of production and process activities related to your Class II Non-Rebreathing Valves, Product Code CBP, exported and intended for the US market, I identified multiple raw materials and components used in the manufacturing of medical devices exported and intended for the …

    This observation opens with FDA's standard wording for 21 CFR 820.86.

  • Hospira, Inc.2016-06-08 · Observation 8

    The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria. Specifically, you did not identify by suitable means, the acceptance status of product, to indicate the conformance or nonconformance with acceptance criteria. You did not perform final acceptance activities for iSecure 2, a class II device to assure the device is functioning as intended. iSecure 2 is used in the withdrawal and …

    This observation opens with FDA's standard wording for 21 CFR 820.86.

Warning letters stating a violation of 21 CFR 820.86

5 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.

“Failure to adequately identify by suitable means the acceptance status of product, to indicate the conformance or nonconformance of product with acceptance criteria, as required by 21 CFR 820.86.”
“Failure to identify by suitable means the acceptance status of product, to indicate the conformance or nonconformance of product with acceptance criteria, as required by 21 CFR 820.86.”

Each letter below states the violation in these words.

“Failure to identify by suitable means the acceptance status of the product, to indicate the conformance or nonconformance of product with acceptance criteria, as required by 21 CFR 820.86.”

Often cited at the same inspection

Of the 102 inspections where FDA cited 21 CFR 820.86, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 820.86

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.86; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.