WHPM Inc.

5358 Irwindale Ave Ste B, Irwindale, California, United States

FEI 3003102569

FDA has 8 inspections on file for WHPM Inc. in Irwindale, California (FEI 3003102569), from 2008 to 2024. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 6 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.

The most recent OAI inspection ended on Nov 12, 2008. FDA has issued 2 warning letters under this FEI, the latest on (CDRH · Office of Medical Device and Radiological Health Division 3). No court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.184 and 21 CFR 820.50(a)(3).

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
8, 2008–2024
1 OAI · 3 VAI · 4 NAI
Inspections with 483 observations shown
3 of 8
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Aug 8, 2024
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    2. Procedures for product review,disposition lack of/inadequate

      Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been established.

    3. Periodic equipment inspections

      Periodic inspections of equipment were not conducted to ensure adherence to applicable maintenance schedules.

    4. Acceptance status

      The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria.

    5. DHR - not or inadequately maintained

      A device history record has not been maintained.

    Show 1 more citation
    1. Acceptable supplier records, inadequate records

      Records of acceptable suppliers and contractors have not been established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Acceptable supplier records, inadequate records

      Records of acceptable contractors have not been established.

    2. Info evaluated to determine if event was reportable

      The written MDR procedure does not include documentation and recordkeeping requirements for all information that was evaluated to determine if an event was reportable.

    3. Lack of or inadequate procedures for distribution

      Procedures for control and distribution of finished devices have not been adequately established.

  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  5. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  7. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices16 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Identification procedures, Lack of or inadequate procedures

      Procedures for identifying product throughout all stages of receipt, production, packaging, distribution, and installation were not established, defined, documented, complete, and implemented.

    2. Quality Audit/Reaudit - conducted

      Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.

    3. Lack of or inadequate DHR procedures

      Procedures have not been established to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.

    4. Management review - defined interval, sufficient frequency

      Management reviews were not conducted at defined intervals.

    5. Evaluation of suppliers, contractors, etc., requirements

      Adequate quality requirements that must be met by suppliers, contractors, and consultants were not established, defined, documented, and implemented.

    Show 11 more citations
    1. Design history file

      The design history file was not established or maintained.

    2. Lack of or inadequate procedures for labeling

      Labeling procedures were not established, defined, documented, complete, and implemented.

    3. Establish the design and development plan

      A design and development plan has not been established, defined, documented, and implemented.

    4. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure that equipment is routinely calibrated, inspected, checked, and maintained were not established, defined, documented, complete, and implemented.

    5. Process control procedures, Lack of or inadequate procedures

      Process controls do not provide for documented instructions, standard operating procedures (SOPs), and methods that define and control the manner of production, monitoring and control of process parameters and component and device characteristics during production, and criteria for workmanship expressed in documented standards or by means of identified and approved representative samples.

    6. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not implemented.

    7. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities were not established, defined, documented, complete, and implemented.

    8. Training

      Employees have not been adequately trained.

    9. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated and approved according to established procedures.

    10. Lack of procedures, or not maintained

      Document control procedures were not implemented.

    11. General

      Complaint handling procedures for reviewing and evaluating complaints have not been implemented.

  8. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Bioresearch MonitoringCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Aug 8, 2024. It asked for a Form 483.

  • Inspection ending Jan 30, 2024VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Aug 8, 2024FDA-FOIA-2024-7108Form 483· 1 of 5 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 19 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 8 inspections, with how many of them cited each.

Show all 19

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003102569 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003102569. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.