21 CFR 820.80: Receiving, in-process, and finished device acceptance
FDA's inspection data lists 2,540 Form 483 citations of 21 CFR 820.80 at 1,903 facilities, from October 2008 to January 12, 2026. 37 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 2,540
- At 2,305 inspections
- Facilities cited
- 1,903
- Distinct FEIs
- Most recent citation
- Jan 12, 2026
- Innova Vascular Inc
- Warning letters
- 37
- State a violation of it
The regulation
Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).
§ 820.80 Receiving, in-process, and finished device acceptance.
(a) General. Each manufacturer shall establish and maintain procedures for acceptance activities. Acceptance activities include inspections, tests, or other verification activities.
(b) Receiving acceptance activities. Each manufacturer shall establish and maintain procedures for acceptance of incoming product. Incoming product shall be inspected, tested, or otherwise verified as conforming to specified requirements. Acceptance or rejection shall be documented.
(c) In-process acceptance activities. Each manufacturer shall establish and maintain acceptance procedures, where appropriate, to ensure that specified requirements for in-process product are met. Such procedures shall ensure that in-process product is controlled until the required inspection and tests or other verification activities have been completed, or necessary approvals are received, and are documented.
(d) Final acceptance activities. Each manufacturer shall establish and maintain procedures for finished device acceptance to ensure that each production run, lot, or batch of finished devices meets acceptance criteria. Finished devices shall be held in quarantine or otherwise adequately controlled until released. Finished devices shall not be released for distribution until:
(1) The activities required in the DMR are completed;
(2) the associated data and documentation is reviewed;
(3) the release is authorized by the signature of a designated individual(s); and
(4) the authorization is dated.
(e) Acceptance records. Each manufacturer shall document acceptance activities required by this part. These records shall include:
(1) The acceptance activities performed;
(2) the dates acceptance activities are performed;
(3) the results;
(4) the signature of the individual(s) conducting the acceptance activities; and
(5) where appropriate the equipment used. These records shall be part of the DHR.
21 CFR Part 820: Quality Management System Regulation, Subpart HRead on eCFR
Citations by year
Form 483 citations of 21 CFR 820.80 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 232 in 2012
| Year | Form 483 citations |
|---|---|
| 2008 | 20 |
| 2009 | 169 |
| 2010 | 186 |
| 2011 | 218 |
| 2012 | 232 |
| 2013 | 213 |
| 2014 | 188 |
| 2015 | 186 |
| 2016 | 145 |
| 2017 | 182 |
| 2018 | 151 |
| 2019 | 159 |
| 2020 | 41 |
| 2021 | 64 |
| 2022 | 80 |
| 2023 | 101 |
| 2024 | 102 |
| 2025 | 101 |
| 2026 | 2 |
How FDA words this citation
FDA's short descriptions and citation text on these 2,540 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Lack of or inadequate final acceptance procedures666 citations
- Lack of or inadequate procedures - Acceptance activities559 citations
- Lack of or inadequate receiving acceptance procedures536 citations
- Documentation432 citations
- Lack of or inadequate In-process acceptance procedures183 citations
Show 3 more
- Incoming acceptance records, documentation108 citations
- Incoming inspection, testing, verification23 citations
- No release signature9 citations
Citation text
- “Procedures for finished device acceptance have not been adequately established.”455 citations
- “Procedures for acceptance of incoming product have not been adequately established.”348 citations
- “Procedures for acceptance activities have not been adequately established.”346 citations
- “Procedures for acceptance activities have not been established.”167 citations
Show 2 more
- “Procedures for finished device acceptance have not been established.”155 citations
- “Procedures for acceptance of incoming product have not been established.”143 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 820.80, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Osseofuse International Inc.2026-01-07 · FEI 3008494722Lack of or inadequate final acceptance procedures
- Peter Schiff Enterprises2025-12-31 · FEI 1000112898Lack of or inadequate final acceptance procedures
- DiAgam Societe Anonyme (S.A.)2025-12-18 · FEI 3003749380Lack of or inadequate In-process acceptance procedures
- My Medical Store LLC2025-12-18 · FEI 3024988980Lack of or inadequate receiving acceptance procedures
- Mercury Enterprises, Inc. dba Mercury Medical2025-12-11 · FEI 1024404Lack of or inadequate receiving acceptance procedures
Show 4 more
- Specialty Health Products CO, LLC2025-12-11 · FEI 2027347Lack of or inadequate receiving acceptance procedures
- Hatch Product Development LLC2025-12-11 · FEI 3030438338Lack of or inadequate final acceptance procedures
What the observations say
Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 820.80 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 3 observations tie this way. The highlighted sentence is the matched wording.
Foremount Enterprise Co., Ltd.2025-06-19 · Observation 6
Procedures for acceptance activities have not been adequately established. Specifically, on June 18, 2025, during the inspection of production and process activities I identified multiple boxes Top Shelf, Par (b) (4) 2, components (approximately (b) (4) ), Lot# (b) (4) which per your firm's staff is used in the manufacturing of your Class II Non-Rebreathing Valves, Product Code CBP. However, per your material specifications CPR One-Way Value, …
This observation opens with FDA's standard wording for 21 CFR 820.80(a).
Engage Uni LLC2021-11-12 · Observation 1
Procedures for acceptance activities have not been adequately established. Specifically, Your firm's machining and finishing supplier for (b) (4), (b) (4), and (b) (4) failed to identify results of in-process inspection that deviated from the specifications documented on the inspection form, and these out of specification measurements were subsequently unidentified during DHR review conducted by your firm upon acceptance of finished devices. All …
This observation opens with FDA's standard wording for 21 CFR 820.80(a).
Alphatec Spine, Inc.2015-03-13 · Observation 9
Procedures for acceptance activities have not been adequately established. Specifically, Document/Procedure No.: QCP-042 - Supplied Material Inspection, Revision J is not adequately defined in that Section No.: 6.9.8.1 provides instructions, "(b) (4)." However, the instructions do not define where on the (b) (4) (i.e., beginning, middle, end, etc.) to take the measurements. For example, regarding the receipt of Part No.: (b) (4) Lot No.: (b) (4) …
This observation opens with FDA's standard wording for 21 CFR 820.80(a).
Warning letters stating a violation of 21 CFR 820.80
37 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.
“Failure to establish and maintain procedures for acceptance of incoming product, as required by 21 CFR 820.80(b).”
- Olympus Medical Systems CorporationIssued 2026-10-07
“Failure to establish and maintain procedures for finished device acceptance to ensure that each production run, lot, or batch of finished devices meets acceptance criteria, as required by 21 CFR 820.80(d).”
Each letter below states the violation in these words.
- Peter Schiff EnterprisesIssued 2026-08-26
- Physitemp Instruments, LLCIssued 2025-11-12
“Failure to adequately establish procedures for finished device acceptance, as required by 21 CFR § 820.80(d).”
- Diasol, Inc.Issued 2026-01-29
“Failure to adequately establish and maintain procedures for finished device acceptance to ensure that each production run, lot, or batch of finished devices meets acceptance criteria, as required by 21 CFR 820.80(d).”
- Abbott Diabetes Care, Inc.Issued 2026-01-23
“Failure to establish and maintain procedures for acceptance of incoming product. Incoming product shall be inspected, tested, or otherwise verified as conforming to specified requirements. Acceptance or rejection shall be documented, as required by 21 CFR 820.80(b).”
- Meridian Bioscience, IncIssued 2025-12-22
Often cited at the same inspection
Of the 2,305 inspections where FDA cited 21 CFR 820.80, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action1,267 (55%)
- 21 CFR 820.198Complaint files1,125 (49%)
- 21 CFR 820.50Purchasing controls900 (39%)
- 21 CFR 820.30Design controls879 (38%)
- 21 CFR 820.90Nonconforming product705 (31%)
- 21 CFR 820.184Device history record609 (26%)
Ask FDAgent about 21 CFR 820.80
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.80; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.