Engage Uni LLC
3505 Lake Lynda Dr, Orlando, Florida, United States
FEI 3016249968
FDA has one inspection on file for Engage Uni LLC in Orlando, Florida (FEI 3016249968). It ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA posted the Form 483 from that inspection. FDAgent read its 2 observations from the PDF; they are listed below. One FOIA request has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 1, in 2021
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 1 request, latest Aug 3, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesPosted 483 · 2 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Inadequate acceptance activities procedures
Procedures for acceptance activities have not been adequately established. Specifically, Your firm's machining and finishing supplier for (b) (4), (b) (4), and (b) (4) failed to identify results of in-process inspection that deviated from the specifications documented on the inspection form, and these out of specification measurements were subsequently unidentified during DHR review conducted by your firm upon acceptance of finished devices. All samples from Lots (b) (4) and (b) (4) of (b) (4), Part No. (b) (4) measured below the specification indicated for "Item #2," and were identified as "Accepts," in the supplier's provided manufacturing records. Additionally, your firm did not document Deviation PD01184, which initiated a change in the visual inspection procedure of implants, in the DHR reviews conducted for Lots (b) (4) and (b) (4) of (b) (4), Part No. (b) (4) This deviation was included in the manufacturing records provided by your firm's cleaning, packaging, and sterilization supplier, which were subject to DHR review.
- Observation 2
Inadequate CAPA trending procedures
Procedures for corrective and preventive action have not been adequately established. Specifically, Your firm does not have procedures established for the trending and evaluation of quality data to identify potential causes of nonconforming product or quality problems. Additionally, your firm has not conducted trending of quality data during 2021 for the purposes of identifying potential causes of nonconforming product or quality problems by means of the CAPA subsystem.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Aug 3, 2026. It asked for a Form 483.
Inspection ending Nov 12, 2021NAI
1 request. In the inspection history above. FDA has posted its Form 483.
- Aug 3, 2026FDA-FOIA-2026-7416Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3016249968 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3016249968. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.