Skip to content

Engage Uni LLC Form 483, November 2021

FDA issued this Form 483 to Engage Uni LLC in Orlando, Florida (FEI 3016249968) at the end of a device inspection that ended November 12, 2021. It lists 2 observations. FDA classified the inspection No Action Indicated (NAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Nov 19, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Nov 12, 2021
Devices · Postmarket Assurance: Devices
FDA's outcome
NAI
No Action Indicated
Posted by FDA
Nov 19, 2026
1,833 days after the inspection
Facility
FEI 3016249968
Orlando, Florida

The 2 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Inadequate acceptance activities procedures

    Procedures for acceptance activities have not been adequately established. Specifically, Your firm's machining and finishing supplier for (b) (4), (b) (4), and (b) (4) failed to identify results of in-process inspection that deviated from the specifications documented on the inspection form, and these out of specification measurements were subsequently unidentified during DHR review conducted by your firm upon acceptance of finished devices. All samples from Lots (b) (4) and (b) (4) of (b) (4), Part No. (b) (4) measured below the specification indicated for "Item #2," and were identified as "Accepts," in the supplier's provided manufacturing records. Additionally, your firm did not document Deviation PD01184, which initiated a change in the visual inspection procedure of implants, in the DHR reviews conducted for Lots (b) (4) and (b) (4) of (b) (4), Part No. (b) (4) This deviation was included in the manufacturing records provided by your firm's cleaning, packaging, and sterilization supplier, which were subject to DHR review.

  2. 2

    Inadequate CAPA trending procedures

    Procedures for corrective and preventive action have not been adequately established. Specifically, Your firm does not have procedures established for the trending and evaluation of quality data to identify potential causes of nonconforming product or quality problems. Additionally, your firm has not conducted trending of quality data during 2021 for the purposes of identifying potential causes of nonconforming product or quality problems by means of the CAPA subsystem.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.