Foremount Enterprise Co., Ltd.
Shengang District, No. 257 Changping Rd., Sec. 5; Shen Gang Qushi, Tai Zhong Shi, Taiwan
FEI 3010402691
FDA has one inspection on file for Foremount Enterprise Co., Ltd. in Tai Zhong Shi, Taiwan (FEI 3010402691). It ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA posted the Form 483 from that inspection. FDAgent read its 9 observations from the PDF; they are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 1, in 2025
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 1 request, latest Apr 1, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesPosted 483 · 9 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 121 CFR 830
Labels lack required unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830. Specifically, your firm has not ensured that your line of private label manufactured Class II Non-Rebreathing Valves, Product Code CBP medical devices your firm has exported and intended for the US market, bear a unique device identifier (UDI).
- Observation 221 CFR 803
No established complaint handling procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established. Specifically, your firm is a Manufacturer of Class II Non-Rebreathing Valves, Product Code CBP and Emergency Manual (Resuscitator) Ventilators, Product Code BTM, medical devices exported and intended for the US market, and your firm's Customer Complaint Handling Procedure, QP824, V.05, dated May 27, 2025, has no requirement to ensure your firm is receiving complaints for your medical devices exported and intended for the US market from your US market distributors to ensure your firm reviews and evaluates all complaints in a uniform and timely manner to determine whether an investigation is necessary and are evaluated to determine whether the complaint represents an event, which should be reported to the USFDA under 21 CFR Part 803 - Medical Device Reporting (MDR).
Show observations 3 to 9
- Observation 321 CFR 803
MDR procedures not developed and maintained
Written MDR procedures have not been developed and maintained. Specifically, your firm is a Manufacturer of Class II Non-Rebreathing Valves, Product Code CBP and Emergency Manual (Resuscitator) Ventilators, Product Code BTM, medical devices exported and intended for the US market, and your firm's Medical Device Report (MDR) procedure Product Vigilance and Market Surveillance Procedure, QP752, V.10, dated December 20, 2024, is not adequately established as to date your firm has not established procedure(s) to ensure your firm is receiving complaints for your medical devices exported and intended for the US market from your US market distributors to ensure your firm reviews and evaluates all complaints in a uniform and timely manner to determine whether an investigation is necessary and are evaluated to determine whether the complaint represents an event, that reasonably suggest the failure of a device to meet its performance specifications or otherwise perform as intended, which should be reported to the USFDA under 21 CFR Part 803 - Medical Device Reporting (MDR).
- Observation 4
Ultrasonic welding process not adequately validated
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures. Specifically, your firm's Validation Management Procedure, QP732, V.05, dated March 26, 2025, and Validation Work Instruction, WI7312, V.02, dated March 27, 2025, have not been adequately established. Your firm is a Manufacturer and Contract Manufacturer of various medical devices to include Class II Non-Rebreathing Valves, Product Code CBP exported and intended for the US market. Your firm conducts Ultrasonic Welding on machines (b) (4) in the manufacturing for your Class II Non-Rebreathing Valves, Product Code CBP exported and intended for the US market; however, your firm was not able to provide protocols and objective evidence to support your related Installation Qualification (IQ), Operational Qualification (OQ), and Performance Qualification (PQ) Reports FM-RN-01 and FM-RN-02 dated October 22, 2024.
- Observation 5
CAPA procedures not adequately established
Procedures for corrective and preventive action have not been adequately established. Specifically, your firm's corrective and preventative action (CAPA) procedure Problem Improvement Procedure, QP851, Ver.20, dated December 25, 2024, has not been adequately established. Your firm was not able to provide evidence of an established CAPA system.
- Observation 6
Acceptance procedures inadequate for incoming components
Procedures for acceptance activities have not been adequately established. Specifically, on June 18, 2025, during the inspection of production and process activities I identified multiple boxes Top Shelf, Par (b) (4) 2, components (approximately (b) (4) ), Lot# (b) (4) which per your firm's staff is used in the manufacturing of your Class II Non-Rebreathing Valves, Product Code CBP. However, per your material specifications CPR One-Way Value, Item# (b) (4) for the Class II Non-Rebreathing Valves, Product Code CBP, the correct Top Shelf component is Part# (b) (4) Additionally, your incoming acceptance activities conducted on Lot (b) (4) dated May 21, 2025, incorrectly documented Part# (b) (4) which was ultimately accepted by your firm to be used in the manufacturing of Class II Non-Rebreathing Valves, Product Code CBP, for medical devices potentially exported and intended for the US market.
- Observation 7
Product acceptance status not identified
The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria. Specifically, on June 18, 2025, during the inspection of production and process activities related to your Class II Non-Rebreathing Valves, Product Code CBP, exported and intended for the US market, I identified multiple raw materials and components used in the manufacturing of medical devices exported and intended for the market, which were not properly identified to ensure that only product which has passed the required acceptance activities are distributed or used.
- Observation 8
Device history record procedures inadequate
Procedures for device history records have not been adequately established. Your firm's Device History Record (DHR) procedure Product Identification and Traceability Procedure, QP751, V.18, dated July 12, 2024, has not been adequality established. Your firm was not able to provide DHRs for your Class II Non-Rebreathing Valves, Product Code CBP, for medical devices exported and intended for the US market, which demonstrated the devices were manufactured in accordance with the Device Maser Record (DMR).
- Observation 921 CFR 80321 CFR 80721 CFR 820
Personnel lack FDA regulatory training
Personnel do not have the necessary training and experience to perform their jobs. Specifically, throughout the inspection key members of your firm who participated and provided relevant information were not familiar with the USFDA regulations and requirements for Manufacturers and Contract Manufacturers exporting medical devices intended for the US market to include: a) 21 CFR Part 803 - Medical Device Reporting b) 21 CFR Part 807 - Establishment Registration and Device Listing for Manufacturers and Initial Importers of Devices c) 21 CFR Part 820 - Quality System Regulation
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Apr 1, 2026. It asked for a Form 483.
Inspection date not given
1 request. FDA released records in answer to a request for its Form 483 on Jun 30, 2026. FDAgent holds no copy that FDA posted.
- Apr 1, 2026FDA-FOIA-2026-3209Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3010402691 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3010402691. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.