Foremount Enterprise Co., Ltd Form 483, June 2025
FDA issued this Form 483 to Foremount Enterprise Co., Ltd in Tai Zhong Shi, Taiwan (FEI 3010402691) at the end of a device inspection that ended June 19, 2025. It lists 9 observations. FDA classified the inspection Official Action Indicated (OAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Oct 26, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Jun 19, 2025
- Devices · Compliance: Devices
- FDA's outcome
- OAI
- Official Action Indicated
- Posted by FDA
- Oct 26, 2026
- 494 days after the inspection
The 9 observations on this Form 483
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.
- 1
Labels lack required unique device identifier
21 CFR 830The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830. Specifically, your firm has not ensured that your line of private label manufactured Class II Non-Rebreathing Valves, Product Code CBP medical devices your firm has exported and intended for the US market, bear a unique device identifier (UDI).
- 2
No established complaint handling procedures
21 CFR 803Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established. Specifically, your firm is a Manufacturer of Class II Non-Rebreathing Valves, Product Code CBP and Emergency Manual (Resuscitator) Ventilators, Product Code BTM, medical devices exported and intended for the US market, and your firm's Customer Complaint Handling Procedure, QP824, V.05, dated May 27, 2025, has no requirement to ensure your firm is receiving complaints for your medical devices exported and intended for the US market from your US market distributors to ensure your firm reviews and evaluates all complaints in a uniform and timely manner to determine whether an investigation is necessary and are evaluated to determine whether the complaint represents an event, which should be reported to the USFDA under 21 CFR Part 803 - Medical Device Reporting (MDR).
- 3
MDR procedures not developed and maintained
21 CFR 803Written MDR procedures have not been developed and maintained. Specifically, your firm is a Manufacturer of Class II Non-Rebreathing Valves, Product Code CBP and Emergency Manual (Resuscitator) Ventilators, Product Code BTM, medical devices exported and intended for the US market, and your firm's Medical Device Report (MDR) procedure Product Vigilance and Market Surveillance Procedure, QP752, V.10, dated December 20, 2024, is not adequately established as to date your firm has not established procedure(s) to ensure your firm is receiving complaints for your medical devices exported and intended for the US market from your US market distributors to ensure your firm reviews and evaluates all complaints in a uniform and timely manner to determine whether an investigation is necessary and are evaluated to determine whether the complaint represents an event, that reasonably suggest the failure of a device to meet its performance specifications or otherwise perform as intended, which should be reported to the USFDA under 21 CFR Part 803 - Medical Device Reporting (MDR).
- 4
Ultrasonic welding process not adequately validated
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures. Specifically, your firm's Validation Management Procedure, QP732, V.05, dated March 26, 2025, and Validation Work Instruction, WI7312, V.02, dated March 27, 2025, have not been adequately established. Your firm is a Manufacturer and Contract Manufacturer of various medical devices to include Class II Non-Rebreathing Valves, Product Code CBP exported and intended for the US market. Your firm conducts Ultrasonic Welding on machines (b) (4) in the manufacturing for your Class II Non-Rebreathing Valves, Product Code CBP exported and intended for the US market; however, your firm was not able to provide protocols and objective evidence to support your related Installation Qualification (IQ), Operational Qualification (OQ), and Performance Qualification (PQ) Reports FM-RN-01 and FM-RN-02 dated October 22, 2024.
- 5
CAPA procedures not adequately established
Procedures for corrective and preventive action have not been adequately established. Specifically, your firm's corrective and preventative action (CAPA) procedure Problem Improvement Procedure, QP851, Ver.20, dated December 25, 2024, has not been adequately established. Your firm was not able to provide evidence of an established CAPA system.
- 6
Acceptance procedures inadequate for incoming components
Procedures for acceptance activities have not been adequately established. Specifically, on June 18, 2025, during the inspection of production and process activities I identified multiple boxes Top Shelf, Par (b) (4) 2, components (approximately (b) (4) ), Lot# (b) (4) which per your firm's staff is used in the manufacturing of your Class II Non-Rebreathing Valves, Product Code CBP. However, per your material specifications CPR One-Way Value, Item# (b) (4) for the Class II Non-Rebreathing Valves, Product Code CBP, the correct Top Shelf component is Part# (b) (4) Additionally, your incoming acceptance activities conducted on Lot (b) (4) dated May 21, 2025, incorrectly documented Part# (b) (4) which was ultimately accepted by your firm to be used in the manufacturing of Class II Non-Rebreathing Valves, Product Code CBP, for medical devices potentially exported and intended for the US market.
- 7
Product acceptance status not identified
The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria. Specifically, on June 18, 2025, during the inspection of production and process activities related to your Class II Non-Rebreathing Valves, Product Code CBP, exported and intended for the US market, I identified multiple raw materials and components used in the manufacturing of medical devices exported and intended for the market, which were not properly identified to ensure that only product which has passed the required acceptance activities are distributed or used.
- 8
Device history record procedures inadequate
Procedures for device history records have not been adequately established. Your firm's Device History Record (DHR) procedure Product Identification and Traceability Procedure, QP751, V.18, dated July 12, 2024, has not been adequality established. Your firm was not able to provide DHRs for your Class II Non-Rebreathing Valves, Product Code CBP, for medical devices exported and intended for the US market, which demonstrated the devices were manufactured in accordance with the Device Maser Record (DMR).
- 9
Personnel lack FDA regulatory training
21 CFR 80321 CFR 80721 CFR 820Personnel do not have the necessary training and experience to perform their jobs. Specifically, throughout the inspection key members of your firm who participated and provided relevant information were not familiar with the USFDA regulations and requirements for Manufacturers and Contract Manufacturers exporting medical devices intended for the US market to include: a) 21 CFR Part 803 - Medical Device Reporting b) 21 CFR Part 807 - Establishment Registration and Device Listing for Manufacturers and Initial Importers of Devices c) 21 CFR Part 820 - Quality System Regulation
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related