21 CFR 820.30: Design controls
FDA's inspection data lists 6,018 Form 483 citations of 21 CFR 820.30 at 2,834 facilities, from October 2008 to January 16, 2026. 102 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 6,018
- At 3,804 inspections
- Facilities cited
- 2,834
- Distinct FEIs
- Most recent citation
- Jan 16, 2026
- DASISIMULATIONS, LLC
- Warning letters
- 102
- State a violation of it
The regulation
Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).
§ 820.30 Design controls.
(a) General. (1) Each manufacturer of any class III or class II device, and the class I devices listed in paragraph (a)(2) of this section, shall establish and maintain procedures to control the design of the device in order to ensure that specified design requirements are met.
(2) The following class I devices are subject to design controls:
(i) Devices automated with computer software; and
(ii) The devices listed in the following chart.
Section | Device
868.6810 | Catheter, Tracheobronchial Suction.
878.4460 | Glove, Surgeon's.
880.6760 | Restraint, Protective.
892.5650 | System, Applicator, Radionuclide, Manual.
892.5740 | Source, Radionuclide Teletherapy.
(b) Design and development planning. Each manufacturer shall establish and maintain plans that describe or reference the design and development activities and define responsibility for implementation. The plans shall identify and describe the interfaces with different groups or activities that provide, or result in, input to the design and development process. The plans shall be reviewed, updated, and approved as design and development evolves.
(c) Design input. Each manufacturer shall establish and maintain procedures to ensure that the design requirements relating to a device are appropriate and address the intended use of the device, including the needs of the user and patient. The procedures shall include a mechanism for addressing incomplete, ambiguous, or conflicting requirements. The design input requirements shall be documented and shall be reviewed and approved by a designated individual(s). The approval, including the date and signature of the individual(s) approving the requirements, shall be documented.
(d) Design output. Each manufacturer shall establish and maintain procedures for defining and documenting design output in terms that allow an adequate evaluation of conformance to design input requirements. Design output procedures shall contain or make reference to acceptance criteria and shall ensure that those design outputs that are essential for the proper functioning of the device are identified. Design output shall be documented, reviewed, and approved before release. The approval, including the date and signature of the individual(s) approving the output, shall be documented.
(e) Design review. Each manufacturer shall establish and maintain procedures to ensure that formal documented reviews of the design results are planned and conducted at appropriate stages of the device's design development. The procedures shall ensure that participants at each design review include representatives of all functions concerned with the design stage being reviewed and an individual(s) who does not have direct responsibility for the design stage being reviewed, as well as any specialists needed. The results of a design review, including identification of the design, the date, and the individual(s) performing the review, shall be documented in the design history file (the DHF).
(f) Design verification. Each manufacturer shall establish and maintain procedures for verifying the device design. Design verification shall confirm that the design output meets the design input requirements. The results of the design verification, including identification of the design, method(s), the date, and the individual(s) performing the verification, shall be documented in the DHF.
(g) Design validation. Each manufacturer shall establish and maintain procedures for validating the device design. Design validation shall be performed under defined operating conditions on initial production units, lots, or batches, or their equivalents. Design validation shall ensure that devices conform to defined user needs and intended uses and shall include testing of production units under actual or simulated use conditions. Design validation shall include software validation and risk analysis, where appropriate. The results of the design validation, including identification of the design, method(s), the date, and the individual(s) performing the validation, shall be documented in the DHF.
(h) Design transfer. Each manufacturer shall establish and maintain procedures to ensure that the device design is correctly translated into production specifications.
(i) Design changes. Each manufacturer shall establish and maintain procedures for the identification, documentation, validation or where appropriate verification, review, and approval of design changes before their implementation.
(j) Design history file. Each manufacturer shall establish and maintain a DHF for each type of device. The DHF shall contain or reference the records necessary to demonstrate that the design was developed in accordance with the approved design plan and the requirements of this part.
21 CFR Part 820: Quality Management System Regulation, Subpart CRead on eCFR
Citations by year
Form 483 citations of 21 CFR 820.30 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 558 in 2012
| Year | Form 483 citations |
|---|---|
| 2008 | 57 |
| 2009 | 361 |
| 2010 | 435 |
| 2011 | 502 |
| 2012 | 558 |
| 2013 | 488 |
| 2014 | 481 |
| 2015 | 487 |
| 2016 | 345 |
| 2017 | 402 |
| 2018 | 361 |
| 2019 | 326 |
| 2020 | 94 |
| 2021 | 143 |
| 2022 | 225 |
| 2023 | 224 |
| 2024 | 237 |
| 2025 | 281 |
| 2026 | 11 |
How FDA words this citation
FDA's short descriptions and citation text on these 6,018 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Design changes - Lack of or Inadequate Procedures992 citations
- Design control - no procedures679 citations
- Design Validation - Risk analysis not performed/inadequate579 citations
- Design validation- Lack of or inadequate procedures498 citations
- Design history file471 citations
Show 3 more
- Design verification - Lack of or inadequate procedures283 citations
- Design review - Lack of or inadequate procedures225 citations
- Design verification - output does not meet input requirement203 citations
Citation text
- “Procedures for design change have not been adequately established.”715 citations
- “Procedures for design control have not been established.”589 citations
- “Procedures for design validation have not been adequately established.”395 citations
- “Risk analysis is inadequate.”223 citations
Show 2 more
- “Procedures for design verification have not been adequately established.”222 citations
- “Design verification does not confirm that design output meets design input requirements.”203 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 820.30, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- DASISIMULATIONS, LLC2026-01-16 · FEI 3030707552Design Validation - Risk analysis not performed/inadequate
- Innova Vascular Inc2026-01-12 · FEI 3020256033Design Validation - Risk analysis not performed/inadequate
Show 4 more
- Osseofuse International Inc.2026-01-07 · FEI 3008494722Design changes - Lack of or Inadequate Procedures
- Ashland Specialty Ingredients, G.P.2025-12-22 · FEI 3010785603Design verification - Lack of or inadequate procedures
What the observations say
Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 820.30 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 13 observations tie this way. The highlighted sentence is the matched wording.
Catalent Belgium SA2022-08-19 · Observation 7
The design history file was not established. Specifically, your firm has neither a procedure specifications, nor a device history file for customer's designs a[illegible]
This observation opens with FDA's standard wording for 21 CFR 820.30(j).
LATE NIGHT CONCEPTS INC. (DBA BRUX NIGHT GUARD)2021-10-28 · Observation 1
The design history file was not established. Specifically, your firm has not established a Design History File for Brux Night Guard per you SOP (Design Control, Doc #: 73-01, date: 10/23/2021).
This observation opens with FDA's standard wording for 21 CFR 820.30(j).
Dr. Reddy's Laboratories Ltd.2019-06-21 · Observation 1
Acceptance criteria were not established prior to the performance of validation activities. 4 A Ps . : (b) (4) . Specifically, is not clear how your firm determined the acceptance criteria for the Functionality Testing (Break loose force &, Glide force) conducted as part of finished product release specification for IB Injection®® mg@nL. For example, your firm explained that the worst-case condition was considered for the applied injection @@ …
This observation opens with FDA's standard wording for 21 CFR 820.30(g).
Show 2 more
Sun Pharmaceutical Industries Ltd.2019-06-11 · Observation 3
OBSERVATION 3 Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established. Specifically, your design plan for ~b)() __(Injection, E_J<> pcg/mL, tbH > mL & tbH > mL) associated to <b><> was approved on May 20[illegible]. However, your firm began the design activities associated to this product on August 2012. Additionally, your firm did not have an …
This observation opens with FDA's standard wording for 21 CFR 820.30(b).
Meridian Medical Technologies, LLC a Subsidiary of Kindeva Drug Delivery, L.P.2017-03-24 · Observation 6
Procedures for design input have not been adequately established. Specifically, A. The design inputs/requirements of the EpiPen, EpiPen Jr., and (b) (4) products are not appropriate for the intended use of the products in regards to the reliability of the delivery system. The reliability inputs/requirements of the device according to your PRD/TRD 16-001 Rev 1 do not include a system level reliability design input/requirement. 1. Your …
This observation opens with FDA's standard wording for 21 CFR 820.30(c).
Warning letters stating a violation of 21 CFR 820.30
102 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.
“Failure to establish and maintain procedures for validating the device design to ensure that devices conform to defined user needs and intended uses and validation shall include testing of production units under actual or simulated use conditions, as required by 21 CFR 820.30(g).”
- Olympus Medical Systems CorporationIssued 2026-10-07
“Failure to establish and maintain procedures to control the design of the device in order to ensure that specified design requirements are met, as required by 21 CFR 820.30(a).”
- Peter Schiff EnterprisesIssued 2026-08-26
“Failure to establish and maintain procedures for the identification, documentation, validation or where appropriate verification, review, and approval of design changes before their implementation, as required by 21 CFR 820.30(i).”
- Next Medical Products Company, LLCIssued 2026-08-19
“Failure to establish and maintain design validation procedures to ensure that devices conform to defined user needs and intended uses and shall include testing of production units under actual or simulated use conditions, as required by 21 CFR 820.30(g).”
- Specialty Health Products CO, LLCIssued 2026-07-07
“Failure to adequately establish and maintain procedures for the identification, documentation, validation or where appropriate verification, review, and approval of design changes before their implementation, as required by 21 CFR 820.30(i).”
- Happiest Baby, Inc.Issued 2026-06-15
“Failure to establish and maintain procedures to control the design of the device in order to ensure that specified design requirements are met, as required by 21 CFR 820.30(a), and failure to establish and maintain procedures for the identification, documentation, validation or verification, review, and approval of design changes before their implementation, as required by 21 CFR 820.30(i).”
- BlephEx, LLCIssued 2026-06-03
Often cited at the same inspection
Of the 3,804 inspections where FDA cited 21 CFR 820.30, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action2,091 (55%)
- 21 CFR 820.198Complaint files1,801 (47%)
- 21 CFR 820.50Purchasing controls1,325 (35%)
- 21 CFR 820.22Quality audit899 (24%)
- 21 CFR 820.80Receiving, in-process, and finished device acceptance879 (23%)
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?857 (23%)
Ask FDAgent about 21 CFR 820.30
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.30; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.