21 CFR 820.100: Corrective and preventive action
FDA's inspection data lists 6,080 Form 483 citations of 21 CFR 820.100 at 4,108 facilities, from October 2008 to July 16, 2026. 93 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 6,080
- At 5,653 inspections
- Facilities cited
- 4,108
- Distinct FEIs
- Most recent citation
- Jul 16, 2026
- Hollywog LLC
- Warning letters
- 93
- State a violation of it
The regulation
Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).
§ 820.100 Corrective and preventive action.
(a) Each manufacturer shall establish and maintain procedures for implementing corrective and preventive action. The procedures shall include requirements for:
(1) Analyzing processes, work operations, concessions, quality audit reports, quality records, service records, complaints, returned product, and other sources of quality data to identify existing and potential causes of nonconforming product, or other quality problems. Appropriate statistical methodology shall be employed where necessary to detect recurring quality problems;
(2) Investigating the cause of nonconformities relating to product, processes, and the quality system;
(3) Identifying the action(s) needed to correct and prevent recurrence of nonconforming product and other quality problems;
(4) Verifying or validating the corrective and preventive action to ensure that such action is effective and does not adversely affect the finished device;
(5) Implementing and recording changes in methods and procedures needed to correct and prevent identified quality problems;
(6) Ensuring that information related to quality problems or nonconforming product is disseminated to those directly responsible for assuring the quality of such product or the prevention of such problems; and
(7) Submitting relevant information on identified quality problems, as well as corrective and preventive actions, for management review.
(b) All activities required under this section, and their results, shall be documented.
21 CFR Part 820: Quality Management System Regulation, Subpart JRead on eCFR
Citations by year
Form 483 citations of 21 CFR 820.100 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 475 in 2012
| Year | Form 483 citations |
|---|---|
| 2008 | 73 |
| 2009 | 425 |
| 2010 | 429 |
| 2011 | 451 |
| 2012 | 475 |
| 2013 | 470 |
| 2014 | 438 |
| 2015 | 431 |
| 2016 | 373 |
| 2017 | 455 |
| 2018 | 371 |
| 2019 | 354 |
| 2020 | 87 |
| 2021 | 153 |
| 2022 | 221 |
| 2023 | 297 |
| 2024 | 253 |
| 2025 | 314 |
| 2026 | 10 |
How FDA words this citation
FDA's short descriptions and citation text on these 6,080 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Lack of or inadequate procedures4,330 citations
- Documentation1,314 citations
- Procedures68 citations
- Identification of actions needed65 citations
- CAPA verification/validation of corrective/preventive action58 citations
Show 3 more
- Investigation procedures37 citations
- Analysis of data sources35 citations
- Analysis of data/reports from data sources30 citations
Citation text
- “Procedures for corrective and preventive action have not been adequately established.”3,304 citations
- “Corrective and preventive action activities and/or results have not been adequately documented.”905 citations
- “Procedures for corrective and preventive action have not been established.”875 citations
- “Corrective and preventive action activities and/or results have not been documented.”279 citations
Show 2 more
- “Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.”65 citations
- “Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.”58 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 820.100, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
Show 4 more
What the observations say
Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 820.100 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 6 observations tie this way. The highlighted sentence is the matched wording.
Implantica CE Reflux Ltd2025-10-24 · Observation 1
Procedures for corrective and preventive action have not been adequately established. Specifically, the following deficiencies were found in these CAPA records: CAPA 00001: Inadequate Corrective Actions (ineffective training) and Effectiveness Check Plan (no interim checks) CAPA 00003: Inadequate Effectiveness verification plan for systemic QMS issue (no interim checks) CAPA 00002: 1. Investigation - Missing Process-Parameter Documentation The …
This observation opens with FDA's standard wording for 21 CFR 820.100(a).
Foremount Enterprise Co., Ltd.2025-06-19 · Observation 5
Procedures for corrective and preventive action have not been adequately established. Specifically, your firm's corrective and preventative action (CAPA) procedure Problem Improvement Procedure, QP851, Ver.20, dated December 25, 2024, has not been adequately established. Your firm was not able to provide evidence of an established CAPA system.
This observation opens with FDA's standard wording for 21 CFR 820.100(a).
Engage Uni LLC2021-11-12 · Observation 2
Procedures for corrective and preventive action have not been adequately established. Specifically, Your firm does not have procedures established for the trending and evaluation of quality data to identify potential causes of nonconforming product or quality problems. Additionally, your firm has not conducted trending of quality data during 2021 for the purposes of identifying potential causes of nonconforming product or quality problems by …
This observation opens with FDA's standard wording for 21 CFR 820.100(a).
Show 2 more
GPC Medical Ltd.2019-10-24 · Observation 3
Procedures for Corrective and Preventive Action have not been adequately established. CAPA procedure / QSP / 12 identifies product non conformance (receiving and finished product) as a Quality Data source for CAPA. The firm log book of product non conformance is not analyzed to identify recurring non conformance for Corrective actions. There have been no CAPA actions for product non conformance in the last three years.
This observation opens with FDA's standard wording for 21 CFR 820.100(a).
Alphatec Spine, Inc.2015-03-13 · Observation 5
Corrective and preventive action activities and/or results have not been adequately documented. Specifically, the firm acquired the (b) (4) (spinal implant device) from another medical device company in (b) (4). Previously, on (b) (4), the (b) (4) device was recalled, however, the firm has no records of the recall/corrections and removals on file to include receipt and final disposition of the returned units under recall nor any records …
This observation opens with FDA's standard wording for 21 CFR 820.100(b).
Warning letters stating a violation of 21 CFR 820.100
93 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.
“Failure to identify the action(s) needed to correct and prevent the recurrence of nonconforming product and other quality problems, as required by 21 CFR 820.100(a)(3).”
- Olympus Medical Systems CorporationIssued 2026-10-07
“Failure to establish procedures for corrective and preventive action, as required by 21 CFR 820.100(a).”
- Peter Schiff EnterprisesIssued 2026-08-26
“Failure to establish procedures for implementing corrective and preventive action, as required by 21 CFR 820.100(a).”
- Next Medical Products Company, LLCIssued 2026-08-19
“Failure to adequately establish and maintain procedures for implementing corrective and preventive action, as required by 21 CFR 820.100(a).”
- Happiest Baby, Inc.Issued 2026-06-15
“Failure to establish and maintain procedures for implementing corrective and preventive action, as required by 21 CFR 820.100(a), and failure to document corrective and preventive action activities and results, as required by 21 CFR 820.100(b).”
- BlephEx, LLCIssued 2026-06-03
“Failure to adequately establish and maintain procedures for corrective and preventive action, as required by 21 CFR 820.100(a).”
- ZIIP, Inc.Issued 2026-05-22
Often cited at the same inspection
Of the 5,653 inspections where FDA cited 21 CFR 820.100, the share where it also cited each of these sections.
- 21 CFR 820.198Complaint files2,869 (51%)
- 21 CFR 820.30Design controls2,091 (37%)
- 21 CFR 820.50Purchasing controls1,699 (30%)
- 21 CFR 820.90Nonconforming product1,418 (25%)
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1,267 (22%)
- 21 CFR 820.22Quality audit1,232 (22%)
Ask FDAgent about 21 CFR 820.100
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
Topics that count this section:Device CAPA and complaints
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.100; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.