GPC Medical Ltd.
Plot C-3, Sec B-1, UPSIDC, Trans Delhi Signature City, Loni,, Loni, India
FEI 3006513362
FDA has 2 inspections on file for GPC Medical Ltd. in Loni, India (FEI 3006513362), from 2011 to 2019. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA posted the Form 483 from that inspection. FDAgent read its 5 observations from the PDF; they are listed below. One FOIA request has asked FDA for this inspection's 483.
FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 2, 2011–2019
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- 1 posted 483 not matched to an inspection
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 1 request, latest Sep 29, 2025
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices5 citations published
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
Inadequate process control procedures for cleaning
Process Control Procedures that describe any process controls necessary to ensure conformance to the specifications have not been adequately established. For example: The Procedure for [illegible] does not identify the temperature that the process is operated at. The Cleaning process validation was performed at a temperature of [illegible]. During a walk through of the manufacturing facility on Tuesday October 22, I observed the (b)(4) Cleaning process operating at a temperature of [illegible]. Also, the procedure for the (b)(4) cleaning does not provide sufficient detail for how parts will be rinsed after [illegible] cleaning such as volume of water used and rinsing time. The Procedure for (b)(4) identifies that the (b)(4) consists of equal parts of [illegible]. The Validation of the process used an (b)(4) containing (b)(4)
- Observation 2
Inadequate in-process acceptance procedures for machining
Procedures for acceptance of in-process product have not been adequately established. Specifically, during a walk through of the Manufacturing Production area I observed the Machining of 4.0MM Canulated Cancellous Screw / LOT (b)(4). The First Piece Approval document indicated that the thread length of the piece was 7mm and three pieces were measured and were reported as 7.0mm, 7.1mm and 7.1mm. The screw length is 20mm and the thread length should be 8mm. Re-measurement determined that the screw thread length was actually 8mm.
Show observations 3 to 5
- Observation 3
CAPA procedures not analyzing product nonconformances
Procedures for Corrective and Preventive Action have not been adequately established. CAPA procedure / QSP / 12 identifies product non conformance (receiving and finished product) as a Quality Data source for CAPA. The firm log book of product non conformance is not analyzed to identify recurring non conformance for Corrective actions. There have been no CAPA actions for product non conformance in the last three years.
- Observation 4
Inadequate validation of cleaning process
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated. Specifically, the process validation for cleaning did not demonstrate that the acceptance was satisfied. The testing after criteria for cleaning only reported a (b)(4) (b)(4) difference to [illegible] and the acceptance criteria established that there would be no "neat" water. Also cleaning validation did not define or describe that the product will be rinsed after cleaning in the enzymatic reagent. Further the chemicals of the enzymatic reagent were not tested for as part of the residual testing.
- Observation 5
No maintenance schedule for RO water system
Schedules for maintenance of equipment have not been adequately established. Specifically, the Reverse (b)(4) Osmosis water system is reportedly serviced by [illegible]. There are no service reports of these activities. The firm has not evaluated the water quality of the RO water used for the cleaning since 2017.
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training records
Personnel training is not documented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Sep 29, 2025. It asked for a Form 483.
Inspection ending Oct 24, 2019OAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on May 27, 2026. FDAgent holds no copy that FDA posted.
- Sep 29, 2025FDA-FOIA-2025-11368Form 483· 1 of 3 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 1 section of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.
- 21 CFR 820.25Personnel1 of 1
Observations read from posted 483 PDFs carry no CFR section of FDA’s, so they are not counted here.
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006513362 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006513362. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.