FDA issued this Form 483 to GPC Medical Limited in Loni, India (FEI 3006513362) at the end of a device inspection that ended October 24, 2019. It lists 5 observations. FDA classified the inspection Official Action Indicated (OAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 19, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Oct 24, 2019
Devices · Postmarket Assurance: Devices
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Oct 19, 2026
2,552 days after the inspection
Facility
FEI 3006513362
Loni, India

The 5 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Inadequate process control procedures for cleaning

    Process Control Procedures that describe any process controls necessary to ensure conformance to the specifications have not been adequately established. For example: The Procedure for [illegible] does not identify the temperature that the process is operated at. The Cleaning process validation was performed at a temperature of [illegible]. During a walk through of the manufacturing facility on Tuesday October 22, I observed the (b)(4) Cleaning process operating at a temperature of [illegible]. Also, the procedure for the (b)(4) cleaning does not provide sufficient detail for how parts will be rinsed after [illegible] cleaning such as volume of water used and rinsing time. The Procedure for (b)(4) identifies that the (b)(4) consists of equal parts of [illegible]. The Validation of the process used an (b)(4) containing (b)(4)

  2. 2

    Inadequate in-process acceptance procedures for machining

    Procedures for acceptance of in-process product have not been adequately established. Specifically, during a walk through of the Manufacturing Production area I observed the Machining of 4.0MM Canulated Cancellous Screw / LOT (b)(4). The First Piece Approval document indicated that the thread length of the piece was 7mm and three pieces were measured and were reported as 7.0mm, 7.1mm and 7.1mm. The screw length is 20mm and the thread length should be 8mm. Re-measurement determined that the screw thread length was actually 8mm.

  3. 3

    CAPA procedures not analyzing product nonconformances

    Procedures for Corrective and Preventive Action have not been adequately established. CAPA procedure / QSP / 12 identifies product non conformance (receiving and finished product) as a Quality Data source for CAPA. The firm log book of product non conformance is not analyzed to identify recurring non conformance for Corrective actions. There have been no CAPA actions for product non conformance in the last three years.

  4. 4

    Inadequate validation of cleaning process

    A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated. Specifically, the process validation for cleaning did not demonstrate that the acceptance was satisfied. The testing after criteria for cleaning only reported a (b)(4) (b)(4) difference to [illegible] and the acceptance criteria established that there would be no "neat" water. Also cleaning validation did not define or describe that the product will be rinsed after cleaning in the enzymatic reagent. Further the chemicals of the enzymatic reagent were not tested for as part of the residual testing.

  5. 5

    No maintenance schedule for RO water system

    Schedules for maintenance of equipment have not been adequately established. Specifically, the Reverse (b)(4) Osmosis water system is reportedly serviced by [illegible]. There are no service reports of these activities. The firm has not evaluated the water quality of the RO water used for the cleaning since 2017.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.