Alphatec Spine, Inc.
1950 Camino Vida Roble, Carlsbad, California, United States
FEI 2027467
FDA has 8 inspections on file for Alphatec Spine, Inc. in Carlsbad, California (FEI 2027467), from 2009 to 2022. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Feb 11, 2010. FDA has issued 3 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 8, 2009–2022
- 2 OAI · 3 VAI · 3 NAI
- Inspections with 483 observations shown
- 5 of 8
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesPosted 483 · 1 citation
Read this Form 483 on its own page →Form 483 PDF on fda.gov
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesPosted 483 · 9 observations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.
- Observation 1
MDR malfunction reports not submitted timely
FDA-483 Item 1 is a repeat observation An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur. Specifically, Between 1/2009 and 4/25/2014, the firm has on file 32 complaint reports for the Illico Screw Extender (Part No.'s 73724, 73772 and 73779) for dimples/tabs partially or fully breaking off during the implant surgical procedure, which were not reported to the FDA as MDR malfunction reports. The firm's rationale for not reporting these events as MDR reportable was because there was no associated harm to the patient or delay in the procedure. However, this decision pathway is flawed in that the definition of a malfunction event does not necessitate that patient harm (i.e., serious injury or death) or procedural delay to include surgical intervention must have occurred, rather it necessitates the probability of patient harm occurring or the need to conduct surgical intervention to prevent harm, if the event were to recur. In addition, the firm has on file five (5) additional complaints for screw extender broken tabs, which were reported as MDR serious injury reports due to a delay in the surgical procedure in an attempt to remove the broken off tabs. See Table 1 below for a listing of the 32 complaint reports, not reported as MDR malfunction reports: Table 1 - Listing of Illico Screw Extender Complaints for broken dimples/tabs (1/2009 through 4/25/2014) Complaint No.: | Investigation Results | Report Date | Part No.: CPL-001406 | Tab broken off | 2/20/12 | 73724 CPL-001900 | Tab broken off | 8/14/12 | 73724 CPL-001991 | Both tabs broken off | 9/27/12 | 73724 CPL-002111 | Both tabs broken off | 11/18/12 | 73724 CPL-002137 | Both tabs broken off | 11/27/12 | 73724 CPL-002214 | Both tabs broken off | 1/2/13 | 73724 CPL-002316 | Tab broken off | 1/31/13 | 73724 CPL-002592 | Tab broken off | 5/15/13 | 73724 CPL-002705 | Both tabs broken off | 6/20/13 | 73772 CPL-002734 | Both tabs broken off | 7/8/13 | 73772 CPL-002804 | Tab broken off | 7/22/13 | 73724 CPL-002813 | Tab broken off | 7/26/13 | 73724 CPL-002918 | Both tabs broken off | 9/12/13 | 73772 CPL-002928 | Tab broken off | 9/17/13 | 73724 CPL-002929 | Both tabs broken off | 9/17/13 | 73772 CPL-002960 | Tab broken off | 9/26/13 | 73724 CPL-002991 | Both tabs broken off | 10/9/13 | 73772 CPL-003055 | Tab broken off | 11/20/13 | 73724 CPL-003106 | Tab broken off | 12/18/13 | 73772 CPL-003113 | Tab broken off | 12/19/13 | 73772 CPL-003134 | Both tabs broken off | 1/8/14 | 73724 CPL-003201 | Tab broken off | 2/13/14 | 73724 CPL-003234 | Balls of flaps breaking off | 3/3/14 | 73772 CPL-003293 | Tab broken off | 4/2/14 | 73724 CPL-003294 | Tab broken off | 4/2/14 | 73724 CPL-003295 | Tab broken off | 4/2/14 | 73724 CPL-003298 | Tab broken off | 4/2/14 | 73779 CPL-003310 | Tab broken off | 4/7/14 | 73779 CPL-003312 | Tab broken off | 4/7/14 | 73772 CPL-003313 | Tab broken off | 4/7/14 | 73779 CPL-003368 | Tab broken off | 4/23/14 | 73779 CPL-00369 | Tab broken off | 4/23/14 | 73724
- Observation 2
Design not validated under use conditions
The design was not validated under actual or simulated use conditions. Specifically, originally/per Change Order (CO) No.: 09-196, release date 5/18/2009 through 8/5/2009, the firm implemented the launch of the Illico MIS Screw Extension (SE) System as part of the Design History File (DHF) No.: 07-021 via Validation Report No.: TR-419 - Design Validation and Simulated Use Lab for the Illico SE Posterior Fixation System, Revision A. On 11/17/2011, per CO No.: (b) (4), the firm released Part No.: (b) (4) (b) (4) for use with Part No.: (b) (4) (Illico Reduction Screw Extender). However, the (b) (4) design was not validated in that CO No.: (b) (4) indicates, "Verification and validation is not required. All of the same mechanisms that have already been validated are being used." This statement refers to Validation Report No.: TR-419. Yet, this statement is invalid in that the (b) (4) design/PN: (b) (4) was not part of the design associated with Validation Report No.: TR-419. In addition, post the release of CO No.: (b) (4), the firm's number of complaints for "broken tabs," has increased (i.e., prior to the release of CO No.: (b) (4) the firm had five (5) complaints on file from 1/2009 through 1/2013; post the release of CO No. (b) (4) (the first set built date was (b) (4)) the firm has approximately 28 complaint reports on file ranging from 5/2013 through 4/25/2014, which were MDR reportable events.
Show observations 3 to 9
- Observation 3
Design history file not established
The design history file was not established. Specifically, the firm's (b) (4) (spinal implant device) was acquired from another medical device company in (b) (4). However, the firm has no design history file (DHF) for the (b) (4) device prior to (b) (4) (i.e., Drawing No.: (b) (4), Revision D) and there is only one (1) DHF record post (b) (4) (i.e., Drawing No.: (b) (4) - (b) (4), Revision R, release date 8/3/2009). The (b) (4) family of medical devices was cleared for marketing via multiple 510 (k) submissions to include 2005 and 2007, and there was a product recall for the (b) (4). In addition, in the event the firm needs to conduct a failure investigation, implement a product recall or submit a MDR report for the (b) (4) device, there would be a gap in the investigation/analysis in that the DHF to include device specifications (i.e., design outputs) would be unavailable for review.
- Observation 4
Device history record inadequately maintained
A device history record has not been adequately maintained. Specifically, the firm acquired the (b) (4) (spinal implant device) from another medical device company in (b) (4). However, the only device history records (DHRs) on file for Part No.: (b) (4)-(b) (4), Final Assembly, (b) (4), Revision B/Catalog No.: (b) (4) are dated (b) (4) and (b) (4) and for Revision C of Part No.: (b) (4)/Catalog No.: (b) (4) they are dated (b) (4) through (b) (4). The (b) (4) family of medical devices was cleared for marketing via multiple 510 (k) submissions to include 2005, 2007, 2008 and 2009. On (b) (4), the FDA was notified of manufactured and distributed (b) (4) devices under a product recall. In addition, post the acquisition, the firm is responsible for conducting a failure investigation, implementing a product recall or submitting a MDR report, if necessary for the (b) (4) device. However, the lack of DHR's would result in an inadequate failure investigation in that there would be a gap where the DHR data would be unavailable for review.
- Observation 5
CAPA activities not adequately documented
Corrective and preventive action activities and/or results have not been adequately documented. Specifically, the firm acquired the (b) (4) (spinal implant device) from another medical device company in (b) (4). Previously, on (b) (4), the (b) (4) device was recalled, however, the firm has no records of the recall/corrections and removals on file to include receipt and final disposition of the returned units under recall nor any records associated with what preventive actions were put into place to prevent recurrence of the problem.
- Observation 6
Quality system procedures not established
Quality system procedures and instructions have not been established. Specifically, the firm acquired the (b) (4) (i.e., spinal implant (b) (4)) medical device family from another medical device company in (b) (4). However, the firm has no procedures in place to ensure the quality system records associated with the acquired device to include device history records, drawings, specifications, procedures, etc., were received at the time of the acquisition.
- Observation 7
CAPA procedures inadequately established
Procedures for corrective and preventive action have not been adequately established. Specifically, a) Procedure No.: SOP-025 - Corrective and Preventive Action (CAPA) System, Revision L is not defined in that the procedure does not provide instructions or a linkage to an action or alert level for the (b) (4) complaint trending data, which is reviewed at the (b) (4) trend review meetings. For example, the (b) (4) complaint rate by product line is determined by the "(b) (4)." The determined unit of display is in percentage. However, the firm has not established an alert or action level to meet the established threshold specified within Section 6.1.1.1 of Procedure No.: SOP-025, "(b) (4)" b) Procedure No.: SOP-025 is not defined in that the Frequency Table (frequency of the event occurrence) listed within Section 6.2.1.1 does not provide a definition for the quantitative estimation formula associated with (b) (4); (b) (4); (b) (4); and (b) (4).
- Observation 8
Design validation procedures inadequately established
Procedures for design validation have not been adequately established. Specifically, Procedure No.: PRO-000003 - Risk Management, Revision F is not defined in that Table 2 - (b) (4) (b) (4)
- Observation 9
Acceptance procedures inadequately established
Procedures for acceptance activities have not been adequately established. Specifically, Document/Procedure No.: QCP-042 - Supplied Material Inspection, Revision J is not adequately defined in that Section No.: 6.9.8.1 provides instructions, "(b) (4)." However, the instructions do not define where on the (b) (4) (i.e., beginning, middle, end, etc.) to take the measurements. For example, regarding the receipt of Part No.: (b) (4) Lot No.: (b) (4) on (b) (4), which is the (b) (4) stock for use in the machining/manufacturing of the firm's spinal implant devices, the instructions do not define where on the (b) (4) (i.e., beginning, middle, end, etc.) to take the measurements.
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Training records
Personnel training is not documented.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been established.
Production processes
Production processes were not controlled to ensure that a device conforms to its specifications.
Show 2 more citations
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Submission of Supplement to Individual Report Within One Month
A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: DevicesPosted 483 · 11 citations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Identification of actions needed
Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.
Essential design outputs
Design outputs that are essential for the proper functioning of the device were not completely identified.
Uniform and timely processing
Complaint handling procedures have not been defined to ensure that all complaints are processed in a uniform and timely manner.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been fully validated and approved according to established procedures.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Show 6 more citations
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not implemented.
Type of Reportable Event
An individual medical device manufacturer report submitted per FDA Form 3500A did not indicate in Block H the type of reportable event.
Process control procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications were not defined and implemented.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Documentation
Process validation activities and results have not been fully documented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not implemented.
Analysis of data sources
Appropriate sources of quality data are not adequately analyzed to identify existing and potential causes of nonconforming product and other quality problems.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure that equipment is routinely calibrated were not defined and implemented.
Not approved or obsolete document retrieval
Documents that were obsolete were observed at a location where they could be used.
Calibration documentation
There is incomplete documentation of calibration dates for measurement equipment .
Show 8 more citations
Approval of process equipment
Process equipment has not been approved.
Time to report - 10 days
A report of the required information regarding device correction and removal actions was not sent to FDA within 10 days of initiating the correction or removal.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been fully validated and approved according to established procedures.
Process control procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications were not defined.
Identification of actions needed
Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Environmental control Lack of or inadequate procedures
Appropriate procedures have not been defined for controlling environmental conditions.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities were not implemented.
Regulations FDA cited here
The 15 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 820.70Production and process controls3 of 4
- 21 CFR 820.75Process validation3 of 4
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 4
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 4
- 21 CFR 803.52If I am a manufacturer, what information must I submit in my individual adverse event reports?1 of 4
- 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 4
Show all 15
- 21 CFR 806.10Reports of corrections and removals1 of 4
- 21 CFR 820.20Management responsibility1 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 4
- 21 CFR 820.90Nonconforming product1 of 4
- 21 CFR 820.198Complaint files1 of 4
Observations read from posted 483 PDFs carry no CFR section of FDA’s, so they are not counted here.
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Other FEIs recorded as Alphatec Spine, Inc.
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Alphatec Spine, Inc.FEI 3019878714Memphis, Tennessee, United States
Last inspected 2024-10-29NAI
Alphatec Spine, Inc.FEI 3011277638Hunt Valley, Maryland, United States
Last inspected 2016-03-16NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2027467 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2027467. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.