Alphatec Spine, Inc. Form 483, February 2010

FDA issued this Form 483 to Alphatec Spine, Inc. in Carlsbad, California (FEI 2027467) at the end of a device inspection that ended February 11, 2010. FDA classified the inspection Official Action Indicated (OAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 20, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Feb 11, 2010
Devices · Postmarket Assurance: Devices
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Oct 20, 2026
6,095 days after the inspection
Facility
FEI 2027467
Carlsbad, California

FDA's 11 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Identification of actions needed

    Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.

  2. Essential design outputs

    Design outputs that are essential for the proper functioning of the device were not completely identified.

  3. Uniform and timely processing

    Complaint handling procedures have not been defined to ensure that all complaints are processed in a uniform and timely manner.

  4. Lack of or inadequate process validation

    A process whose results cannot be fully verified by subsequent inspection and test has not been fully validated and approved according to established procedures.

  5. Nonconforming product control

    Products that do not conform to specifications are not adequately controlled.

  6. Lack of or inadequate procedures

    The procedures for implementing corrective and preventive actions were not implemented.

  7. Type of Reportable Event

    An individual medical device manufacturer report submitted per FDA Form 3500A did not indicate in Block H the type of reportable event.

  8. Process control procedures

    Process control procedures that describe any process controls necessary to ensure conformance to specifications were not defined and implemented.

  9. Individual Report of Malfunction

    An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

  10. Report of Death or Serious Injury

    An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

  11. Documentation

    Process validation activities and results have not been fully documented.

After this inspection

FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.

  • Warning LetterJun 21, 2010CDRH · 130 days after this inspection ended

Other Form 483s FDA posted for this facility

Every Form 483 FDA posts under FEI 2027467, newest first.

All inspections at this facility →

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.