Alphatec Spine, Inc. Form 483, March 2015

FDA issued this Form 483 to Alphatec Spine, Inc. in Carlsbad, California (FEI 2027467) at the end of a device inspection that ended March 13, 2015. It lists 9 observations. FDA classified the inspection Voluntary Action Indicated (VAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 20, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Mar 13, 2015
Devices · Postmarket Assurance: Devices
FDA's outcome
VAI
Voluntary Action Indicated
Posted by FDA
Oct 20, 2026
4,239 days after the inspection
Facility
FEI 2027467
Carlsbad, California

The 9 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    MDR malfunction reports not submitted timely

    FDA-483 Item 1 is a repeat observation An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur. Specifically, Between 1/2009 and 4/25/2014, the firm has on file 32 complaint reports for the Illico Screw Extender (Part No.'s 73724, 73772 and 73779) for dimples/tabs partially or fully breaking off during the implant surgical procedure, which were not reported to the FDA as MDR malfunction reports. The firm's rationale for not reporting these events as MDR reportable was because there was no associated harm to the patient or delay in the procedure. However, this decision pathway is flawed in that the definition of a malfunction event does not necessitate that patient harm (i.e., serious injury or death) or procedural delay to include surgical intervention must have occurred, rather it necessitates the probability of patient harm occurring or the need to conduct surgical intervention to prevent harm, if the event were to recur. In addition, the firm has on file five (5) additional complaints for screw extender broken tabs, which were reported as MDR serious injury reports due to a delay in the surgical procedure in an attempt to remove the broken off tabs. See Table 1 below for a listing of the 32 complaint reports, not reported as MDR malfunction reports: Table 1 - Listing of Illico Screw Extender Complaints for broken dimples/tabs (1/2009 through 4/25/2014) Complaint No.: | Investigation Results | Report Date | Part No.: CPL-001406 | Tab broken off | 2/20/12 | 73724 CPL-001900 | Tab broken off | 8/14/12 | 73724 CPL-001991 | Both tabs broken off | 9/27/12 | 73724 CPL-002111 | Both tabs broken off | 11/18/12 | 73724 CPL-002137 | Both tabs broken off | 11/27/12 | 73724 CPL-002214 | Both tabs broken off | 1/2/13 | 73724 CPL-002316 | Tab broken off | 1/31/13 | 73724 CPL-002592 | Tab broken off | 5/15/13 | 73724 CPL-002705 | Both tabs broken off | 6/20/13 | 73772 CPL-002734 | Both tabs broken off | 7/8/13 | 73772 CPL-002804 | Tab broken off | 7/22/13 | 73724 CPL-002813 | Tab broken off | 7/26/13 | 73724 CPL-002918 | Both tabs broken off | 9/12/13 | 73772 CPL-002928 | Tab broken off | 9/17/13 | 73724 CPL-002929 | Both tabs broken off | 9/17/13 | 73772 CPL-002960 | Tab broken off | 9/26/13 | 73724 CPL-002991 | Both tabs broken off | 10/9/13 | 73772 CPL-003055 | Tab broken off | 11/20/13 | 73724 CPL-003106 | Tab broken off | 12/18/13 | 73772 CPL-003113 | Tab broken off | 12/19/13 | 73772 CPL-003134 | Both tabs broken off | 1/8/14 | 73724 CPL-003201 | Tab broken off | 2/13/14 | 73724 CPL-003234 | Balls of flaps breaking off | 3/3/14 | 73772 CPL-003293 | Tab broken off | 4/2/14 | 73724 CPL-003294 | Tab broken off | 4/2/14 | 73724 CPL-003295 | Tab broken off | 4/2/14 | 73724 CPL-003298 | Tab broken off | 4/2/14 | 73779 CPL-003310 | Tab broken off | 4/7/14 | 73779 CPL-003312 | Tab broken off | 4/7/14 | 73772 CPL-003313 | Tab broken off | 4/7/14 | 73779 CPL-003368 | Tab broken off | 4/23/14 | 73779 CPL-00369 | Tab broken off | 4/23/14 | 73724

  2. 2

    Design not validated under use conditions

    The design was not validated under actual or simulated use conditions. Specifically, originally/per Change Order (CO) No.: 09-196, release date 5/18/2009 through 8/5/2009, the firm implemented the launch of the Illico MIS Screw Extension (SE) System as part of the Design History File (DHF) No.: 07-021 via Validation Report No.: TR-419 - Design Validation and Simulated Use Lab for the Illico SE Posterior Fixation System, Revision A. On 11/17/2011, per CO No.: (b) (4), the firm released Part No.: (b) (4) (b) (4) for use with Part No.: (b) (4) (Illico Reduction Screw Extender). However, the (b) (4) design was not validated in that CO No.: (b) (4) indicates, "Verification and validation is not required. All of the same mechanisms that have already been validated are being used." This statement refers to Validation Report No.: TR-419. Yet, this statement is invalid in that the (b) (4) design/PN: (b) (4) was not part of the design associated with Validation Report No.: TR-419. In addition, post the release of CO No.: (b) (4), the firm's number of complaints for "broken tabs," has increased (i.e., prior to the release of CO No.: (b) (4) the firm had five (5) complaints on file from 1/2009 through 1/2013; post the release of CO No. (b) (4) (the first set built date was (b) (4)) the firm has approximately 28 complaint reports on file ranging from 5/2013 through 4/25/2014, which were MDR reportable events.

  3. 3

    Design history file not established

    The design history file was not established. Specifically, the firm's (b) (4) (spinal implant device) was acquired from another medical device company in (b) (4). However, the firm has no design history file (DHF) for the (b) (4) device prior to (b) (4) (i.e., Drawing No.: (b) (4), Revision D) and there is only one (1) DHF record post (b) (4) (i.e., Drawing No.: (b) (4) - (b) (4), Revision R, release date 8/3/2009). The (b) (4) family of medical devices was cleared for marketing via multiple 510 (k) submissions to include 2005 and 2007, and there was a product recall for the (b) (4). In addition, in the event the firm needs to conduct a failure investigation, implement a product recall or submit a MDR report for the (b) (4) device, there would be a gap in the investigation/analysis in that the DHF to include device specifications (i.e., design outputs) would be unavailable for review.

  4. 4

    Device history record inadequately maintained

    A device history record has not been adequately maintained. Specifically, the firm acquired the (b) (4) (spinal implant device) from another medical device company in (b) (4). However, the only device history records (DHRs) on file for Part No.: (b) (4)-(b) (4), Final Assembly, (b) (4), Revision B/Catalog No.: (b) (4) are dated (b) (4) and (b) (4) and for Revision C of Part No.: (b) (4)/Catalog No.: (b) (4) they are dated (b) (4) through (b) (4). The (b) (4) family of medical devices was cleared for marketing via multiple 510 (k) submissions to include 2005, 2007, 2008 and 2009. On (b) (4), the FDA was notified of manufactured and distributed (b) (4) devices under a product recall. In addition, post the acquisition, the firm is responsible for conducting a failure investigation, implementing a product recall or submitting a MDR report, if necessary for the (b) (4) device. However, the lack of DHR's would result in an inadequate failure investigation in that there would be a gap where the DHR data would be unavailable for review.

  5. 5

    CAPA activities not adequately documented

    Corrective and preventive action activities and/or results have not been adequately documented. Specifically, the firm acquired the (b) (4) (spinal implant device) from another medical device company in (b) (4). Previously, on (b) (4), the (b) (4) device was recalled, however, the firm has no records of the recall/corrections and removals on file to include receipt and final disposition of the returned units under recall nor any records associated with what preventive actions were put into place to prevent recurrence of the problem.

  6. 6

    Quality system procedures not established

    Quality system procedures and instructions have not been established. Specifically, the firm acquired the (b) (4) (i.e., spinal implant (b) (4)) medical device family from another medical device company in (b) (4). However, the firm has no procedures in place to ensure the quality system records associated with the acquired device to include device history records, drawings, specifications, procedures, etc., were received at the time of the acquisition.

  7. 7

    CAPA procedures inadequately established

    Procedures for corrective and preventive action have not been adequately established. Specifically, a) Procedure No.: SOP-025 - Corrective and Preventive Action (CAPA) System, Revision L is not defined in that the procedure does not provide instructions or a linkage to an action or alert level for the (b) (4) complaint trending data, which is reviewed at the (b) (4) trend review meetings. For example, the (b) (4) complaint rate by product line is determined by the "(b) (4)." The determined unit of display is in percentage. However, the firm has not established an alert or action level to meet the established threshold specified within Section 6.1.1.1 of Procedure No.: SOP-025, "(b) (4)" b) Procedure No.: SOP-025 is not defined in that the Frequency Table (frequency of the event occurrence) listed within Section 6.2.1.1 does not provide a definition for the quantitative estimation formula associated with (b) (4); (b) (4); (b) (4); and (b) (4).

  8. 8

    Design validation procedures inadequately established

    Procedures for design validation have not been adequately established. Specifically, Procedure No.: PRO-000003 - Risk Management, Revision F is not defined in that Table 2 - (b) (4) (b) (4)

  9. 9

    Acceptance procedures inadequately established

    Procedures for acceptance activities have not been adequately established. Specifically, Document/Procedure No.: QCP-042 - Supplied Material Inspection, Revision J is not adequately defined in that Section No.: 6.9.8.1 provides instructions, "(b) (4)." However, the instructions do not define where on the (b) (4) (i.e., beginning, middle, end, etc.) to take the measurements. For example, regarding the receipt of Part No.: (b) (4) Lot No.: (b) (4) on (b) (4), which is the (b) (4) stock for use in the machining/manufacturing of the firm's spinal implant devices, the instructions do not define where on the (b) (4) (i.e., beginning, middle, end, etc.) to take the measurements.

After this inspection

FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.

  • Warning LetterJul 16, 2015CDRH · 125 days after this inspection ended

Other Form 483s FDA posted for this facility

Every Form 483 FDA posts under FEI 2027467, newest first.

All inspections at this facility →

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.