21 CFR 803.50: If I am a manufacturer, what reporting requirements apply to me?
FDA's inspection data lists 993 Form 483 citations of 21 CFR 803.50 at 714 facilities, from October 2008 to July 24, 2026. 33 warning letters state a violation of it.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 993
- At 897 inspections
- Facilities cited
- 714
- Distinct FEIs
- Most recent citation
- Jul 24, 2026
- King Brand Healthcare Products
- Warning letters
- 33
- State a violation of it
The regulation
The current text of 21 CFR 803.50 from the Electronic Code of Federal Regulations.
§ 803.50 If I am a manufacturer, what reporting requirements apply to me?
(a) If you are a manufacturer, you must report to us the information required by § 803.52 in accordance with the requirements of § 803.12(a), no later than 30 calendar days after the day that you receive or otherwise become aware of information, from any source, that reasonably suggests that a device that you market:
(1) May have caused or contributed to a death or serious injury or
(2) Has malfunctioned and this device or a similar device that you market would be likely to cause or contribute to a death or serious injury, if the malfunction were to recur.
(b) What information does FDA consider “reasonably known” to me?
(1) You must submit all information required in this subpart E that is reasonably known to you. We consider the following information to be reasonably known to you:
(i) Any information that you can obtain by contacting a user facility, importer, or other initial reporter;
(ii) Any information in your possession; or
(iii) Any information that you can obtain by analysis, testing, or other evaluation of the device.
(2) You are responsible for obtaining and submitting to us information that is incomplete or missing from reports submitted by user facilities, importers, and other initial reporters.
(3) You are also responsible for conducting an investigation of each event and evaluating the cause of the event. If you cannot submit complete information on a report, you must provide a statement explaining why this information was incomplete and the steps you took to obtain the information. If you later obtain any required information that was not available at the time you filed your initial report, you must submit this information in a supplemental report under § 803.56 in accordance with the requirements of § 803.12(a).
21 CFR Part 803: Medical Device Reporting, Subpart ERead on eCFR
Citations by year
Form 483 citations of 21 CFR 803.50 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 101 in 2010
| Year | Form 483 citations |
|---|---|
| 2008 | 11 |
| 2009 | 70 |
| 2010 | 101 |
| 2011 | 84 |
| 2012 | 71 |
| 2013 | 79 |
| 2014 | 78 |
| 2015 | 81 |
| 2016 | 59 |
| 2017 | 65 |
| 2018 | 51 |
| 2019 | 43 |
| 2020 | 9 |
| 2021 | 18 |
| 2022 | 36 |
| 2023 | 40 |
| 2024 | 43 |
| 2025 | 45 |
| 2026 | 9 |
How FDA words this citation
FDA's short descriptions and citation text on these 993 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Individual Report of Malfunction550 citations
- Report of Death or Serious Injury402 citations
- Reporting Information Reasonably Known29 citations
- Conducting an Investigation10 citations
- Providing Incomplete or Missing Information2 citations
Citation text
- “An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.”550 citations
- “An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.”402 citations
- “An MDR report submitted to FDA did not include all information that was reasonably known to the manufacturer.”29 citations
- “An MDR report with incomplete information was submitted to FDA without a statement explaining why such information was incomplete and the steps taken to obtain the information.”8 citations
Show 2 more
- “The firm did not conduct an investigation of an event and evaluate the cause of an event.”2 citations
- “The firm did not obtain and provide FDA with information that is incomplete or missing from reports submitted by user facilities, distributors, and other initial reporters.”1 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 803.50, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
Show 4 more
Warning letters stating a violation of 21 CFR 803.50
33 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.
“Failure to report the information required by § 803.52 in accordance with the requirements of § 803.12(a), no later than 30 calendar days after the day that your firm received or otherwise became aware of information, from any source, that reasonably suggests that a device that your firm markets may have caused or contributed to a death or serious injury, as required by 21 CFR 803.50(a)(1).”
- Lion Street Medical, d/b/a Pensar Medical LLCIssued 2026-08-21
“Failure to submit a report to FDA no later than 30 calendar days after the day that your firm receives or otherwise becomes aware of information, from any source, that reasonably suggests that a device that your firm markets has malfunctioned and this device or a similar device that it markets would be likely to cause or contribute to a death or serious injury, if the malfunction were to recur, as required by 21 CFR 803.50(a)(2).”
- ZOLL Medical CorporationIssued 2026-04-30
“Failure to report to us the information required by § 803.52 in accordance with the requirements of § 803.12(a), no later than 30 calendar days after the day that your firm received or otherwise became aware of information, from any source, that reasonably suggests that a device that it markets has malfunctioned and this device or a similar device that your firm markets would be likely to cause or contribute to a death or serious injury, if the malfunction were to recur, as required by 21 CFR 803.50(a)(2).”
- Diasol, Inc.Issued 2026-01-29
“Failure to report the information required by § 803.52 in accordance with the requirements of §803.12(a), no later than 30 calendar days after the day that your firm received or otherwise became aware of information, from any source, that reasonably suggests that a device that it markets has malfunctioned and this device or a similar device that your firm markets would be likely to cause or contribute to a death or serious injury, if the malfunction were to recur, as required by 21 CFR 803.50(a)(2).”
- Unomedical Device S.A. de C.V.Issued 2026-01-08
“Failure to report to us the information required by § 803.52 in accordance with therequirements of § 803.12(a), no later than 30 calendar days after the day that your firm receives or otherwise become aware of information, from any source, that reasonably suggests that a device that your firm markets has malfunctioned and this device or a similar device that it markets would be likely to cause or contribute to a death or serious injury, if the malfunction were to recur, as required by 21 CFR 803.50(a)(2).”
- Meridian Bioscience, IncIssued 2025-12-22
“Failure to report to us the information required by § 803.52 in accordance with the requirements of § 803.12(a), no later than 30 calendar days after the day that your firm received or otherwise became aware of information, from any source, that reasonably suggests that a device that your firm markets may have caused or contributed to a death or serious injury, as required by 21 CFR 803.50(a)(1).”
- MAST Biosurgery USA, Inc.Issued 2025-12-17
Often cited at the same inspection
Of the 897 inspections where FDA cited 21 CFR 803.50, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action487 (54%)
- 21 CFR 820.198Complaint files457 (51%)
- 21 CFR 820.30Design controls348 (39%)
- 21 CFR 820.50Purchasing controls192 (21%)
- 21 CFR 820.75Process validation175 (20%)
- 21 CFR 820.70Production and process controls155 (17%)
Ask FDAgent about 21 CFR 803.50
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
Topics that count this section:Device CAPA and complaints
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 803.50; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.