BioTissue Holdings Inc
8305 Nw 27th St Ste 101, Doral, Florida, United States
FEI 3009809074
FDA has 11 inspections on file for BioTissue Holdings Inc in Doral, Florida (FEI 3009809074), from 2010 to 2026. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 803.50(a)(1).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 11, 2010–2026
- 1 OAI · 4 VAI · 6 NAI
- Inspections with 483 observations shown
- 5 of 11
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 9 requests, latest Sep 2, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
11 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionBiologics · Human Cellular, Tissue, and Gene TherapiesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 8.2.1
Information gathered in feedback process not used as input into risk management
Information gathered in the feedback process does not adequately serve as potential input into risk management for monitoring and maintaining product requirements as well as the product realization or improvement processes.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
- ISO 13485:2016 Clause 8.5.2
Corrective action not taken
Corrective action was not adequately taken to eliminate the cause of nonconformities in order to prevent recurrence.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated where necessary.
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Procedures not adequately established or maintained
Document control procedures have not been adequately maintained.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Blood and Blood Products11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Deviations not reported to FDA
HCT/P deviations relating to core CGTP requirements that occurred in your establishment and at an establishment under contract, agreement, or arrangement with your establishment were not reported to FDA.
Documentation of activities
Documentation of facility cleaning and sanitation activities was not maintained.
Testing, screening, donor eligibility procedures
Procedures for all steps performed in the screening of donor eligibility of HCT/Ps were not established and maintained.
Corrective actions
Appropriate corrective actions were not taken related to the inspections of environmental control systems.
Monitoring--cross contamination, exposure to CD
Environmental conditions are not monitored when such conditions could cause contamination or cross contamination of HCT/Ps or equipment.
Show 6 more citations
Sampling representative (in-process HCT/Ps)
Sampling of in-process HCT/Ps was not representative of the material to be evaluated.
Use prior to verification
Supplies and reagents were used before they were verified to meet specifications designed to prevent the introduction, transmission, or spread of communicable disease.
Cleaned, sanitized per established schedules
Equipment used for manufacturing HCT/Ps was not cleaned, sanitized and maintained according to established schedules.
Validation & approval--established procedures
Processes with results which could not be fully verified by inspection and tests, were not validated and approved according to established procedures.
Risk factors, clinical evidence
Donors were not screened by a review of relevant medical records for risk factors and clinical evidence of communicable disease agents and diseases.
Causing contamination, increasing risks
HCT/Ps were not processed in a way that does not cause contamination or cross contamination during processing and that does not increase the risk of introduction, transmission, or spread of communicable disease.
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene Therapies1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Nine requests to FDA under the Freedom of Information Act named this site, most recently on Sep 2, 2026.
Inspection ending Feb 18, 2026VAI
4 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Apr 23, 2026. FDAgent holds no copy that FDA posted.
- Sep 2, 2026FDA-FOIA-2026-8574Form 483 + EIROpen
- May 19, 2026FDA-FOIA-2026-5148Form 483· 1 of 4 sitesRecords released
- Apr 24, 2026FDA-FOIA-2026-4114Form 483· 1 of 5 sitesRecords released
- Apr 1, 2026FDA-FOIA-2026-3161Form 483Records released
Inspection ending Mar 6, 2024VAI
2 requests. In the inspection history above.
- Aug 31, 2026FDA-FOIA-2026-8436Form 483 + EIROpen
- Jun 7, 2024FDA-FOIA-2024-5009Form 483· 1 of 3 sitesClosed
Inspection ending Jul 16, 2019VAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Apr 7, 2025FDA-FOIA-2025-3221Form 483Closed over fees
- Dec 22, 2023FDA-FOIA-2023-11418Form 483Closed
Inspection date not given
1 request.
- Feb 6, 2019FDA-FOIA-2019-940EIROpen
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 17 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 11 inspections, with how many of them cited each.
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 5
- 21 CFR 820.100Corrective and preventive action2 of 5
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 5
- 21 CFR 820.30Design controls1 of 5
- 21 CFR 820.40[Reserved]1 of 5
- 21 CFR 820.70Production and process controls1 of 5
- 21 CFR 820.90Nonconforming product1 of 5
- 21 CFR 820.198Complaint files1 of 5
Show all 17
- 21 CFR 1271.47What procedures must I establish and maintain?1 of 5
- 21 CFR 1271.75How do I screen a donor?1 of 5
- 21 CFR 1271.190Facilities1 of 5
- 21 CFR 1271.195Environmental control and monitoring1 of 5
- 21 CFR 1271.200Equipment1 of 5
- 21 CFR 1271.210Supplies and reagents1 of 5
- 21 CFR 1271.220Processing and process controls1 of 5
- 21 CFR 1271.230Process validation1 of 5
- 21 CFR 1271.350Reporting1 of 5
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3009809074 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3009809074. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.