21 CFR 1271.220: Processing and process controls
FDA's inspection data lists 28 Form 483 citations of 21 CFR 1271.220 at 23 facilities, from March 2009 to July 30, 2025.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 28
- At 25 inspections
- Facilities cited
- 23
- Distinct FEIs
- Most recent citation
- Jul 30, 2025
- Nelson Laboratories, LLC
- Warning letters
- 0
- State a violation of it
The regulation
The current text of 21 CFR 1271.220 from the Electronic Code of Federal Regulations.
§ 1271.220 Processing and process controls.
(a) General. If you are an establishment that processes HCT/Ps, you must process each HCT/P in a way that does not cause contamination or cross-contamination during processing, and that prevents the introduction, transmission, or spread of communicable disease through the use of the HCT/P.
(b) Pooling. Human cells or tissue from two or more donors must not be pooled (placed in physical contact or mixed in a single receptacle) during manufacturing.
(c) In-process control and testing. You must ensure that specified requirements, consistent with paragraph (a) of this section, for in-process controls are met, and that each in-process HCT/P is controlled until the required inspection and tests or other verification activities have been completed, or necessary approvals are received and documented. Sampling of in-process HCT/Ps must be representative of the material to be evaluated.
(d) Dura mater. (1) When there is a published validated process that reduces the risk of transmissible spongiform encephalopathy, you must use this process for dura mater (or an equivalent process that you have validated), unless following this process adversely affects the clinical utility of the dura mater.
(2) When you use a published validated process, you must verify such a process in your establishment.
21 CFR Part 1271: Human Cells, Tissues, and Cellular and Tissue-Based Products, Subpart DRead on eCFR
Citations by year
Form 483 citations of 21 CFR 1271.220 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 3 in 2009
| Year | Form 483 citations |
|---|---|
| 2009 | 3 |
| 2010 | 2 |
| 2011 | 1 |
| 2012 | 1 |
| 2013 | 3 |
| 2014 | 2 |
| 2016 | 1 |
| 2017 | 2 |
| 2018 | 2 |
| 2019 | 2 |
| 2021 | 1 |
| 2022 | 1 |
| 2023 | 2 |
| 2024 | 2 |
| 2025 | 3 |
How FDA words this citation
FDA's short descriptions and citation text on these 28 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Causing contamination, increasing risks19 citations
- Specified requirements for in-process controls5 citations
- In-process HCT/Ps-control2 citations
- Sampling representative (in-process HCT/Ps)2 citations
Citation text
- “HCT/Ps were not processed in a way that does not cause contamination or cross contamination during processing.”8 citations
- “HCT/Ps were not processed in a way that does not cause contamination or cross contamination during processing and that does not increase the risk of introduction, transmission, or spread of communicable disease.”6 citations
- “HCT/Ps were not processed in a way that does not increase the risk of introduction, transmission, or spread of communicable disease.”5 citations
- “Specified requirements for in-process controls were not met.”5 citations
Show 2 more
- “In-process HCT/Ps were not controlled until required inspections, tests, or other verification activities were completed.”2 citations
- “Sampling of in-process HCT/Ps was not representative of the material to be evaluated.”2 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 1271.220, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Berkeley Advanced Biomaterials, LLC2024-11-22 · FEI 3014298539Causing contamination, increasing risks
Show 4 more
Often cited at the same inspection
Of the 25 inspections where FDA cited 21 CFR 1271.220, the share where it also cited each of these sections.
- 21 CFR 1271.230Process validation13 (52%)
- 21 CFR 1271.195Environmental control and monitoring12 (48%)
- 21 CFR 1271.160Establishment and maintenance of a quality program10 (40%)
- 21 CFR 1271.200Equipment9 (36%)
- 21 CFR 1271.190Facilities8 (32%)
- 21 CFR 1271.270Records8 (32%)
Ask FDAgent about 21 CFR 1271.220
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 1271.220; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.