21 CFR 820.70: Production and process controls
FDA's inspection data lists 2,818 Form 483 citations of 21 CFR 820.70 at 1,925 facilities, from October 2008 to January 16, 2026. 28 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 2,818
- At 2,301 inspections
- Facilities cited
- 1,925
- Distinct FEIs
- Most recent citation
- Jan 16, 2026
- Marson Medical, Inc.
- Warning letters
- 28
- State a violation of it
The regulation
Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).
§ 820.70 Production and process controls.
(a) General. Each manufacturer shall develop, conduct, control, and monitor production processes to ensure that a device conforms to its specifications. Where deviations from device specifications could occur as a result of the manufacturing process, the manufacturer shall establish and maintain process control procedures that describe any process controls necessary to ensure conformance to specifications. Where process controls are needed they shall include:
(1) Documented instructions, standard operating procedures (SOP's), and methods that define and control the manner of production;
(2) Monitoring and control of process parameters and component and device characteristics during production;
(3) Compliance with specified reference standards or codes;
(4) The approval of processes and process equipment; and
(5) Criteria for workmanship which shall be expressed in documented standards or by means of identified and approved representative samples.
(b) Production and process changes. Each manufacturer shall establish and maintain procedures for changes to a specification, method, process, or procedure. Such changes shall be verified or where appropriate validated according to § 820.75, before implementation and these activities shall be documented. Changes shall be approved in accordance with § 820.40.
(c) Environmental control. Where environmental conditions could reasonably be expected to have an adverse effect on product quality, the manufacturer shall establish and maintain procedures to adequately control these environmental conditions. Environmental control system(s) shall be periodically inspected to verify that the system, including necessary equipment, is adequate and functioning properly. These activities shall be documented and reviewed.
(d) Personnel. Each manufacturer shall establish and maintain requirements for the health, cleanliness, personal practices, and clothing of personnel if contact between such personnel and product or environment could reasonably be expected to have an adverse effect on product quality. The manufacturer shall ensure that maintenance and other personnel who are required to work temporarily under special environmental conditions are appropriately trained or supervised by a trained individual.
(e) Contamination control. Each manufacturer shall establish and maintain procedures to prevent contamination of equipment or product by substances that could reasonably be expected to have an adverse effect on product quality.
(f) Buildings. Buildings shall be of suitable design and contain sufficient space to perform necessary operations, prevent mixups, and assure orderly handling.
(g) Equipment. Each manufacturer shall ensure that all equipment used in the manufacturing process meets specified requirements and is appropriately designed, constructed, placed, and installed to facilitate maintenance, adjustment, cleaning, and use.
(1) Maintenance schedule. Each manufacturer shall establish and maintain schedules for the adjustment, cleaning, and other maintenance of equipment to ensure that manufacturing specifications are met. Maintenance activities, including the date and individual(s) performing the maintenance activities, shall be documented.
(2) Inspection. Each manufacturer shall conduct periodic inspections in accordance with established procedures to ensure adherence to applicable equipment maintenance schedules. The inspections, including the date and individual(s) conducting the inspections, shall be documented.
(3) Adjustment. Each manufacturer shall ensure that any inherent limitations or allowable tolerances are visibly posted on or near equipment requiring periodic adjustments or are readily available to personnel performing these adjustments.
(h) Manufacturing material. Where a manufacturing material could reasonably be expected to have an adverse effect on product quality, the manufacturer shall establish and maintain procedures for the use and removal of such manufacturing material to ensure that it is removed or limited to an amount that does not adversely affect the device's quality. The removal or reduction of such manufacturing material shall be documented.
(i) Automated processes. When computers or automated data processing systems are used as part of production or the quality system, the manufacturer shall validate computer software for its intended use according to an established protocol. All software changes shall be validated before approval and issuance. These validation activities and results shall be documented.
21 CFR Part 820: Quality Management System Regulation, Subpart GRead on eCFR
Citations by year
Form 483 citations of 21 CFR 820.70 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 232 in 2010
| Year | Form 483 citations |
|---|---|
| 2008 | 38 |
| 2009 | 184 |
| 2010 | 232 |
| 2011 | 230 |
| 2012 | 232 |
| 2013 | 220 |
| 2014 | 225 |
| 2015 | 196 |
| 2016 | 177 |
| 2017 | 208 |
| 2018 | 179 |
| 2019 | 159 |
| 2020 | 38 |
| 2021 | 48 |
| 2022 | 91 |
| 2023 | 106 |
| 2024 | 119 |
| 2025 | 130 |
| 2026 | 6 |
How FDA words this citation
FDA's short descriptions and citation text on these 2,818 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Process control procedures, Lack of or inadequate procedures599 citations
- Environmental control Lack of or inadequate procedures504 citations
- Software validation for automated processes308 citations
- Production processes281 citations
- Maintenance schedule, Lack of or inadequate schedule240 citations
Show 3 more
- Production and Process Change Procedures, lack of or Inad.225 citations
- Contamination control, Lack of or inadequate procedures149 citations
- Documentation of software validation92 citations
Citation text
- “Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.”423 citations
- “Procedures to control environmental conditions have not been adequately established.”355 citations
- “Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.”167 citations
- “Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.”156 citations
Show 2 more
- “Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.”103 citations
- “Procedures to control environmental conditions have not been established.”102 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 820.70, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Marson Medical, Inc.2026-01-16 · FEI 1000550978Process control procedures, Lack of or inadequate procedures
- HG Medical Gmbh2026-01-09 · FEI 3006272282Production and Process Change Procedures, lack of or Inad.
- J Sterling Industries, LLC2026-01-08 · FEI 3008879776Environmental control Lack of or inadequate procedures
- DiAgam Societe Anonyme (S.A.)2025-12-18 · FEI 3003749380Equipment Installation, Placement, Specified Requirements
- MEDTRONIC GRAND RAPIDS2025-12-15 · FEI 3003955307Environmental control Lack of or inadequate procedures
Show 4 more
- MedLine Industries, LP (NAMIC Division)2025-12-12 · FEI 3015910259Maintenance schedule, Lack of or inadequate schedule
- Specialty Health Products CO, LLC2025-12-11 · FEI 2027347Process control procedures, Lack of or inadequate procedures
- Next Medical Products Company, LLC2025-12-10 · FEI 2523891Environmental control Lack of or inadequate procedures
What the observations say
Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 820.70 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 2 observations tie this way. The highlighted sentence is the matched wording.
Synergy Health Ede BV2023-10-12 · Observation 1
Procedures for changes to a process have not been adequately established. Specifically, the process treatment [illegible] factor for xray sterilization in the Venlo facility was changed from [illegible] reportedly on September 11, 2023. The change was not documented and approved in the change control system.
This observation opens with FDA's standard wording for 21 CFR 820.70(b).
Legacy Pharmaceutical Packaging LLC2020-01-16 · Observation 4
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established. Specifically, maintenance activities, including the reason (problem), maintenance performed, date and individual(s) performing the maintenance activities, were not documented when maintenance was accomplished on the vision system for performance qualification lot numberi (5) (4).
This observation opens with FDA's standard wording for 21 CFR 820.70(g)(1).
Warning letters stating a violation of 21 CFR 820.70
28 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.
“Failure to establish and maintain procedures where environmental conditions that could reasonably be expected to have an adverse effect on product quality, as required by 21 CFR 820.70(c).”
- Next Medical Products Company, LLCIssued 2026-08-19
“Failure to develop, conduct, control, and monitor production processes to ensure that a device conforms to its specifications, as required by 21 CFR 820.70(a).”
- Specialty Health Products CO, LLCIssued 2026-07-07
“Failure to establish and maintain schedules for the adjustment, cleaning, and other maintenance of equipment to ensure that manufacturing specifications are met, as required by 21 CFR 820.70(g)(1).”
- Medline Industries, LPIssued 2026-03-25
“Failure to adequately establish process control procedures that describe any process controls necessary to ensure conformance to specifications, as required by 21 CFR § 820.70(a).”
- Diasol, Inc.Issued 2026-01-29
“Failure to adequately establish procedures to prevent contamination or product by substances that may have an adverse effect on product quality, as required by 21 CFR 820.70(e).”
- MAST Biosurgery USA, Inc.Issued 2025-12-17
“Failure to establish and maintain procedures to prevent contamination of equipment or product by substances that could reasonably be expected to have an adverse effect on product quality as required by 21 CFR 820.70(e).”
- Qianjiang Kingphar Medical Material Co Ltd.Issued 2025-10-03
Often cited at the same inspection
Of the 2,301 inspections where FDA cited 21 CFR 820.70, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action1,223 (53%)
- 21 CFR 820.198Complaint files931 (40%)
- 21 CFR 820.75Process validation736 (32%)
- 21 CFR 820.30Design controls696 (30%)
- 21 CFR 820.50Purchasing controls622 (27%)
- 21 CFR 820.90Nonconforming product613 (27%)
Ask FDAgent about 21 CFR 820.70
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.70; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.