Skip to content

DiAgam Societe Anonyme (S.A.)

Rue Du Parc Industriel 40, Ath, Belgium

FEI 3003749380

FDA has 3 inspections on file for DiAgam Societe Anonyme (S.A.) in Ath, Belgium (FEI 3003749380), from 2020 to 2025. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA posted the Form 483 from that inspection. FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
3, 2020–2025
1 OAI · 1 VAI · 1 NAI
Inspections with 483 observations shown
2 of 3
Form 483s posted by FDA
↗
Requested under FOIA
1 request, latest Apr 2, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesPosted 483 · 4 citations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate In-process acceptance procedures

      Procedures for the acceptance and control of in-process product have not been established.

    2. Equipment Installation, Placement, Specified Requirements

      The appropriate design, construction, placement, and installation of manufacturing equipment have not been ensured and equipment used in the manufacturing process does not meet specified requirements.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    4. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR documentation of label release {see also 820.184}

      The DHR does not include records of examination and release of device labeling, including date and signature of the examiner.

    2. Document change records, maintained.

      Records of changes to documents were not maintained.

    3. Documented evaluation

      The evaluation of potential suppliers was not documented.

    4. Design validation - documentation

      The results of design validation, including identification of the design, method(s), the date and the individual(s) performing validation, were not documented in the design history file.

    5. Design transfer - Lack of or inadequate procedures

      Procedures for design transfer have not been established.

    Show 1 more citation
    1. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Apr 2, 2026. It asked for a Form 483.

  • Inspection date not given

    1 request. FDA released records in answer to a request for its Form 483 on Jun 23, 2026. FDAgent holds no copy that FDA posted.

    1. Apr 2, 2026FDA-FOIA-2026-3325Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003749380 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003749380. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.