21 CFR 820.75: Process validation
FDA's inspection data lists 2,444 Form 483 citations of 21 CFR 820.75 at 1,700 facilities, from October 2008 to January 16, 2026. 47 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 2,444
- At 2,169 inspections
- Facilities cited
- 1,700
- Distinct FEIs
- Most recent citation
- Jan 16, 2026
- Marson Medical, Inc.
- Warning letters
- 47
- State a violation of it
The regulation
Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).
§ 820.75 Process validation.
(a) Where the results of a process cannot be fully verified by subsequent inspection and test, the process shall be validated with a high degree of assurance and approved according to established procedures. The validation activities and results, including the date and signature of the individual(s) approving the validation and where appropriate the major equipment validated, shall be documented.
(b) Each manufacturer shall establish and maintain procedures for monitoring and control of process parameters for validated processes to ensure that the specified requirements continue to be met.
(1) Each manufacturer shall ensure that validated processes are performed by qualified individual(s).
(2) For validated processes, the monitoring and control methods and data, the date performed, and, where appropriate, the individual(s) performing the process or the major equipment used shall be documented.
(c) When changes or process deviations occur, the manufacturer shall review and evaluate the process and perform revalidation where appropriate. These activities shall be documented.
21 CFR Part 820: Quality Management System Regulation, Subpart GRead on eCFR
Citations by year
Form 483 citations of 21 CFR 820.75 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 204 in 2015
| Year | Form 483 citations |
|---|---|
| 2008 | 21 |
| 2009 | 145 |
| 2010 | 168 |
| 2011 | 190 |
| 2012 | 184 |
| 2013 | 177 |
| 2014 | 200 |
| 2015 | 204 |
| 2016 | 140 |
| 2017 | 192 |
| 2018 | 147 |
| 2019 | 149 |
| 2020 | 32 |
| 2021 | 63 |
| 2022 | 86 |
| 2023 | 98 |
| 2024 | 115 |
| 2025 | 131 |
| 2026 | 2 |
How FDA words this citation
FDA's short descriptions and citation text on these 2,444 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Lack of or inadequate process validation1,453 citations
- Documentation426 citations
- Lack/Inad procedure-Monitoring/Control of Validated Proces265 citations
- Process changes - review, evaluation and revalidation143 citations
- Documentation of validated process performance89 citations
Show 3 more
- Documentation - review in response to changes or deviations62 citations
- Documentation - specific items5 citations
- Performed by qualified individual1 citations
Citation text
- “A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.”862 citations
- “A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.”468 citations
- “Procedures for monitoring and control of process parameters for a validated process have not been adequately established.”182 citations
- “A validated process was not revalidated when changes or process deviations occurred.”76 citations
Show 2 more
- “Process validation activities and results have not been adequately documented.”68 citations
- “Process validation activities have not been adequately documented.”68 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 820.75, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Ashland Specialty Ingredients, G.P.2025-12-22 · FEI 3010785603Lack of or inadequate process validation
- BioMed Laboratories, LLC.2025-12-19 · FEI 3006784948Process changes - review, evaluation and revalidation
- MedLine Industries, LP (NAMIC Division)2025-12-12 · FEI 3015910259Lack of or inadequate process validation
Show 4 more
- Creative Foam Corporation - Bremen2025-11-24 · FEI 3001814065Lack/Inad procedure-Monitoring/Control of Validated Proces
- OLYMPUS MEDICAL SYSTEMS CORP. Hinode Plant2025-11-21 · FEI 3003637092Lack of or inadequate process validation
What the observations say
Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 820.75 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 5 observations tie this way. The highlighted sentence is the matched wording.
Implantica CE Reflux Ltd2025-10-24 · Observation 4
Process validation activities have not been adequately documented. Specifically, the stated purpose of the validation protocol (test instructions) RS DVP0004 "RefluxStop Radiopacity and Barium Sulphate," was not congruent with the conclusions in the corresponding output report RS DVR0004. The protocol documentation was inaccurate.
This observation opens with FDA's standard wording for 21 CFR 820.75(a).
Foremount Enterprise Co., Ltd.2025-06-19 · Observation 4
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures. Specifically, your firm's Validation Management Procedure, QP732, V.05, dated March 26, 2025, and Validation Work Instruction, WI7312, V.02, dated March 27, 2025, have not been adequately established. Your firm is a Manufacturer and Contract Manufacturer of various medical devices to include …
This observation opens with FDA's standard wording for 21 CFR 820.75(a).
Synergy Health Ede BV2023-10-12 · Observation 2
Procedures for monitoring and control of process parameters for a validated process have not been adequately established. Specifically, the conveyor speed in the Xray treatment area has not been adequately defined. Process monitoring is performed at [illegible] According to the operating technician the system allows speed settings between 0 and 4.5 m/minute. The described working range (undocumented) for the conveyor speed is between [illegible]
This observation opens with FDA's standard wording for 21 CFR 820.75(b).
Show 2 more
Legacy Pharmaceutical Packaging LLC2020-01-16 · Observation 1
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures. Specifically, the following was noted from review of the Final Report for Performance Qualification for Packaging 1 (5) (4), dated on or about 12/23/19: A. The process control verification protocol (QP 19-416, -417 and -418, part 8.0) documents the Vision System Challenge tests for …
This observation opens with FDA's standard wording for 21 CFR 820.75(a).
Sun Pharmaceutical Industries Ltd.2019-06-11 · Observation 4
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures. Specifically, the requalification of your [illegible] under Protocol No.: QUA-S/0253 Requalification Protocol/Report for [illegible] (equipment ID: 240) does not demonstrate that the air supply, assembly air, and clamp air pressures required per your reference document were …
This observation opens with FDA's standard wording for 21 CFR 820.75(a).
Warning letters stating a violation of 21 CFR 820.75
47 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.
“Failure to adequately ensure that when the results of a process cannot be fully verified by subsequent inspection and test that the process shall be validated with a high degree of assurance and approved according to established procedure, as required by 21 CFR 820.75(a).”
Each letter below states the violation in these words.
- Olympus Medical Systems CorporationIssued 2026-10-07
- Flextronics America LLCIssued 2026-01-27
“Failure to revalidate a validated process when changes or process deviations occurred, as required by 21 CFR 820.75(c).”
- IsoTis OrthoBiologics, Inc.Issued 2026-02-24
“Failure to adequately validate a process whose results cannot be fully verified by subsequent inspection and test according to established procedures, as required by 21 CFR § 820.75(a).”
- Diasol, Inc.Issued 2026-01-29
“Failure to validate a process with a high degree of assurance and approved according to established procedures, when the results of a process cannot be fully verified by subsequent inspection and test. The validation activities and results, including the date and signature of the individual(s) approving the validation and where appropriate the major equipment validated, shall be documented, as required by 21 CFR 820.75(a).”
- Meridian Bioscience, IncIssued 2025-12-22
“Failure to adequately ensure that when the results of a process cannot be fully verified by subsequent inspection and test, the process shall be validated with a high degree of assurance and approved according to established procedures, as required by 21 CFR 820.75(a).”
- Physitemp Instruments, LLCIssued 2025-11-12
Often cited at the same inspection
Of the 2,169 inspections where FDA cited 21 CFR 820.75, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action1,158 (53%)
- 21 CFR 820.198Complaint files810 (37%)
- 21 CFR 820.70Production and process controls736 (34%)
- 21 CFR 820.30Design controls727 (34%)
- 21 CFR 820.50Purchasing controls599 (28%)
- 21 CFR 820.90Nonconforming product594 (27%)
Ask FDAgent about 21 CFR 820.75
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
Topics that count this section:Process validation
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.75; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.