OLYMPUS MEDICAL SYSTEMS CORP. Hinode Plant
34-3, Oazahirai; Hinode-Machi, Nishitama-Gun, Japan
FEI 3003637092
FDA has 4 inspections on file for OLYMPUS MEDICAL SYSTEMS CORP. Hinode Plant in Nishitama-Gun, Japan (FEI 3003637092), from 2010 to 2025. The latest ended on and was classified OAI, official action indicated. FDA has issued 2 warning letters under this FEI, the latest on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.
FDA’s Import Alert 89-04 Red List names this firm at this address. FDA printed no FEI on the entry. No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 4, 2010–2025
- 2 OAI · 1 VAI · 1 NAI
- Inspections with 483 observations shown
- 1 of 4
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest May 1, 2026
- Import refusals
- None under this FEI
- Import alerts
- 89-04Red List
- Matched on FDA’s firm name and street address
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Process changes - review, evaluation and revalidation
A validated process was not revalidated when changes or process deviations occurred.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on May 1, 2026.
Inspection ending Nov 21, 2025OAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jun 18, 2026. FDAgent holds no copy that FDA posted.
- Jan 14, 2026FDA-FOIA-2026-497Form 483· 1 of 2 sitesRecords released
Inspection date not given
1 request. None of FDA’s log entries for these requests say records were released.
- May 1, 2026FDA-FOIA-2026-4360EIR· 1 of 8 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 3 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 4 inspections.
- 21 CFR 820.75Process validation1 of 1
- 21 CFR 820.90Nonconforming product1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov
Date published 2026-10-19
- FDA’s entry
- OLYMPUS MEDICAL SYSTEMS CORP. Hinode Plant · 34-3, Oazahirai; Hinode-Machi, Nishitama-Gun, Tokyo JAPAN
- FDA’s inspection record for FEI 3003637092
- OLYMPUS MEDICAL SYSTEMS CORP. Hinode Plant · 34-3, Oazahirai; Hinode-Machi, Nishitama-Gun, Japan
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
3 products
- 77 E - - OBNasopharyngoscope (Flexible Or Rigid)
- 77 E - - OQBronchoscope (Flexible Or Rigid)
- 78 N - - EZSystem, Imaging, Gastrointestinal, Wireless, Capsule
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003637092 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003637092. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.