21 CFR 820.120: Device labeling
FDA's inspection data lists 463 Form 483 citations of 21 CFR 820.120 at 400 facilities, from October 2008 to December 11, 2025. 6 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 463
- At 432 inspections
- Facilities cited
- 400
- Distinct FEIs
- Most recent citation
- Dec 11, 2025
- Specialty Health Products CO, LLC
- Warning letters
- 6
- State a violation of it
The regulation
Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).
§ 820.120 Device labeling.
Each manufacturer shall establish and maintain procedures to control labeling activities.
(a) Label integrity. Labels shall be printed and applied so as to remain legible and affixed during the customary conditions of processing, storage, handling, distribution, and where appropriate use.
(b) Labeling inspection. Labeling shall not be released for storage or use until a designated individual(s) has examined the labeling for accuracy including, where applicable, the correct unique device identifier (UDI) or universal product code (UPC), expiration date, control number, storage instructions, handling instructions, and any additional processing instructions. The release, including the date and signature of the individual(s) performing the examination, shall be documented in the DHR.
(c) Labeling storage. Each manufacturer shall store labeling in a manner that provides proper identification and is designed to prevent mixups.
(d) Labeling operations. Each manufacturer shall control labeling and packaging operations to prevent labeling mixups. The label and labeling used for each production unit, lot, or batch shall be documented in the DHR.
(e) Control number. Where a control number is required by § 820.65, that control number shall be on or shall accompany the device through distribution.
21 CFR Part 820: Quality Management System Regulation, Subpart KRead on eCFR
Citations by year
Form 483 citations of 21 CFR 820.120 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 49 in 2012
| Year | Form 483 citations |
|---|---|
| 2008 | 7 |
| 2009 | 26 |
| 2010 | 29 |
| 2011 | 28 |
| 2012 | 49 |
| 2013 | 39 |
| 2014 | 36 |
| 2015 | 24 |
| 2016 | 21 |
| 2017 | 29 |
| 2018 | 42 |
| 2019 | 31 |
| 2020 | 5 |
| 2021 | 17 |
| 2022 | 29 |
| 2023 | 16 |
| 2024 | 20 |
| 2025 | 15 |
How FDA words this citation
FDA's short descriptions and citation text on these 463 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Lack of or inadequate procedures for labeling248 citations
- Records, DHR {see also 820.184(e)}70 citations
- DHR documentation of label release {see also 820.184}60 citations
- Examination for accuracy34 citations
- Labeling inspection - UDI24 citations
Show 3 more
- Mixups13 citations
- Remain legible7 citations
- Control number {see also 820.65, 820.184(f)}5 citations
Citation text
- “Procedures to control labeling activities have not been adequately established.”114 citations
- “Procedures to control labeling activities have not been established.”109 citations
- “Labels and labeling used for each finished product, lot, or batch, were not sufficiently documented in the DHR.”65 citations
- “The DHR does not include records of examination and release of device labeling, including date and signature of the examiner.”43 citations
Show 2 more
- “Labeling was not sufficiently examined by a designated individual for accuracy including unique device identifier (UDI) or universal product code (UPC).”24 citations
- “The DHR does not include complete records of examination and release of device labeling, including date and signature of the examiner.”13 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 820.120, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Specialty Health Products CO, LLC2025-12-11 · FEI 2027347Lack of or inadequate procedures for labeling
Show 4 more
- Atmos Medizintechnik Gmbh & Co. KG2025-01-16 · FEI 3003169291DHR documentation of label release {see also 820.184}
Warning letters stating a violation of 21 CFR 820.120
6 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.
“Failure to adequately establish procedures for the control of labeling operations to ensure that the label and labeling used for each production unit, lot, or batch is documented in the Device History Record (DHR), as required under 21 CFR 820.120(d).”
- Leltek Inc.Issued 2025-07-17
“Failure to establish and maintain adequate procedures to control labeling activities, as required by 21 CFR 820.120.”
- DFI Co., Ltd.Issued 2025-05-06
“Failure to establish procedures to control labeling activities, as required by 21 CFR 820.120.”
- Robbins Instruments, LLCIssued 2025-01-21
“Failure to establish and maintain acceptance procedures to control labeling activities, as required by 21 CFR 820.120.”
- Rolence Ent. Inc.Issued 2024-10-18
“Failure to establish and maintain adequate procedures to control labeling activities, including ensuring that labeling shall not be released for storage or use until a designated individual(s) has examined the labeling for accuracy including, where applicable, among other things, the correct unique device identifier (UDI) or universal product code (UPC), as required by 21 CFR 820.120(b).”
- Caltag Medsystems LtdIssued 2023-08-31
“Failure to establish and maintain procedures to control labeling activities as required by 21 CFR 820.120.”
- Forcemech International LLCIssued 2022-09-21
Often cited at the same inspection
Of the 432 inspections where FDA cited 21 CFR 820.120, the share where it also cited each of these sections.
- 21 CFR 820.100Corrective and preventive action212 (49%)
- 21 CFR 820.198Complaint files206 (48%)
- 21 CFR 820.80Receiving, in-process, and finished device acceptance171 (40%)
- 21 CFR 820.30Design controls166 (38%)
- 21 CFR 820.50Purchasing controls153 (35%)
- 21 CFR 820.184Device history record150 (35%)
Ask FDAgent about 21 CFR 820.120
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.120; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.