Specialty Health Products CO, LLC
2514 E Mohawk Ln Ste 111, Phoenix, Arizona, United States
FEI 2027347
FDA has 4 inspections on file for Specialty Health Products CO, LLC in Phoenix, Arizona (FEI 2027347), from 2011 to 2025. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 15 citations from this inspection: 14 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 801, the device labeling rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 4, 2011–2025
- 1 OAI · 0 VAI · 3 NAI
- Inspections with 483 observations shown
- 1 of 4
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Feb 24, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices15 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.
Show 10 more citations
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Lack of or inadequate procedures for labeling
Procedures to control labeling activities have not been adequately established.
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been maintained.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Feb 24, 2026. It asked for a Form 483.
Inspection ending Dec 11, 2025OAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 24, 2026. FDAgent holds no copy that FDA posted.
- Feb 24, 2026FDA-FOIA-2026-1914Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 4 inspections.
- 21 CFR 801.20Label to bear a unique device identifier1 of 1
- 21 CFR 820.22Quality audit1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.40[Reserved]1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.70Production and process controls1 of 1
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
Show all 14
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.120Device labeling1 of 1
- 21 CFR 820.181Device master record1 of 1
- 21 CFR 820.184Device history record1 of 1
- 21 CFR 820.198Complaint files1 of 1
- 21 CFR 820.200Servicing1 of 1
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2027347 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2027347. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.