Skip to content

21 CFR 820.200: Servicing

FDA's inspection data lists 330 Form 483 citations of 21 CFR 820.200 at 282 facilities, from October 2008 to December 11, 2025. 3 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
330
At 314 inspections
Facilities cited
282
Distinct FEIs
Most recent citation
Dec 11, 2025
Specialty Health Products CO, LLC
Warning letters
3
State a violation of it

The regulation

Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).

§ 820.200 Servicing.

(a) Where servicing is a specified requirement, each manufacturer shall establish and maintain instructions and procedures for performing and verifying that the servicing meets the specified requirements.

(b) Each manufacturer shall analyze service reports with appropriate statistical methodology in accordance with § 820.100.

(c) Each manufacturer who receives a service report that represents an event which must be reported to FDA under part 803 of this chapter shall automatically consider the report a complaint and shall process it in accordance with the requirements of § 820.198.

(d) Service reports shall be documented and shall include:

(1) The name of the device serviced;

(2) Any unique device identifier (UDI) or universal product code (UPC), and any other device identification(s) and control number(s) used;

(3) The date of service;

(4) The individual(s) servicing the device;

(5) The service performed; and

(6) The test and inspection data.

21 CFR Part 820: Quality Management System Regulation, Subpart NRead on eCFR

Citations by year

Form 483 citations of 21 CFR 820.200 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 36 in 2011

Form 483 citations of 21 CFR 820.200 by inspection year
YearForm 483 citations
20088
200918
201016
201136
201231
201324
201416
201516
201617
201729
201831
201916
20207
202112
202221
202312
202410
202510

How FDA words this citation

FDA's short descriptions and citation text on these 330 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Servicing - Lack of or inadequate procedures197 citations
  • Service reports71 citations
  • Analyzing service report31 citations
  • Service reports/MDRs/complaints14 citations
  • Test and inspection data8 citations
Show 3 more
  • No Service reports5 citations
  • Service date1 citations
  • Service performed1 citations

Citation text

  • “Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.”65 citations
  • “Service reports do not include the required information.”51 citations
  • “Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been established.”40 citations
  • “Procedures or instructions for performing servicing activities have not been adequately established.”32 citations
Show 2 more
  • “Service reports were not analyzed following appropriate statistical methods.”31 citations
  • “Procedures or instructions for performing servicing activities have not been established.”18 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 820.200, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

  • Specialty Health Products CO, LLC2025-12-11 · FEI 2027347Servicing - Lack of or inadequate procedures
  • Medidenta2025-11-05 · FEI 1000113887Servicing - Lack of or inadequate procedures
  • Drake Precision Dental Laboratory2025-11-05 · FEI 3002706372Servicing - Lack of or inadequate procedures
  • Dane Technologies2025-08-25 · FEI 3003857817Servicing - Lack of or inadequate procedures
  • ATZ Manufacturing2025-07-24 · FEI 3014653196Servicing - Lack of or inadequate procedures
  • Horizon Scientific2025-04-18 · FEI 3006457401Servicing - Lack of or inadequate procedures
Show 4 more

Warning letters stating a violation of 21 CFR 820.200

3 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.

“Failure to adequately establish procedures for performing servicing activities and verifying that servicing meets specified requirements, as required by 21 CFR 820.200(a).”

Each letter below states the violation in these words.

“Failure to establish and maintain instructions and procedures for performing and verifying that the servicing meets the specified requirements, as required by 21 CFR 820.200.”

Often cited at the same inspection

Of the 314 inspections where FDA cited 21 CFR 820.200, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 820.200

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.200; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.