Skip to content

Drake Precision Dental Laboratory

8510 Crown Crescent Ct, Charlotte, North Carolina, United States

FEI 3002706372

FDA has 3 inspections on file for Drake Precision Dental Laboratory in Charlotte, North Carolina (FEI 3002706372), from 2017 to 2025. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 8 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.184 and 21 CFR 820.72(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
3, 2017–2025
0 OAI · 0 VAI · 3 NAI
Inspections with 483 observations shown
2 of 3
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Mar 5, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Evaluation, timeliness, identification

      Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.

    2. Rationale documented for no investigation

      Records for complaints where no investigation was made do not include required information.

    3. Management representative

      No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.

    4. Servicing - Lack of or inadequate procedures

      Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been established.

    5. Equipment control activity documentation

      Equipment calibrations have not been documented.

    Show 3 more citations
    1. Documentation of validated process performance

      There is no documentation of monitoring and control methods and data and the major equipment used for a validated process.

    2. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.

    3. Documentation

      Corrective and preventive action activities and/or results have not been documented.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Document review, approval by designated individual

      Documents were not reviewed and not approved by designated individual(s) prior to issuance .

    2. Equipment control activity documentation

      Equipment calibrations and checks have not been documented.

    3. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

    4. DMR - not or inadequately maintained

      A device master record has not been maintained.

    5. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    Show 1 more citation
    1. Documentation

      Process validation activities and results have not been documented.

  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Mar 5, 2026. It asked for a Form 483.

  • Inspection ending Nov 5, 2025NAI

    1 request. In the inspection history above.

    1. Mar 5, 2026FDA-FOIA-2026-2256Form 483· 1 of 5 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.

Show all 10

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002706372 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002706372. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.