Form 483 citations
893
At 884 inspections
Facilities cited
769
Distinct FEIs
Most recent citation
Mar 3, 2026
Medivis, Inc.
Warning letters
10
State a violation of it

The regulation

Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).

§ 820.181 Device master record.

Each manufacturer shall maintain device master records (DMR's). Each manufacturer shall ensure that each DMR is prepared and approved in accordance with § 820.40. The DMR for each type of device shall include, or refer to the location of, the following information:

(a) Device specifications including appropriate drawings, composition, formulation, component specifications, and software specifications;

(b) Production process specifications including the appropriate equipment specifications, production methods, production procedures, and production environment specifications;

(c) Quality assurance procedures and specifications including acceptance criteria and the quality assurance equipment to be used;

(d) Packaging and labeling specifications, including methods and processes used; and

(e) Installation, maintenance, and servicing procedures and methods.

21 CFR Part 820: Quality Management System Regulation, Subpart MRead on eCFR

Citations by year

Form 483 citations of 21 CFR 820.181 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 111 in 2011

Form 483 citations of 21 CFR 820.181 by inspection year
YearForm 483 citations
200811
200954
201062
2011111
201285
201373
201473
201563
201644
201766
201850
201958
202010
202117
202231
202332
202426
202525
20262

How FDA words this citation

FDA's short descriptions and citation text on these 893 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • DMR - not or inadequately maintained795 citations
  • DMR device specifications64 citations
  • DMR QA procedures and specifications17 citations
  • DMR production process specifications14 citations
  • DMR packaging and labeling specifications3 citations

Citation text

  • “A device master record has not been adequately maintained.”445 citations
  • “A device master record has not been maintained.”347 citations
  • “The device master record does not include or refer to the location of all specifications.”28 citations
  • “The device master record does not include or refer to the location of device software specifications.”16 citations
Show 2 more
  • “The device master record does not include or refer to the location of all production and process specifications.”12 citations
  • “The device master record does not include or refer to the location of all quality assurance procedures and specifications.”10 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 820.181, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

  • Medivis, Inc.2026-03-03 · FEI 3015536273DMR - not or inadequately maintained
  • ZIIP INC2026-01-12 · FEI 3029740697DMR - not or inadequately maintained
  • Specialty Health Products CO, LLC2025-12-11 · FEI 2027347DMR - not or inadequately maintained
  • ASEPTIKITS, LLC2025-12-03 · FEI 3015422213DMR - not or inadequately maintained
  • Hakko Co., Ltd.2025-11-13 · FEI 3002807220DMR - not or inadequately maintained
  • Lightwave2025-10-22 · FEI 3009241253DMR - not or inadequately maintained
Show 4 more
  • Tuxedo Imaging2025-09-19 · FEI 3021736604DMR - not or inadequately maintained
  • Exactech, Inc.2025-08-27 · FEI 1038671DMR - not or inadequately maintained
  • Pharmacaribe, LLC2025-08-20 · FEI 3006500694DMR - not or inadequately maintained
  • Diasol, Inc2025-08-19 · FEI 3007137482DMR - not or inadequately maintained

What the observations say

Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 820.181 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 1 observation ties this way. The highlighted sentence is the matched wording.

  • LATE NIGHT CONCEPTS INC. (DBA BRUX NIGHT GUARD)2021-10-28 · Observation 3

    A device master record has not been maintained. Specifically, a Device Master Record has not been maintained for Brux Night Guard.

    This observation opens with FDA's standard wording for 21 CFR 820.181.

Warning letters stating a violation of 21 CFR 820.181

10 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.

“Failure to maintain device master records (DMR's), as required by 21 CFR 820.181.”
“Failure to establish and maintain device master records, as required by 21 CFR 820.181.”
“Failure to maintain device master records as required by 21 CFR 820.181.”
“Failure to adequately maintain device master records (DMR's), as required by 21 CFR820.181”
“Failure to maintain Device Master Records (DMR’s), as required by 21 CFR 820.181.”
“Failure to maintain a device master record, as required by 21 CFR 820.181.”

Often cited at the same inspection

Of the 884 inspections where FDA cited 21 CFR 820.181, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 820.181

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.181; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.