Diasol, Inc
310 S 43rd Ave Ste B, Phoenix, Arizona, United States
FEI 3007137482
FDA has 12 inspections on file for Diasol, Inc in Phoenix, Arizona (FEI 3007137482), from 2009 to 2025. The latest ended on and was classified OAI, official action indicated. FDA has issued 3 warning letters under this FEI, the latest on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 18 citations from this inspection: 15 under 21 CFR Part 820, the device quality system rules, 2 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 801, the device labeling rules. They are listed below. Two of the three FOIA requests for this site have asked FDA for this inspection's 483.
No court action or import refusal is filed under this FEI.
16 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 12, 2009–2025
- 5 OAI · 4 VAI · 3 NAI
- Inspections with 483 observations shown
- 9 of 12
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Mar 23, 2026
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
12 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices18 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis was not performed and is incomplete.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Show 13 more citations
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by contractors and consultants have not been adequately established.
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
Buildings
Buildings are not of suitable design to perform necessary operations.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of System for Timely Submission of Reports
The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA.
Design history file
The design history file was not established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
Equipment design and installation
Equipment used in the manufacturing process has not been appropriately designed, constructed and installed to facilitate maintenance, adjustment, cleaning, and use.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Info evaluated to determine if event was reportable
The written MDR procedure does not include documentation and recordkeeping requirements for all information that was evaluated to determine if an event was reportable.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Lack of or inadequate In-process acceptance procedures
Procedures for the acceptance and control of in-process product have not been adequately established.
Show 1 more citation
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Product rework documentation, DHR {see also 820.184}
Rework and reevaluation activities have not been documented in the device history record.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices17 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - documentation
The design verification results, including method(s), were not documented in the design history file.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Show 12 more citations
Equipment Installation, Placement, Specified Requirements
The equipment used in the manufacturing process does not meet specified requirements.
Lack of Written MDR Procedures
Written MDR procedures have not been developed and implemented.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers were not evaluated and selected based on their ability to meet specified requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Investigation of device failures
Complaints involving the possible failure of a device and labeling to meet any of its specifications were not where necessary.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Design control - no procedures
Procedures for design control have not been established.
Not approved or obsolete document retrieval
Documents that were not approved were observed at a location where they could be and are being used.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Documentation
Process validation activities and results have not been documented.
Lack of or inadequate In-process acceptance procedures
Procedures for the acceptance of in-process product have not been adequately established.
Equipment Installation, Placement, Specified Requirements
The appropriate design, construction, placement, and installation of manufacturing equipment have not been ensured.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation
Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.
DHR content - manufacturing dates
The device history record does not include manufacturing dates.
Lack of or inadequate In-process acceptance procedures
Acceptance procedures to ensure that specified requirements for in-process product are met were not implemented.
Training records
Employee training is not fully documented.
Lack of or inadequate final acceptance procedures
Procedures for acceptance or rejection of finished device production runs, lots, or batches were not documented.
Show 5 more citations
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.
DHR content
The device history record does not demonstrate the device is manufactured in accordance with the device master record.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not implemented.
Equipment control activity documentation
Equipment calibrations have not been documented.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure that equipment is routinely calibrated were not established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration documentation
There is incomplete documentation of calibration dates, the individual performing each calibration, and the next calibration date for measurement and test equipment .
Equipment suitability & capability
Certain measuring and test equipment is not suitable for its intended purposes or capable of producing valid results.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.
Environmental control Lack of or inadequate procedures
Appropriate procedures have not been followed for controlling environmental conditions.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Equipment suitability & capability
Certain test equipment is not suitable for its intended purposes or capable of producing valid results.
Environmental control Lack of or inadequate procedures
Appropriate procedures have not been defined, documented, and followed for controlling environmental conditions.
Training temp work under special environmental conditions
Maintenance and other personnel who are required to work temporarily under special environmental conditions are not appropriately trained or supervised by a trained individual.
Calibration documentation
There is incomplete documentation of calibration dates, the individual performing each calibration, and the next calibration date for measurement and test equipment .
Nonconforming product, Lack of or inadequate procedures
Procedures were not complete and followed for the control of products that do not conform to specifications.
Show 2 more citations
Contamination control, Lack of or inadequate procedures
Procedures have not been followed to prevent contamination of equipment or product by certain substances.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Mar 23, 2026. Each one asked for a Form 483.
Inspection ending Aug 19, 2025OAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Dec 3, 2025. FDAgent holds no copy that FDA posted.
- Mar 23, 2026FDA-FOIA-2026-2867Form 483Records released
- Sep 30, 2025FDA-FOIA-2025-11466Form 483· 1 of 5 sitesRecords released
Inspection ending Feb 28, 2024VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 30, 2024FDA-FOIA-2024-9627Form 483· 1 of 3 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 18 sections of 21 CFR in FDA’s published citations for 9 of this facility’s 12 inspections, with how many of them cited each.
- 21 CFR 820.70Production and process controls6 of 9
- 21 CFR 820.75Process validation6 of 9
- 21 CFR 820.80Receiving, in-process, and finished device acceptance6 of 9
- 21 CFR 820.90Nonconforming product5 of 9
- 21 CFR 820.100Corrective and preventive action5 of 9
- 21 CFR 820.72Inspection, measuring, and test equipment4 of 9
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?3 of 9
- 21 CFR 820.30Design controls3 of 9
Show all 18
- 21 CFR 820.50Purchasing controls3 of 9
- 21 CFR 820.198Complaint files3 of 9
- 21 CFR 820.22Quality audit2 of 9
- 21 CFR 820.25Personnel2 of 9
- 21 CFR 820.40[Reserved]2 of 9
- 21 CFR 801.20Label to bear a unique device identifier1 of 9
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 9
- 21 CFR 820.20Management responsibility1 of 9
- 21 CFR 820.181Device master record1 of 9
- 21 CFR 820.184Device history record1 of 9
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3007137482 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3007137482. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.