Diasol, Inc

310 S 43rd Ave Ste B, Phoenix, Arizona, United States

FEI 3007137482

FDA has 12 inspections on file for Diasol, Inc in Phoenix, Arizona (FEI 3007137482), from 2009 to 2025. The latest ended on and was classified OAI, official action indicated. FDA has issued 3 warning letters under this FEI, the latest on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 18 citations from this inspection: 15 under 21 CFR Part 820, the device quality system rules, 2 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 801, the device labeling rules. They are listed below. Two of the three FOIA requests for this site have asked FDA for this inspection's 483.

No court action or import refusal is filed under this FEI.

16 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
12, 2009–2025
5 OAI · 4 VAI · 3 NAI
Inspections with 483 observations shown
9 of 12
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
3 requests, latest Mar 23, 2026
Warning letters
· CDRH ↗
3 on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

12 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices18 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    2. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. Design Validation - Risk analysis not performed/inadequate

      Risk analysis was not performed and is incomplete.

    5. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

    Show 13 more citations
    1. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by contractors and consultants have not been adequately established.

    5. Label to bear a unique device identifier

      The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.

    6. Buildings

      Buildings are not of suitable design to perform necessary operations.

    7. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    8. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    9. Lack of System for Timely Submission of Reports

      The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA.

    10. Design history file

      The design history file was not established.

    11. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    12. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been adequately established.

    13. Equipment design and installation

      Equipment used in the manufacturing process has not been appropriately designed, constructed and installed to facilitate maintenance, adjustment, cleaning, and use.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Info evaluated to determine if event was reportable

      The written MDR procedure does not include documentation and recordkeeping requirements for all information that was evaluated to determine if an event was reportable.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    4. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    5. Lack of or inadequate In-process acceptance procedures

      Procedures for the acceptance and control of in-process product have not been adequately established.

    Show 1 more citation
    1. Procedures for product review,disposition lack of/inadequate

      Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been documented in the device history record.

    2. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    3. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    4. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices17 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design verification - documentation

      The design verification results, including method(s), were not documented in the design history file.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    4. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    5. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

    Show 12 more citations
    1. Equipment Installation, Placement, Specified Requirements

      The equipment used in the manufacturing process does not meet specified requirements.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and implemented.

    3. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential suppliers were not evaluated and selected based on their ability to meet specified requirements.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    6. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    7. Investigation of device failures

      Complaints involving the possible failure of a device and labeling to meet any of its specifications were not where necessary.

    8. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    9. Design control - no procedures

      Procedures for design control have not been established.

    10. Not approved or obsolete document retrieval

      Documents that were not approved were observed at a location where they could be and are being used.

    11. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    12. Maintenance schedule, Lack of or inadequate schedule

      Schedules for the adjustment, cleaning, and other maintenance of equipment have not been established.

  5. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. Incoming acceptance records, documentation

      Acceptance or rejection of incoming product was not documented.

  7. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  8. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  9. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    2. Documentation

      Process validation activities and results have not been documented.

    3. Lack of or inadequate In-process acceptance procedures

      Procedures for the acceptance of in-process product have not been adequately established.

    4. Equipment Installation, Placement, Specified Requirements

      The appropriate design, construction, placement, and installation of manufacturing equipment have not been ensured.

  10. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigation

      Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.

    2. DHR content - manufacturing dates

      The device history record does not include manufacturing dates.

    3. Lack of or inadequate In-process acceptance procedures

      Acceptance procedures to ensure that specified requirements for in-process product are met were not implemented.

    4. Training records

      Employee training is not fully documented.

    5. Lack of or inadequate final acceptance procedures

      Procedures for acceptance or rejection of finished device production runs, lots, or batches were not documented.

    Show 5 more citations
    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.

    2. DHR content

      The device history record does not demonstrate the device is manufactured in accordance with the device master record.

    3. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not implemented.

    4. Equipment control activity documentation

      Equipment calibrations have not been documented.

    5. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure that equipment is routinely calibrated were not established.

  11. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Calibration documentation

      There is incomplete documentation of calibration dates, the individual performing each calibration, and the next calibration date for measurement and test equipment .

    2. Equipment suitability & capability

      Certain measuring and test equipment is not suitable for its intended purposes or capable of producing valid results.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.

    4. Environmental control Lack of or inadequate procedures

      Appropriate procedures have not been followed for controlling environmental conditions.

  12. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Equipment suitability & capability

      Certain test equipment is not suitable for its intended purposes or capable of producing valid results.

    2. Environmental control Lack of or inadequate procedures

      Appropriate procedures have not been defined, documented, and followed for controlling environmental conditions.

    3. Training temp work under special environmental conditions

      Maintenance and other personnel who are required to work temporarily under special environmental conditions are not appropriately trained or supervised by a trained individual.

    4. Calibration documentation

      There is incomplete documentation of calibration dates, the individual performing each calibration, and the next calibration date for measurement and test equipment .

    5. Nonconforming product, Lack of or inadequate procedures

      Procedures were not complete and followed for the control of products that do not conform to specifications.

    Show 2 more citations
    1. Contamination control, Lack of or inadequate procedures

      Procedures have not been followed to prevent contamination of equipment or product by certain substances.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Mar 23, 2026. Each one asked for a Form 483.

  • Inspection ending Aug 19, 2025OAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Dec 3, 2025. FDAgent holds no copy that FDA posted.

    1. Mar 23, 2026FDA-FOIA-2026-2867Form 483Records released
    2. Sep 30, 2025FDA-FOIA-2025-11466Form 483· 1 of 5 sitesRecords released
  • Inspection ending Feb 28, 2024VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Oct 30, 2024FDA-FOIA-2024-9627Form 483· 1 of 3 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 18 sections of 21 CFR in FDA’s published citations for 9 of this facility’s 12 inspections, with how many of them cited each.

Show all 18

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3007137482 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3007137482. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.