Exactech, Inc.
2320 Nw 66th Ct, Gainesville, Florida, United States
FEI 1038671
FDA has 14 inspections on file for Exactech, Inc. in Gainesville, Florida (FEI 1038671), from 2008 to 2025. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 9 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two of the seven FOIA requests for this site have asked FDA for this inspection's 483.
FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Seven regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 14, 2008–2025
- 2 OAI · 5 VAI · 7 NAI
- Inspections with 483 observations shown
- 5 of 14
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 7 requests, latest FY2023
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
14 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate and is incomplete.
Design input - documentation
Design input requirements were not adequately documented.
Show 4 more citations
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Process changes - review, evaluation and revalidation
A validated process was not reviewed and evaluated and revalidated when changes or process deviations occurred.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a specification have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Sampling plans
Sampling plans are not based on valid statistical rationale.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been established.
Incorrect translation to production specifications
The device design was not correctly translated into production specifications.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
Show 2 more citations
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Show 2 more citations
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Project Evaluation: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Seven requests to FDA under the Freedom of Information Act named this site, most recently on FY2023.
Inspection ending Aug 27, 2025OAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Dec 3, 2025. FDAgent holds no copy that FDA posted.
- Feb 10, 2026FDA-FOIA-2026-1396Form 483Records released
- Sep 30, 2025FDA-FOIA-2025-11466Form 483· 1 of 5 sitesRecords released
Inspection ending Oct 9, 2024VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Mar 4, 2025FDA-FOIA-2025-1940Form 483· 1 of 4 sitesClosed
Inspection ending Sep 26, 2023OAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Feb 27, 2026. FDAgent holds no copy that FDA posted.
- Feb 27, 2026FDA-FOIA-2026-2028Form 483Records released
- Jan 23, 2024FDA-FOIA-2024-750Form 483Closed
Inspection ending Jan 31, 2020VAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on May 14, 2026. FDAgent holds no copy that FDA posted.
- Apr 14, 2026FDA-FOIA-2026-3761Form 483Records released
Inspection date not given
1 request. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-7765Inspection recordsClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 14 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls4 of 5
- 21 CFR 820.100Corrective and preventive action4 of 5
- 21 CFR 820.50Purchasing controls3 of 5
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 5
- 21 CFR 820.75Process validation2 of 5
- 21 CFR 820.25Personnel1 of 5
- 21 CFR 820.70Production and process controls1 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 5
Show all 12
- 21 CFR 820.90Nonconforming product1 of 5
- 21 CFR 820.181Device master record1 of 5
- 21 CFR 820.198Complaint files1 of 5
- 21 CFR 820.250Statistical techniques1 of 5
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1038671 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1038671. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.