FDA has one inspection on file for Auxilium Medical Inc. in Charleston, South Carolina (FEI 3014650138). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 12 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2025
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Sep 29, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Acceptable supplier records, inadequate records
Records of acceptable suppliers and contractors have not been established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals and with sufficient frequency.
Records, DHR {see also 820.184(e)}
Labels and labeling used for each finished product, lot, or batch, were not sufficiently documented in the DHR.
Show 7 more citations
Supplier notification of changes
There is no agreement with suppliers and contractors to notify you of changes in the product or service.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been established.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
DHR documentation of label release {see also 820.184}
The DHR does not include records of examination and release of device labeling, including date and signature of the examiner.
DMR device specifications
The device master record does not include or refer to the location of device software specifications.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Sep 29, 2025. Each one asked for a Form 483.
Inspection ending Mar 26, 2025VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 8, 2025. FDAgent holds no copy that FDA posted.
- Sep 29, 2025FDA-FOIA-2025-11405Form 483· 1 of 4 sitesRecords released
- Jun 16, 2025FDA-FOIA-2025-7703Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.20Management responsibility1 of 1
- 21 CFR 820.22Quality audit1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
- 21 CFR 820.120Device labeling1 of 1
- 21 CFR 820.181Device master record1 of 1
- 21 CFR 820.184Device history record1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3014650138 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3014650138. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.