Skip to content

Fudakang Industrial Co Ltd

Building 2 501, No. 11 Changheng Road; Changping Town, Dongguan, China

FEI 3008167616

FDA has 3 inspections on file for Fudakang Industrial Co Ltd in Dongguan, China (FEI 3008167616), from 2014 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.

1 import refusal is filed under this FEI, on . No warning letter or court action is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2014–2026
0 OAI · 2 VAI · 1 NAI
Inspections with 483 observations shown
1 of 3
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Aug 28, 2026
Warning letters
None on file
Import refusals
1 refusal
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    2. ISO 13485:2016 Clause 8.3.1

      Records of nonconforming product and actions taken not maintained

      Records of the nature of nonconforming product and any subsequent action taken, including the evaluation, and any investigation and the rationale for decisions, were not adequately maintained.

    3. ISO 13485:2016 Clause 4.1.4

      Changes to the QMS not adequately controlled or evaluated

      Changes to quality management system processes were not controlled in accordance with requirements and evaluated for their impact on the quality management system and the medical devices produced.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Aug 28, 2026. It asked for a Form 483.

  • Inspection ending Mar 12, 2026VAI

    1 request. In the inspection history above.

    1. Aug 28, 2026FDA-FOIA-2026-8360Form 483· 1 of 2 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 1 section of 21 CFR in FDA’s published citations for 1 of this facility’s 3 inspections.

  • 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 1

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3008167616, one row per refused product line.

1 refusal on 2017-10-23 · charged NOT LISTED on 1, FRNMFGREG on 1 · no FDA sample analysis

FDA firm profile
  • 2017-10-23

    THERMOMETER, ELECTRONIC, CLINICAL80FLL

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Northern Border ImportsUPS-4513956-6/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3008167616 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3008167616. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.