21 CFR 806.10: Reports of corrections and removals

FDA's inspection data lists 337 Form 483 citations of 21 CFR 806.10 at 284 facilities, from November 2008 to August 14, 2026. 26 warning letters state a violation of it.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
337
At 329 inspections
Facilities cited
284
Distinct FEIs
Most recent citation
Aug 14, 2026
Vitalcor, Inc.
Warning letters
26
State a violation of it

The regulation

The current text of 21 CFR 806.10 from the Electronic Code of Federal Regulations.

§ 806.10 Reports of corrections and removals.

(a) Each device manufacturer or importer shall submit a written report to FDA of any correction or removal of a device initiated by such manufacturer or importer if the correction or removal was initiated:

(1) To reduce a risk to health posed by the device; or

(2) To remedy a violation of the act caused by the device which may present a risk to health unless the information has already been provided as set forth in paragraph (f) of this section or the corrective or removal action is exempt from the reporting requirements under § 806.1(b).

(b) The manufacturer or importer shall submit any report required by paragraph (a) of this section within 10-working days of initiating such correction or removal.

(c) The manufacturer or importer shall include the following information in the report:

(1) The seven digit registration number of the entity responsible for submission of the report of corrective or removal action (if applicable), the month, day, and year that the report is made, and a sequence number (i.e., 001 for the first report, 002 for the second report, 003 etc.), and the report type designation “C” or “R”. For example, the complete number for the first correction report submitted on June 1, 1997, will appear as follows for a firm with the registration number 1234567: 1234567-6/1/97-001-C. The second correction report number submitted by the same firm on July 1, 1997, would be 1234567-7/1/97-002-C etc. For removals, the number will appear as follows: 1234567-6/1/97-001-R and 1234567-7/1/97-002-R, etc. Firms that do not have a seven digit registration number may use seven zeros followed by the month, date, year, and sequence number (i.e. 0000000-6/1/97-001-C for corrections and 0000000-7/1/97-001-R for removals). Reports received without a seven digit registration number will be assigned a seven digit central file number by the district office reviewing the reports.

(2) The name, address, and telephone number of the manufacturer or importer, and the name, title, address, and telephone number of the manufacturer or importer representative responsible for conducting the device correction or removal.

(3) The brand name and the common name, classification name, or usual name of the device and the intended use of the device.

(4) Marketing status of the device, i.e., any applicable premarket notification number, premarket approval number, or indication that the device is a preamendments device, and the device listing number. A manufacturer or importer that does not have an FDA establishment registration number shall indicate in the report whether it has ever registered with FDA.

(5) The unique device identifier (UDI) that appears on the device label or on the device package, or the device identifier, universal product code (UPC), model, catalog, or code number of the device and the manufacturing lot or serial number of the device or other identification number.

(6) The manufacturer's name, address, telephone number, and contact person if different from that of the person submitting the report.

(7) A description of the event(s) giving rise to the information reported and the corrective or removal actions that have been, and are expected to be taken.

(8) Any illness or injuries that have occurred with use of the device. If applicable, include the medical device report numbers.

(9) The total number of devices manufactured or distributed subject to the correction or removal and the number in the same batch, lot, or equivalent unit of production subject to the correction or removal.

(10) The date of manufacture or distribution and the device's expiration date or expected life.

(11) The names, addresses, and telephone numbers of all domestic and foreign consignees of the device and the dates and number of devices distributed to each such consignee.

(12) A copy of all communications regarding the correction or removal and the names and addresses of all recipients of the communications not provided in accordance with paragraph (c)(11) of this section.

(13) If any required information is not immediately available, a statement as to why it is not available and when it will be submitted.

(d) If, after submitting a report under this part, a manufacturer or importer determines that the same correction or removal should be extended to additional lots or batches of the same device, the manufacturer or importer shall within 10-working days of initiating the extension of the correction or removal, amend the report by submitting an amendment citing the original report number assigned according to paragraph (c)(1) of this section, all of the information required by paragraph (c)(2), and any information required by paragraphs (c)(3) through (c)(12) of this section that is different from the information submitted in the original report. The manufacturer or importer shall also provide a statement in accordance with paragraph (c)(13) of this section for any required information that is not readily available.

(e) A report submitted by a manufacturer or importer under this section (and any release by FDA of that report or information) does not necessarily reflect a conclusion by the manufacturer, importer, or FDA that the report or information constitutes an admission that the device caused or contributed to a death or serious injury. A manufacturer or importer need not admit, and may deny, that the report or information submitted under this section constitutes an admission that the device caused or contributed to a death or serious injury.

(f) No report of correction or removal is required under this part, if a report of the correction or removal is required and has been submitted under parts 803 or 1004 of this chapter.

21 CFR Part 806: Medical Devices; Reports of Corrections and Removals, Subpart BRead on eCFR

Citations by year

Form 483 citations of 21 CFR 806.10 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 38 in 2012

Form 483 citations of 21 CFR 806.10 by inspection year
YearForm 483 citations
20082
200920
201037
201130
201238
201329
201431
201529
201629
201716
201812
20197
20202
202110
20229
202311
20249
202512
20264

How FDA words this citation

FDA's short descriptions and citation text on these 337 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Report of risk to health213 citations
  • Report of violation of the Act (see 803.52(e)(9))55 citations
  • Time to report - 10 days49 citations
  • Device number, identification4 citations
  • Domestic & foreign consignees4 citations
Show 3 more
  • Notification of extended action4 citations
  • Event description3 citations
  • Illness report3 citations

Citation text

  • “A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.”213 citations
  • “A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.”55 citations
  • “A report of the required information regarding device correction and removal actions was not sent to FDA within 10 days of initiating the correction or removal.”49 citations
  • “No amended notification with required information was submitted within 10 working days of initiating an extension of the correction or removal action.”4 citations
Show 2 more
  • “The names , addresses and telephone numbers of all domestic and foreign consignees of devices subject to correction or removal actions and number of devices distributed to each consignee were not reported.”4 citations
  • “The total number of devices subject to correction or removal actions was not reported.”4 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 806.10, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

  • Vitalcor, Inc.2026-08-14 · FEI 1450019Report of risk to health
  • ConMed Corporation2026-04-06 · FEI 3004491689Report of violation of the Act (see 803.52(e)(9))
  • Gentuity, LLC2026-03-05 · FEI 3011578544Report of risk to health
  • SUMMA THERAPEUTICS, LLC2026-02-26 · FEI 3027395815Report of violation of the Act (see 803.52(e)(9))
  • Next Medical Products Company, LLC2025-12-10 · FEI 2523891Report of violation of the Act (see 803.52(e)(9))
  • Intuitive Surgical, Inc.2025-11-05 · FEI 3001675293Report of risk to health
Show 4 more
  • Magellan Diagnostics, Inc.2025-08-26 · FEI 1218996Report of violation of the Act (see 803.52(e)(9))
  • Abiomed, Inc.2025-08-05 · FEI 1220648Report of risk to health
  • ASP Global, LLC.2025-08-01 · FEI 3004824601Report of risk to health
  • Omnia Medical2025-07-28 · FEI 3010560653Report of risk to health

What the observations say

Excerpts from Form 483s that FDA posted in its reading room. A Form 483 usually doesn't name the regulation, so FDAgent ties an observation to 21 CFR 806.10 only when it opens with one of FDA's standard citation texts for this section, word for word apart from case, spacing and punctuation. 1 observation ties this way. The highlighted sentence is the matched wording.

  • Philips Medical Systems Nederland B.V.2025-01-17 · Observation 1

    A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA. Specifically, your firm manufactures the IntelliSpace Cardiovascular (ISCV) software product, an integrated multimodality image and information system designed to perform the necessary functions required for import, export, storage, archival, review, analysis, quantification, reporting and database management of digital …

    This observation opens with FDA's standard wording for 21 CFR 806.10(a)(1).

Warning letters stating a violation of 21 CFR 806.10

26 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.

“Failure to submit a report required by 21 CFR 806.10”
“Failure to submit a Report required by 21 CFR 806.10”

Each letter below states the violation in these words.

“Failure to submit any report required by 21 CFR 806.10(a)”

Often cited at the same inspection

Of the 329 inspections where FDA cited 21 CFR 806.10, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 806.10

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

Topics that count this section:Device CAPA and complaints

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 806.10; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.