Philips Medical Systems Nederland B.V.

High Tech Campus 36, Eindhoven, Netherlands

FEI 3000976525

FDA has one inspection on file for Philips Medical Systems Nederland B.V. in Eindhoven, Netherlands (FEI 3000976525). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA posted the Form 483 from that inspection. FDAgent read its 3 observations from the PDF; they are listed below. One FOIA request has asked FDA for this inspection's 483.

No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Form 483 PDF ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
1, in 2025
Inspections with 483 observations shown
1 of 1
Form 483s posted by FDA
↗
Requested under FOIA
1 request, latest Apr 17, 2025
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesPosted 483 · 3 observations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent’s.

    1. Observation 1

      Unreported correction and removal for health risk

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA. Specifically, your firm manufactures the IntelliSpace Cardiovascular (ISCV) software product, an integrated multimodality image and information system designed to perform the necessary functions required for import, export, storage, archival, review, analysis, quantification, reporting and database management of digital cardiovascular images, waveforms and data related to cardiology. On March of 2023, your firm became aware, through internal testing, of a software defect affecting ISCV Version 7.0 that caused data exports from the ISCV to fail, preventing data from being archived, exported, or copied. Version 7.0 was first installed in the U.S. in April of 2023, after you became aware of the defect. Your firm subsequently received 12 complaints (dated July 2023 - November 2023) from customers reporting issues related to patient studies unable to be archived or transferred, including Complaint #4061517 (dated 13-Nov-2023) alleging that a patient expired due to a delay in transmission of study images possibly resulting in a delay in treatment. Complaint #4061517 states that the "risk was not previously assessed in the RMM [Risk Management Matrix]. The severity of harm has increased from what was predicted in the RMM". Your Correction & Removal Review Board made the decision [illegible] (dated 31-Jan-2024). Your firm sent notification letters to customers (March 2024) and [illegible] in April 2024. However, this recall was not reported to the FDA with [illegible] in the Correction and Removal Strategy (b)(4) (dated 25-Mar-2024) as (b)(4) therefore, this is considered not to be a risk to health." Nevertheless, Complaint #4061517 shows that this software defect resulted in the severity of harm to increase from what was predicted in the RMM. Your firm failed to submit a report to the FDA for this recall that was initiated to reduce the risk to health.

    2. Observation 2

      Inadequate risk analysis for product and security

      Risk analysis is inadequate. Specifically, A. Release of the IntelliSpace Cardiovascular (ISCV) Version 8.0 included a new User Requirement Specification (b)(4) dated 27-Feb-2024) shows the following new requirements (b)(4) were added to the ISCV version 8.0: (b)(4) However, your firm failed to include these new product requirements in your ISCV Product Safety Risk Management Matrix (Document (b)(4) dated 21-Nov-2024) to ensure that hazards associated with these new functions are analyzed and evaluated, and that appropriate risk control measures are implemented as needed. B. The Product Safety Risk Management Surveillance Report, Q1 2020, for IntelliSpace Cardiovascular Product Family (dated 05-OCT-2020) documents the results of the quarterly postmarket surveillance performed for the IntelliSpace Cardiovascular (ISCV). The data sources reviewed included the FDA's MAUDE and Recall databases to review experiences with similar or equivalent competitor devices. Records for devices with the same product code as the ISCV (LLZ) were queried. The search identified a Medical Device Report where a hospital reported a patient death due to the system being down and patient images not accessible. Your firm concluded this was not a new hazard because the Risk Management Matrix includes a comparable hazard "Loss of key functionality"; therefore, "no new hazard or hazardous situation". However, your firm failed to consider the increased severity (death) reported in the Medical Device Report. Furthermore, your firm considers [illegible] for inclusion in the Risk Management Matrix (b)(4) dated 08-Jun-2023). C. Your firm received 12 complaints (dated July 2023 - November 2023) from customers reporting issues with the IntelliSpace Cardiovascular (ISCV) software product related to patient studies unable to be archived or transferred, including Complaint #4061517 (dated 13-Nov-2023) alleging that a patient expired due to a delay in transmission of study images possibly resulting in a delay in treatment. Your firm concluded this Complaint #4061517 presents a new risk not previously evaluated within the Risk Management Matrix (RMM) and added the risk to the RMM. However, the new hazard (b)(4) (b)(4) because your firm concluded that the ISCV was not the direct cause of the patient's death. Your firm failed to consider the potential for patient serious injury or death, as alleged in Complaint #4061517. D. Your IntelliSpace Cardiovascular software product that interfaces with other systems and devices within hospitals' cardiology departments as well as across the enterprise system (e.g., Health Information Systems and Electronic Medical Records). Your firm conducted a security risk analysis to address security and privacy related vulnerabilities of the ISCV, including threat modeling. However, review of the Product Security Risk Management Matrix for the IntelliSpace Cardiovascular 8.0 (dated 17-Apr-2024) showed that the threat modeling conducted was incomplete because it failed to define countermeasures to prevent, mitigate, monitor, or respond to the effects of threats to the ISCV throughout its lifecycle.

    Show observation 3
    1. Observation 3

      Inadequate complaint handling and evaluation procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established. Specifically, A. Review of data downloaded from your (b)(4) reporting IntelliSpace Cardiovascular malfunctions (with no deaths or serious injuries reported) were identified as not reportable as Medical Device Reports ("No" entered under "Potentially Reportable Event"); however, there is no valid justification documented to support the decision to not report these malfunctions, including >4,000 (dated Dec 2022 - Nov 2024) reporting "Incorrect, inaccurate, imprecise, or unexpected result" for cardiovascular studies (e.g., Complaints # 5421845, 5358162, 5021318, 4961060, and 2842278), and >2,000 (dated Aug 2023 - Nov 2024) reporting "Data did not transfer / could not be archived" (e.g., Complaints number 5199415, 4999779, 4562214, 4300649, and 3781836). B. Review of data downloaded from your complaint handling system showed 436 complaints reporting malfunctions of the IntelliSpace Cardiovascular software product where the cause of the failures was traced to software coding issues. Reportedly these complaints are escalated to your (b)(4) and a ticket is initiated to correct the software (b)(4) Dated April 2022). However, review of the complaint data download shows that 127 out of the 436 complaints with software defects were not escalated to (b)(4) For example: PR ID Date Opened Short Description (b)(4) [illegible]

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Apr 17, 2025. It asked for a Form 483.

  • Inspection ending Jan 17, 2025OAI

    1 request. In the inspection history above. FDA has posted its Form 483.

    1. Apr 17, 2025FDA-FOIA-2025-3553Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3000976525 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3000976525. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.