Philips Medical Systems Nederland B.V. Form 483, January 2025

FDA issued this Form 483 to Philips Medical Systems Nederland B.V. in Eindhoven, Netherlands (FEI 3000976525) at the end of a device inspection that ended January 17, 2025. It lists 3 observations. FDA classified the inspection Official Action Indicated (OAI) and issued the facility a warning letter on September 9, 2025.

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Oct 27, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Jan 17, 2025
Devices · Compliance: Devices
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Oct 27, 2026
648 days after the inspection
Facility
FEI 3000976525
Eindhoven, Netherlands

The 3 observations on this Form 483

Read from the FDA-posted Form 483 PDF by OCR. FDAgent keeps an observation only when it matches the scanned text; check the PDF before quoting. The short titles are FDAgent's.

  1. 1

    Unreported correction and removal for health risk

    A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA. Specifically, your firm manufactures the IntelliSpace Cardiovascular (ISCV) software product, an integrated multimodality image and information system designed to perform the necessary functions required for import, export, storage, archival, review, analysis, quantification, reporting and database management of digital cardiovascular images, waveforms and data related to cardiology. On March of 2023, your firm became aware, through internal testing, of a software defect affecting ISCV Version 7.0 that caused data exports from the ISCV to fail, preventing data from being archived, exported, or copied. Version 7.0 was first installed in the U.S. in April of 2023, after you became aware of the defect. Your firm subsequently received 12 complaints (dated July 2023 - November 2023) from customers reporting issues related to patient studies unable to be archived or transferred, including Complaint #4061517 (dated 13-Nov-2023) alleging that a patient expired due to a delay in transmission of study images possibly resulting in a delay in treatment. Complaint #4061517 states that the "risk was not previously assessed in the RMM [Risk Management Matrix]. The severity of harm has increased from what was predicted in the RMM". Your Correction & Removal Review Board made the decision [illegible] (dated 31-Jan-2024). Your firm sent notification letters to customers (March 2024) and [illegible] in April 2024. However, this recall was not reported to the FDA with [illegible] in the Correction and Removal Strategy (b)(4) (dated 25-Mar-2024) as (b)(4) therefore, this is considered not to be a risk to health." Nevertheless, Complaint #4061517 shows that this software defect resulted in the severity of harm to increase from what was predicted in the RMM. Your firm failed to submit a report to the FDA for this recall that was initiated to reduce the risk to health.

  2. 2

    Inadequate risk analysis for product and security

    Risk analysis is inadequate. Specifically, A. Release of the IntelliSpace Cardiovascular (ISCV) Version 8.0 included a new User Requirement Specification (b)(4) dated 27-Feb-2024) shows the following new requirements (b)(4) were added to the ISCV version 8.0: (b)(4) However, your firm failed to include these new product requirements in your ISCV Product Safety Risk Management Matrix (Document (b)(4) dated 21-Nov-2024) to ensure that hazards associated with these new functions are analyzed and evaluated, and that appropriate risk control measures are implemented as needed. B. The Product Safety Risk Management Surveillance Report, Q1 2020, for IntelliSpace Cardiovascular Product Family (dated 05-OCT-2020) documents the results of the quarterly postmarket surveillance performed for the IntelliSpace Cardiovascular (ISCV). The data sources reviewed included the FDA's MAUDE and Recall databases to review experiences with similar or equivalent competitor devices. Records for devices with the same product code as the ISCV (LLZ) were queried. The search identified a Medical Device Report where a hospital reported a patient death due to the system being down and patient images not accessible. Your firm concluded this was not a new hazard because the Risk Management Matrix includes a comparable hazard "Loss of key functionality"; therefore, "no new hazard or hazardous situation". However, your firm failed to consider the increased severity (death) reported in the Medical Device Report. Furthermore, your firm considers [illegible] for inclusion in the Risk Management Matrix (b)(4) dated 08-Jun-2023). C. Your firm received 12 complaints (dated July 2023 - November 2023) from customers reporting issues with the IntelliSpace Cardiovascular (ISCV) software product related to patient studies unable to be archived or transferred, including Complaint #4061517 (dated 13-Nov-2023) alleging that a patient expired due to a delay in transmission of study images possibly resulting in a delay in treatment. Your firm concluded this Complaint #4061517 presents a new risk not previously evaluated within the Risk Management Matrix (RMM) and added the risk to the RMM. However, the new hazard (b)(4) (b)(4) because your firm concluded that the ISCV was not the direct cause of the patient's death. Your firm failed to consider the potential for patient serious injury or death, as alleged in Complaint #4061517. D. Your IntelliSpace Cardiovascular software product that interfaces with other systems and devices within hospitals' cardiology departments as well as across the enterprise system (e.g., Health Information Systems and Electronic Medical Records). Your firm conducted a security risk analysis to address security and privacy related vulnerabilities of the ISCV, including threat modeling. However, review of the Product Security Risk Management Matrix for the IntelliSpace Cardiovascular 8.0 (dated 17-Apr-2024) showed that the threat modeling conducted was incomplete because it failed to define countermeasures to prevent, mitigate, monitor, or respond to the effects of threats to the ISCV throughout its lifecycle.

  3. 3

    Inadequate complaint handling and evaluation procedures

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established. Specifically, A. Review of data downloaded from your (b)(4) reporting IntelliSpace Cardiovascular malfunctions (with no deaths or serious injuries reported) were identified as not reportable as Medical Device Reports ("No" entered under "Potentially Reportable Event"); however, there is no valid justification documented to support the decision to not report these malfunctions, including >4,000 (dated Dec 2022 - Nov 2024) reporting "Incorrect, inaccurate, imprecise, or unexpected result" for cardiovascular studies (e.g., Complaints # 5421845, 5358162, 5021318, 4961060, and 2842278), and >2,000 (dated Aug 2023 - Nov 2024) reporting "Data did not transfer / could not be archived" (e.g., Complaints number 5199415, 4999779, 4562214, 4300649, and 3781836). B. Review of data downloaded from your complaint handling system showed 436 complaints reporting malfunctions of the IntelliSpace Cardiovascular software product where the cause of the failures was traced to software coding issues. Reportedly these complaints are escalated to your (b)(4) and a ticket is initiated to correct the software (b)(4) Dated April 2022). However, review of the complaint data download shows that 127 out of the 436 complaints with software defects were not escalated to (b)(4) For example: PR ID Date Opened Short Description (b)(4) [illegible]

After this inspection

FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • The observations are the text FDAgent read from this PDF by OCR, shown only when every printed observation passed FDAgent’s checks against the scanned text.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.