FDA has one inspection on file for SUMMA THERAPEUTICS, LLC in Allston, Massachusetts (FEI 3027395815). It ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 8 citations from this inspection. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 1, in 2026
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest May 25, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.5.6
Procedures for validation of processes not documented
Procedures for validation of processes were not adequately documented.
- ISO 13485:2016 Clause 8.5.2
Corrective action procedures not documented
Corrective action procedures defining the requirements for reviewing the effectiveness of corrective action taken have not been adequately documented.
- ISO 13485:2016 Clause 7.5.7
Sterilization and sterile barrier system validation procedures not documented
Procedures for the validation of processes for sterilization were not adequately documented.
- ISO 13485:2016 Clause 7.3.8
Design and development transfer procedures lack required elements
Procedures for transfer of design and development outputs to manufacturing do not adequately ensure that design and development outputs are verified as suitable for manufacturing before becoming final production specifications.
- ISO 13485:2016 Clause 4.1.5
Control over outsourced processes not proportionate to risk
Controls over outsourced processes that affect product conformity to requirements are not proportionate to the risk involved.
Show 3 more citations
- ISO 13485:2016 Clause 8.2.1
Feedback process procedures not documented
Procedures for the feedback process have not been documented.
Report of violation of the Act (see 803.52(e)(9))
A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.
- ISO 13485:2016 Clause 7.3.7
Design and development validation not performed as planned
Design and development validation was not adequately performed in accordance with planned and documented arrangements to ensure that the resulting product is capable of meeting the requirements for the specified application or intended use.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on May 25, 2026. Each one asked for a Form 483.
Inspection ending Feb 26, 2026NAI
3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 31, 2026. FDAgent holds no copy that FDA posted.
- May 25, 2026FDA-FOIA-2026-5254Form 483Records released
- May 19, 2026FDA-FOIA-2026-5134Form 483· 1 of 4 sitesRecords released
- Apr 24, 2026FDA-FOIA-2026-4114Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 1 section of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 806.10Reports of corrections and removals1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3027395815 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3027395815. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.