Intuitive Surgical, Inc.
1266 Kifer Rd, Sunnyvale, California, United States
FEI 3001675293
FDA has 10 inspections on file for Intuitive Surgical, Inc. in Sunnyvale, California (FEI 3001675293), from 2010 to 2025. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 6 citations from this inspection: 4 under 21 CFR Part 820, the device quality system rules, 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below. Two of the five FOIA requests for this site have asked FDA for this inspection's 483.
FDA issued a warning letter on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 10, 2010–2025
- 2 OAI · 4 VAI · 4 NAI
- Inspections with 483 observations shown
- 6 of 10
- 1 posted 483 not matched to an inspection
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 5 requests, latest May 25, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
10 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Show 1 more citation
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
- NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Availability
Required records were not made readily available for review and copying by the FDA and are not legible.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Complaints
Complaint files are not maintained.
Show 1 more citation
Evaluation, timeliness, identification
Complaints representing events that are MDR reportable were not clearly identified.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Illness report
Illnesses or injuries that have occurred with use of devices subject to corrections or removals have not been reported.
Design input - documentation
Design input requirements were not adequately documented.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of procedures, or not maintained
Document control procedures have not been maintained.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Five requests to FDA under the Freedom of Information Act named this site, most recently on May 25, 2026.
Inspection ending Nov 5, 2025OAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Mar 11, 2026. FDAgent holds no copy that FDA posted.
- May 25, 2026FDA-FOIA-2026-5253Form 483Records released
- Jan 8, 2026FDA-FOIA-2026-267Form 483· 1 of 5 sitesRecords released
Inspection ending Dec 14, 2023VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its EIR on Apr 23, 2026. FDAgent holds no copy that FDA posted.
- Jan 14, 2026FDA-FOIA-2026-502EIRRecords released
- Jan 3, 2024FDA-FOIA-2024-62Form 483Closed
Inspection date not given
1 request. FDA released records in answer to a request for its Form 483 + EIR on Mar 12, 2026. FDAgent holds no copy that FDA posted.
- Dec 4, 2025FDA-FOIA-2025-13609Form 483 + EIRRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 10 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action4 of 6
- 21 CFR 820.30Design controls3 of 6
- 21 CFR 806.10Reports of corrections and removals2 of 6
- 21 CFR 820.50Purchasing controls2 of 6
- 21 CFR 820.198Complaint files2 of 6
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 6
- 21 CFR 820.40[Reserved]1 of 6
- 21 CFR 820.70Production and process controls1 of 6
Show all 11
- 21 CFR 820.75Process validation1 of 6
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 6
- 21 CFR 820.180General requirements1 of 6
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3001675293, one row per refused product line.
1 refusal on 2020-02-07 · charged NOT LISTED on 1 · no FDA sample analysis
FDA firm profile- 2020-02-07
SYSTEM, SURGICAL, COMPUTER CONTROLLED INSTRUMENT79NAY
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-5312943-2/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001675293 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001675293. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.