21 CFR 820.180: General requirements

FDA's inspection data lists 119 Form 483 citations of 21 CFR 820.180 at 113 facilities, from October 2008 to December 5, 2025. 2 warning letters state a violation of it. eCFR shows this section removed on February 4, 2026.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
119
At 117 inspections
Facilities cited
113
Distinct FEIs
Most recent citation
Dec 5, 2025
Jedmed Instrument Company
Warning letters
2
State a violation of it

The regulation

Removed from the CFR on 2026-02-04. The text shown is the last text in force before then (eCFR version of 2016-12-30).

§ 820.180 General requirements.

All records required by this part shall be maintained at the manufacturing establishment or other location that is reasonably accessible to responsible officials of the manufacturer and to employees of FDA designated to perform inspections. Such records, including those not stored at the inspected establishment, shall be made readily available for review and copying by FDA employee(s). Such records shall be legible and shall be stored to minimize deterioration and to prevent loss. Those records stored in automated data processing systems shall be backed up.

(a) Confidentiality. Records deemed confidential by the manufacturer may be marked to aid FDA in determining whether information may be disclosed under the public information regulation in part 20 of this chapter.

(b) Record retention period. All records required by this part shall be retained for a period of time equivalent to the design and expected life of the device, but in no case less than 2 years from the date of release for commercial distribution by the manufacturer.

(c) Exceptions. This section does not apply to the reports required by § 820.20(c) Management review, § 820.22 Quality audits, and supplier audit reports used to meet the requirements of § 820.50(a) Evaluation of suppliers, contractors, and consultants, but does apply to procedures established under these provisions. Upon request of a designated employee of FDA, an employee in management with executive responsibility shall certify in writing that the management reviews and quality audits required under this part, and supplier audits where applicable, have been performed and documented, the dates on which they were performed, and that any required corrective action has been undertaken.

21 CFR Part 820: Quality Management System Regulation, Subpart MRead on eCFR

Citations by year

Form 483 citations of 21 CFR 820.180 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 15 in 2011

Form 483 citations of 21 CFR 820.180 by inspection year
YearForm 483 citations
20084
200911
201013
201115
20129
20136
20146
20155
201615
20174
20188
20193
20216
20223
20233
20242
20256

How FDA words this citation

FDA's short descriptions and citation text on these 119 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Availability76 citations
  • Retention period33 citations
  • Legibility6 citations
  • Storage4 citations

Citation text

  • “Required records were not made readily available for review and copying by the FDA.”19 citations
  • “Required records are not retained for the design and expected life of the device.”17 citations
  • “Required records are not maintained at a location that is reasonably accessible to responsible officials of the manufacturer and to employees of the FDA and were not made readily available for review and copying by the FDA.”14 citations
  • “Required records are not maintained at a location that is reasonably accessible to responsible officials of the manufacturer and to employees of the FDA.”10 citations
Show 2 more
  • “Required records are not stored to minimize deterioration and prevent loss.”10 citations
  • “Required records are not legible.”9 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 820.180, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

Show 4 more

Warning letters stating a violation of 21 CFR 820.180

2 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.

“Failure to adequately maintain records at a location that is reasonably accessible to responsible officials of the manufacturer and to employees of the FDA, as required by 21 CFR 820.180.”
“Failure to adequately maintain all records required by 21 CFR 820 at the manufacturing establishment or other location that is reasonably accessible to responsible officials of the manufacturer and to employees of FDA designated to perform inspections. Such records, including those not stored at the inspected establishment, shall be made readily available for review and copying by FDA employee(s), as required by 21 CFR 820.180.”

Often cited at the same inspection

Of the 117 inspections where FDA cited 21 CFR 820.180, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 820.180

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 820.180; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.