Contec Medical Systems Co., Ltd.
Economic And, No. 112 Qinhuang West Street; Technical Development Zone, Qinhuangdao, China
FEI 3006979678
FDA has 4 inspections on file for Contec Medical Systems Co., Ltd. in Qinhuangdao, China (FEI 3006979678), from 2012 to 2025. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
28 import refusals are filed under this FEI, the latest on . FDA’s Red Lists for Import Alerts 89-04 and 89-08 name this firm at this address. FDA printed no FEI on the entries. No court action is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.198(a).
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 4, 2012–2025
- 1 OAI · 1 VAI · 2 NAI
- Inspections with 483 observations shown
- 2 of 4
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Sep 9, 2025
- Warning letters
- · CDRH ↗
- Import refusals
- 28 refusals
- Import alerts
- 89-04, 89-08Red List
- Matched on FDA’s firm name and street address
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Availability
Required records are not maintained at a location that is reasonably accessible to responsible officials of the manufacturer and to employees of the FDA.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Document change records, maintained.
Records of changes to documents were not adequately maintained.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with.
Production processes
Production processes were not conducted, controlled, and monitored to ensure that a device conforms to its specifications.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Sep 9, 2025. It asked for a Form 483.
Inspection ending Jun 12, 2025OAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Dec 9, 2025. FDAgent holds no copy that FDA posted.
- Sep 9, 2025FDA-FOIA-2025-10777Form 483· 1 of 2 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.180General requirements1 of 2
- 21 CFR 820.184Device history record1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3006979678, one row per refused product line.
28 refusals from 2016-12-01 to 2026-04-14 · charged NOT LISTED on 9, FRNMFGREG on 9 · no FDA sample analysis
FDA firm profile- 2026-04-14
CUFF, BLOOD-PRESSURE74DXQ
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of West Coast ImportsAEK-0359110-3/10002/1
- 2026-03-18
SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN
DEVGMPS 520(f); 801(a)(1)
Division of Southeast Imports1FX-0736395-4/10003/1
- 2026-03-17
DOPPLER, FETAL, ULTRASOUND90LXE
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast Imports799-5006889-6/20003/1
- 2026-01-09
OXIMETER73DQA
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-2202087-3/40004/1
- 2025-12-04
OXIMETER73DQA
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
Division of Southeast ImportsK80-5154281-0/10003/1
Show 23 more
- 2025-12-01
OXIMETER, INFRARED, SPORTING, AVIATION74OCH
DEVGMPS 520(f); 801(a)(1)
Division of Southeast Imports442-6847375-6/20004/1
- 2025-12-01
OXIMETER, INFRARED, SPORTING, AVIATION74OCH
DEVGMPS 520(f); 801(a)(1)
Division of Southeast Imports442-6847375-6/10004/1
- 2025-11-26
SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN
DEVICEGMPS 501(h), 801(a)(1); ADULTERATION
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Northern Border Imports416-0186137-4/10004/1
- 2025-07-03
OXIMETER73DQA
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-8972437-3/33/1
- 2025-07-03
OXIMETER73DQA
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-8972437-3/13/1
- 2025-06-13
SMALL HOSPITAL PRACTICE EQUIPMENT, NEC (NONHUMAN)68NYG99
INCONSPICU 502(c), 801(a)(3); MISBRANDING
Division of Northern Border Imports106-0316683-5/52/1
- 2025-06-13
MONITOR, CARDIAC (INCL. CARDIOTACHOMETER & RATE ALARM)74DRT
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Northern Border Imports106-0316683-5/42/1
- 2025-03-28
OXIMETER73DQA
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsSCS-9353765-6/14/1
- 2025-02-27
MONITOR, PHYSIOLOGICAL, PATIENT74MHX
NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION
Division of Northeast ImportsARE-3872738-9/22/1
- 2025-02-27
MONITOR, PHYSIOLOGICAL, PATIENT74MHX
NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION
Division of Northeast ImportsARE-3872738-9/12/1
- 2024-12-30
OXIMETER73DQA
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsAEK-8025333-7/52/1
- 2023-09-22
CANNULA, NASAL, OXYGEN73CAT
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports231-0213130-4/11/1
- 2020-11-20
OXIMETER73DQA
DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING
LACKS FIRM 502(b)(1), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsSCS-2720010-6/11/1
- 2020-08-18
OXIMETER73DQA
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsAEK-7716180-7/12/1
- 2020-07-02
CATHETER, OXIMETER, FIBEROPTIC74DQE
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsAEK-7542911-5/11/1
- 2020-07-02
OXIMETER73DQA
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsAEK-7530419-3/11/1
- 2020-06-24
OXIMETER73DQA
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsSCS-0671575-1/12/1
- 2020-06-19
OXIMETER73DQA
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsSCS-0637055-7/12/1
- 2020-06-02
OXIMETER73DQA
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsSCS-0413409-6/12/1
- 2019-07-02
OXIMETER73DQA
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsAEK-6061703-9/11/1
- 2018-05-07
OXIMETER73DQA
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Northern Border ImportsUPS-8747149-7/11/1
- 2018-02-09
OXIMETER, WELLNESS74PGJ
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-3939556-9/11/2
- 2016-12-01
MONITOR, PHYSIOLOGICAL, PATIENT74MHX
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsAEK-2764414-2/11/1
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov
Date published 2025-10-02
- FDA’s entry
- Contec Medical Systems Co., Ltd. · Economic And, No. 112 Qinhuang West Street; Technical Development Zone, Qinhuangdao, Hebei CHINA
- FDA’s inspection record for FEI 3006979678
- Contec Medical Systems Co., Ltd. · Economic And, No. 112 Qinhuang West Street; Technical Development Zone, Qinhuangdao, China
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
19 products
- 73 - - - --Anesthesiology
- 74 - - - --Cardiovascular
- 75 - - - --Chemistry
Show 16 more products
- 76 - - - --Dental
- 77 - - - --Ear,Nose And Throat
- 78 - - - --Gastroenterological & Urological
- 79 - - - --General & Plastic Surgery
- 80 - - - --General Hospital/Personal Use
- 81 - - - --Hematology
- 82 - - - --Immunology
- 83 - - - --Microbiology
- 84 - - - --Neurological
- 85 - - - --Obstetrical & Gynecological
- 86 - - - --Ophthalmic
- 87 - - - --Orthopedic
- 88 - - - --Pathology
- 89 - - - --Physical Medicine
- 90 - - - --Radiology
- 91 - - - --Toxicology
- 89-08Red ListDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Import alert pageAlert on fda.gov
Date published 2025-01-24
- FDA’s entry
- Contec Medical Systems Co., Ltd. · Economic And, No. 112 Qinhuang West Street; Technical Development Zone, Qinhuangdao, Hebei CHINA
- FDA’s inspection record for FEI 3006979678
- Contec Medical Systems Co., Ltd. · Economic And, No. 112 Qinhuang West Street; Technical Development Zone, Qinhuangdao, China
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
12 products
- 68 N - - 99Small Hospital Practice Equipment, NEC (Nonhuman)
- 68 Z - - 99Animal Medical Devices, NEC
- 73 B - - ZQMonitor, Breathing Frequency
Show 9 more products
- 73 C - - CKAnalyzer, Gas, Carbon-Dioxide, Gaseous-Phase
- 73 D - - QAOximeter
- 74 D - - RTMonitor, Cardiac (Incl. Cardiotachometer & Rate Alarm)
- 74 D - - SIDetector And Alarm, Arrhythmia
- 74 D - - SKComputer, Blood-Pressure
- 74 D - - XNSystem, Measurement, Blood-Pressure, Non-Invasive
- 74 M - - HXMonitor, Physiological, Patient
- 74 M - - LDMonitor, St Segment with Alarm
- 80 F - - LLThermometer, Electronic, Clinical
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006979678 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006979678. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.