Skip to content

Vitalcor, Inc.

100 Chestnut Ave, Westmont, Illinois, United States

FEI 1450019

FDA has 2 inspections on file for Vitalcor, Inc. in Westmont, Illinois (FEI 1450019), from 2017 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 9 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 803.17.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
2, 2017–2026
0 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Oct 16, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 8.5.2

      Corrective action procedures not documented

      Corrective action procedures defining the requirements for evaluating the need for action to ensure that nonconformities do not recur, planning and documenting action needed and implementing such action and reviewing the effectiveness of corrective action taken have not been adequately documented.

    2. ISO 13485:2016 Clause 8.2.2

      Complaint handling procedures not documented

      Procedures for timely complaint handling have not been adequately documented.

    3. ISO 13485:2016 Clause 4.2.4

      Lack of document control

      Documents required by the quality management system have not been adequately controlled.

    4. ISO 13485:2016 Clause 4.1.5

      Inadequate monitoring and control over outsourced processes

      Outsourced processes that affect product conformity to requirements were not adequately monitored and controlled.

    5. Lack of Written MDR Procedures

      Written MDR procedures have not been developed, maintained and implemented.

    Show 4 more citations
    1. Devices subject to device identification GUDID data submission requirements.

      The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.

    2. ISO 13485:2016 Clause 5.6.1

      Management review procedures not documented

      Procedures for management review have not been adequately documented.

    3. Label to bear a unique device identifier

      The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.

    4. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers have not been adequately established.

    2. Design control - no procedures

      Procedures for design control have not been established.

    3. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    5. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system.

    Show 3 more citations
    1. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    2. Lack of procedures, or not maintained

      Document control procedures have not been established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Oct 16, 2026. Each one asked for a Form 483.

  • Inspection ending Aug 14, 2026VAI

    2 requests. In the inspection history above.

    1. Oct 16, 2026FDA-FOIA-2026-10081Form 483· 1 of 5 sitesOpen
    2. Oct 6, 2026FDA-FOIA-2026-9718Form 483· 1 of 5 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Show all 11

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1450019 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1450019. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.