21 CFR 830.300: Devices subject to device identification data submission requirements

FDA's inspection data lists 112 Form 483 citations of 21 CFR 830.300 at 112 facilities, from June 2023 to September 16, 2026. 3 warning letters state a violation of it.

Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.

Form 483 citations
112
At 112 inspections
Facilities cited
112
Distinct FEIs
Most recent citation
Sep 16, 2026
Maymom LLC
Warning letters
3
State a violation of it

The regulation

The current text of 21 CFR 830.300 from the Electronic Code of Federal Regulations.

§ 830.300 Devices subject to device identification data submission requirements.

(a) In general. The labeler of a device must provide the information required by this subpart for each version or model required to bear a unique device identifier (UDI).

(b) Voluntary submission of information. If a labeler voluntarily includes a UDI on the label of a device under § 801.40, the labeler may also voluntarily submit information concerning that device under this part.

(c) Exclusions. FDA may reject or remove any device identification data where:

(1) The device identifier submitted does not conform to § 830.20;

(2) The information concerns a device that is neither manufactured in the United States nor in interstate commerce in the United States,

(3) The information concerns a product that FDA determines is not a device or a combination product that includes a device constituent part,

(4) The information concerns a device or a combination product that requires, but does not have, FDA premarket approval, licensure, or clearance;

(5) A device that FDA has banned under section 516 of the Federal Food, Drug, and Cosmetic Act; or

(6) FDA has suspended the accreditation of the issuing agency that operates the system used by the labeler.

21 CFR Part 830: Unique Device Identification, Subpart ERead on eCFR

Citations by year

Form 483 citations of 21 CFR 830.300 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).

Peak: 43 in 2025

Form 483 citations of 21 CFR 830.300 by inspection year
YearForm 483 citations
202315
202434
202543
202620

How FDA words this citation

FDA's short descriptions and citation text on these 112 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.

Short descriptions

  • Devices subject to device identification GUDID data submission requirements.112 citations

Citation text

  • “The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.”112 citations

Facilities cited most recently

The latest inspections where FDA cited 21 CFR 830.300, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.

  • Maymom LLC2026-09-16 · FEI 3010196750Devices subject to device identification GUDID data submission requirements.
  • Vitalcor, Inc.2026-08-14 · FEI 1450019Devices subject to device identification GUDID data submission requirements.
  • ECLOSION2026-08-13 · FEI 3030198312Devices subject to device identification GUDID data submission requirements.
  • Angstrom Manufacturing, Inc.2026-08-12 · FEI 3003610374Devices subject to device identification GUDID data submission requirements.
  • CRYON-X LLC2026-08-12 · FEI 3033648890Devices subject to device identification GUDID data submission requirements.
  • Hospitec Usa2026-07-09 · FEI 3008009298Devices subject to device identification GUDID data submission requirements.
Show 4 more
  • MANSBACH HEALTH TOOLS LLC2026-06-02 · FEI 3035111095Devices subject to device identification GUDID data submission requirements.
  • Bluewave Technologies Inc2026-05-22 · FEI 3029300415Devices subject to device identification GUDID data submission requirements.
  • Perry Dynamics Inc2026-05-20 · FEI 3004979815Devices subject to device identification GUDID data submission requirements.
  • Marlen Manufacturing & Development Co.2026-05-06 · FEI 1520020Devices subject to device identification GUDID data submission requirements.

Warning letters stating a violation of 21 CFR 830.300

3 warning letters FDA posted state a violation of this section in a numbered violation sentence. The most recent are below, each linking to FDAgent's page for the letter.

“Failure to ensure that the labeler of a device provide the information required by the subpart for each version or model required to bear a unique device identifier, as required by 21 CFR 830.300.”
“Failure to provide required information to GUDID as required by 21 CFR 830.300(a).”
“Failure to provide required information to Global Unique Device Identification Database (GUDID) as required by 21 CFR 830.300(a).”

Often cited at the same inspection

Of the 112 inspections where FDA cited 21 CFR 830.300, the share where it also cited each of these sections.

Ask FDAgent about 21 CFR 830.300

FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
  • Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
  • A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 830.300; FDAgent quotes that sentence exactly.
  • FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.